Bone Healing Monitoring in Bone Lengthening Using Bioimpedance
Open accessThe DOI matches its Crossref record; 35 of the 35 references it cites were found in a registry or an index.
Abstract
The most common technique of orthopedic surgical procedure for the correction of deformities is bone lengthening by "distraction osteogenesis," which requires periodic and ongoing bone assessment following surgery. Bone impedance is a noninvasive, quantitative method of assessing bone fracture healing. The purpose of this study was to monitor bone healing and determine when fixation devices should be removed. The left tibia of eight male New Zealand white rabbits (2.4 ± 0.4 kg) undergoing osteotomy was attached with a mini-external fixator. The bone length was increased by 1 cm one week after surgery by distracting it 1 mm per day. Before and after osteotomy, as well as every week after, bone impedance was measured in seven frequency ranges using an EVAL-AD5933EBZ board. Three orthopedic surgeons analyzed the radiographs using the Radiographic Union Scale for Tibial (RUST) score. The Kappa Fleiss coefficient was used to determine surgeon agreement, and the Spearman rank correlation coefficient was used to find out the relationship between impedance measurements and RUST scores. Finally, the device removal time was calculated by comparing the bone impedance to the preosteotomy impedance. The agreement of three orthopedic surgeons on radiographs had a Fleiss' Kappa coefficient of 49%, indicating a moderate level of agreement. The Spearman rank correlation coefficient was 0.43, indicating that impedance and radiographic techniques have a direct relationship. Impedance is expected to be used to monitor fractured or lengthened bones in a noninvasive, low-cost, portable, and straightforward manner. Furthermore, when used in conjunction with other qualitative methods such as radiography, impedance can be useful in determining the precise time of device removal.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures












Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The Measurement of Observer Agreement for Categorical Data
Landis, J. R., & Koch, G. G. (1977) · Biometrics · doi:10.2307/2529310 · Cited by 81,148 · cited 2×
“-e Fleiss’ Kappa coefficient was used to evaluate the reliability of agreement among three surgeons [25].”
MethodImproving Bioscience Research Reporting: The ARRIVE Guidelines for Reporting Animal Research
Kilkenny, C., Browne, W. J., Cuthill, I. C., Emerson, M., & Altman, D. G. (2010) · PLoS Biology · doi:10.1371/journal.pbio.1000412 · Cited by 7,300
Improving bioscience research reporting: The ARRIVE guidelines for reporting animal research
Kilkenny, C., Browne, W. J., Cuthill, I. C., Emerson, M., & Altman, D. G. (2010) · Journal of Pharmacology and Pharmacotherapeutics · doi:10.4103/0976-500x.72351 · Cited by 6,277
Geddes, L. A., & Baker, L. E. (1967) · Medical & Biological Engineering · doi:10.1007/bf02474537 · Cited by 1,610
(2008) · Springer eBooks · doi:10.1007/978-3-540-69972-9 · Cited by 1,213
Quantitative computed tomography
Adams, J. E. (2009) · European Journal of Radiology · doi:10.1016/j.ejrad.2009.04.074 · Cited by 458 · cited 1×
“-ere are various quantitative techniques for evaluating bone fractures healing process, such as ultrasound [11], bending stiffness test [12], measurement of bone mineral density [13], quantitative computed tomography (QCT) [14], acoustic emission [15], and vibratory devices [16].”
BackgroundVariability in the Assessment of Fracture-Healing in Orthopaedic Trauma Studies
Corrales, L. A., Morshed, S., Bhandari, M., & Miclau, T. (2008) · The Journal of Bone and Joint Surgery-American Volume · doi:10.2106/jbjs.g.01580 · Cited by 415
Litrenta, J., Tornetta, P., Mehta, S., Jones, C., O'Toole, R. V., Bhandari, M., Kottmeier, S., Ostrum, R., Egol, K., Ricci, W., et al. (2015) · Journal of Orthopaedic Trauma · doi:10.1097/bot.0000000000000390 · Cited by 287 · cited 1×
“-e union of bone after the osteotomy is scored by RUST [19, 24].”
MethodBhandari, M., Fong, K., Sprague, S., Williams, D., & Petrisor, B. (2012) · Journal of Bone and Joint Surgery · doi:10.2106/jbjs.k.01344 · Cited by 179 · cited 1×
“Physician examinations are subjective, which might result in varying and erroneous judgments [3].”
BackgroundCurrent concepts of leg lengthening
Hasler, C. C., & Krieg, A. H. (2012) · Journal of Children s Orthopaedics · doi:10.1007/s11832-012-0391-5 · Cited by 143
Effect of Fatiguing Exercise on Longitudinal Bone Strain as Related to Stress Fracture in Humans
Fyhrie, D. P., Milgrom, C., Hoshaw, S. J., Simkin, A., Dar, S., Drumb, D., & Burr, D. B. (1998) · Annals of Biomedical Engineering · doi:10.1114/1.103 · Cited by 110 · cited 1×
“As a result, some researchers focused on mechanical bone feedback [17, 18], while others used electrical techniques to evaluate fractures [19].”
Ultrasonic monitoring of bone fracture healing
Protopappas, V. C., Vavva, M. G., Fotiadis, D. I., & Malizos, Κ. Ν. (2008) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2008.787 · Cited by 84 · cited 1×
“Computed Tomography (CT), Dual Energy X-ray Absorptiometry (DXA/DEXA), and ultrasound can all be used to diagnose bone fractures, but their use in medical centers is limited due to their high cost and radiation exposure [7, 10].”
