Thromboembolic complications with Pipeline Embolization Device placement: impact of procedure time, number of stents and pre-procedure P2Y12 reaction unit (PRU) value
The DOI matches its Crossref record; 13 of the 13 references it cites were found in a registry or an index.
Abstract
Background: Thromboembolic events after Pipeline Embolization Device (PED) placement remain a feared complication among neuroendovascular surgeons. This study aimed to investigate potential risk factors for thromboembolic events in patients undergoing PED placement. Methods: Medical records of patients who underwent PED placement from April 2011 to August 2013 were reviewed. Variables including pre-procedure P2Y12 reaction unit (PRU) value, procedure time, number of PEDs deployed and perioperative neurovascular complications were recorded. Multivariate analysis was performed to identify risk factors for perioperative thromboembolic complications. Results: Seventy-four patients were identified. Six patients (8.1%) had changes in neurological status after PED placement including five (6.8%) thromboembolic complications and one (1.4%) delayed intracranial hemorrhage; 50.9% of patients had diffusion-weighted imaging (DWI) changes on post-procedural MRI. Longer procedure time (>116 min) and multiple PED placements (>1) were statistically significant risk factors for symptomatic thromboembolic events (p<0.01). A pre-procedural PRU value >208 had an OR of 11.32 (95% CI 0.06 to 212.57) for symptomatic thromboembolic complications, but the result was not statistically significant. Conclusions: DWI changes on MRI occurred at a much higher rate than new neurological symptoms following PED placement. Longer procedure time and multiple PED deployment are associated with higher risks of new neurological changes due to thromboembolic events. There was a trend for an increased risk of a symptomatic thromboembolic event in patients with pre-procedural PRU values >208. Reloading (clopidogrel 600 mg) patients with preoperative PRU >208 was safe and may have a protective effect on thromboembolic events.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Bonello, L., Tantry, U. S., Marcucci, R., Blindt, R., Angiolillo, D. J., Becker, R., Bhatt, D. L., Cattaneo, M., Collet, J. P., Cuisset, T., et al. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2010.04.047 · Cited by 1,149
Bonello, L., Camoin-Jau, L., Arques, S., Boyer, C., Panagides, D., Wittenberg, O., Simeoni, M.-C., Barragan, P., Dignat-George, F., & Paganelli, F. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2007.12.044 · Cited by 556
Price, M. J., Angiolillo, D. J., Teirstein, P. S., Lillie, E., Manoukian, S. V., Berger, P. B., Tanguay, J.-F., Cannon, C. P., & Topol, E. J. (2011) · Circulation · doi:10.1161/circulationaha.111.029165 · Cited by 396
Campo, G., Parrinello, G., Ferraresi, P., Lunghi, B., Tebaldi, M., Miccoli, M., Marchesini, J., Bernardi, F., Ferrari, R., & Valgimigli, M. (2011) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2010.12.047 · Cited by 331
Antiplatelet effects of clopidogrel and bleeding in patients undergoing coronary stent placement
SIBBING, D., SCHULZ, S., BRAUN, S., MORATH, T., STEGHERR, J., MEHILLI, J., SCHÖMIG, A., VON BECKERATH, N., & KASTRATI, A. (2010) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2009.03709.x · Cited by 324
Bonello, L., Camoin-Jau, L., Armero, S., Com, O., Arques, S., Burignat-Bonello, C., Giacomoni, M.-P., Bonello, R., Collet, F., Rossi, P., et al. (2009) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2008.08.048 · Cited by 276
Lee, D. H., Arat, A., Morsi, H., Shaltoni, H. M., Harris, J. R., & Mawad, M. E. (2008) · American Journal of Neuroradiology · doi:10.3174/ajnr.a1070 · Cited by 142
Price, M. J., Berger, P. B., Angiolillo, D. J., Teirstein, P. S., Tanguay, J., Kandzari, D. E., Cannon, C. P., & Topol, E. J. (2009) · American Heart Journal · doi:10.1016/j.ahj.2009.02.012 · Cited by 126
Bonello, L., Arméro, S., Mokhtar, O. A., Mancini, J., Aldebert, P., Saut, N., Bonello‐Palot, N., Barragan, P. T., Arquès, S., Giacomoni, M.-P., et al. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2010.07.004 · Cited by 121
Akbari, S. H. A., Reynolds, M. R., Kadkhodayan, Y., Cross, D. T., & Moran, C. J. (2012) · Journal of NeuroInterventional Surgery · doi:10.1136/neurintsurg-2012-010334 · Cited by 117
Metadata
- Authors
- Tan, L. A., Keigher, K. M., Munich, S. A., Moftakhar, R., & Lopes, D. K.
- Venue
- Journal of NeuroInterventional Surgery
- Publisher
- BMJ
- Published
- 2014-02-19
- Volume
- 7
- Issue
- 3
- Pages
- 217-221
- ISSN
- 1759-8478 (ISSN-L)
- 1759-8486
- OpenAlex
- W2161367561
- Cited by
- 164
- References
- 13
- FWCI
- 9.31
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Thromboembolic complications with Pipeline Embolization Device placement: impact of procedure time, number of stents and pre-procedure P2Y12 reaction unit (PRU) value. https://app.cochat.ai/papers/10.1136/neurintsurg-2014-011111
- The paper
- Tan, L. A., Keigher, K. M., Munich, S. A., Moftakhar, R., & Lopes, D. K. (2015). Thromboembolic complications with Pipeline Embolization Device placement: impact of procedure time, number of stents and pre-procedure P2Y12 reaction unit (PRU) value. Journal of NeuroInterventional Surgery. https://doi.org/10.1136/neurintsurg-2014-011111
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.