BackgroundFracture following lower limb lengthening in children: A series of 58 patients
Launay, F., Younsi, R., Pithioux, M., Chabrand, P. R., Bollini, G., & Jouve, J.-L. (2012) · Orthopaedics & Traumatology Surgery & Research · doi:10.1016/j.otsr.2012.08.005 · Cited by 64 · cited 1×
“6 percent when Osteodensitometry was used [34].”
Smart bone plates can monitor fracture healing
Lin, M. C., Hu, D. P., Marmor, M. T., Herfat, S. T., Bahney, C. S., & Maharbiz, M. M. (2019) · Scientific Reports · doi:10.1038/s41598-018-37784-0 · Cited by 60 · cited 5×
“-e union of bone after the osteotomy is scored by RUST [19, 24].”
BackgroundMethodInfluentialIn vivo measurement of bending stiffness in fracture healing
HENTE, R., Cordey, J., & Perren, S. M. (2003) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-2-8 · Cited by 57 · cited 1×
“-ere are various quantitative techniques for evaluating bone fractures healing process, such as ultrasound [11], bending stiffness test [12], measurement of bone mineral density [13], quantitative computed tomography (QCT) [14], acoustic emission [15], and vibratory devices [16].”
BackgroundNatu, S. S., Ali, I. M., Alam, S., Giri, K. Y., Agarwal, A., & Kulkarni, V. A. (2014) · PubMed Central · doi:10.4103/1735-3327.127179 · Cited by 57
Borchani, W., Aono, K., Lajnef, N., & Chakrabartty, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2496237 · Cited by 56 · cited 1×
“As a result, some researchers focused on mechanical bone feedback [17, 18], while others used electrical techniques to evaluate fractures [19].”
DEXA as a Predictor of Fixator Removal in Distraction Osteogenesis
Saran, N., & Hamdy, R. C. (2008) · Clinical Orthopaedics and Related Research · doi:10.1007/s11999-008-0514-y · Cited by 52
Zhao, L., Fan, Q., Venkatesh, K. P., Park, M. S., & Song, H. R. (2009) · Clinical Orthopaedics and Related Research · doi:10.1007/s11999-009-1011-7 · Cited by 46 · cited 1×
“used radiography to determine the presence of a full cortex at the lengthened site [36].”
MethodDjasim, U. M., Wolvius, E. B., van Neck, J. W., Weinans, H., & van der Wal, K. G. H. (2007) · International Journal of Oral and Maxillofacial Surgery · doi:10.1016/j.ijom.2007.05.012 · Cited by 45 · cited 2×
“Aside frommovement limitations, the device’s prolonged presence on the bone may have a negative social and psychological impact on patients [35].”
BackgroundVannabouathong, C., Sprague, S., & Bhandari, M. (2010) · Injury · doi:10.1016/j.injury.2010.11.048 · Cited by 44
Evaluation of Strength of Healing Fractures With Dual Energy Xray Absorptiometry
Blokhuis, T. J., den Boer, F. C., Bramer, J. A. M., van Lingen, A., Roos, J. C., Bakker, F. C., Patka, P., & Haarman, H. J. T. M. (2000) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-200011000-00035 · Cited by 30 · cited 1×
“-ere are various quantitative techniques for evaluating bone fractures healing process, such as ultrasound [11], bending stiffness test [12], measurement of bone mineral density [13], quantitative computed tomography (QCT) [14], acoustic emission [15], and vibratory devices [16].”
BackgroundMeasurement of bone electrical impedance in fracture healing
Yoshida, T., Kim, W., Kawamoto, K., Hirashima, T., Oka, Y., & Kubo, T. (2009) · Journal of Orthopaedic Science · doi:10.1007/s00776-009-1333-z · Cited by 27 · cited 2×
“Several studies in this field using diverse experiments such as histology, radiography, and micro CT revealed a direct association between bone healing and impedance [21, 26, 32].”
BackgroundEvaluation of bone fracture in animal model using bio-electrical impedance analysis
Khan, M., Sirdeshmukh, S. P. S. M. A., & Javed, K. (2016) · Perspectives in Science · doi:10.1016/j.pisc.2016.06.022 · Cited by 25 · cited 1×
“-ese tests are extremely harmful to children, pregnant women, and the elderly, not to mention quite expensive [6, 8, 9].”
BackgroundLin, M. C., Yang, F. F., Herfat, S. T., Bahney, C. S., Marmor, M. T., & Maharbiz, M. M. (2017) · Journal of Orthopaedic Research® · doi:10.1002/jor.23570 · Cited by 20 · cited 1×
“As a result, impedance values in tissues with less water, such as bone, are greater than impedance values in tissues with a lot of water, such as blood and muscle [29, 30].”
BackgroundEvaluating Bone Union of Distal Radius Fractures by Measuring Impedance Values
Hirashima, T., Kim, W., Kawamoto, K., Yoshida, T., & Kubo, T. (2009) · Orthopedics · doi:10.3928/01477447-20090101-10 · Cited by 19 · cited 3×
“-ese tests are extremely harmful to children, pregnant women, and the elderly, not to mention quite expensive [6, 8, 9].”
BackgroundAssessment of distraction callus in rabbits by monitoring of the electrical impedance of bone
Yoshida, T., Kim, W., Oka, Y., Yamada, N., & Kubo, T. (2010) · Acta Orthopaedica · doi:10.3109/17453674.2010.519910 · Cited by 15 · cited 6×
“Traditionally, orthopedic surgeons determined the time of fixation device removal from bone by examining the patient and taking radiographs [21].”
BackgroundMethodInfluentialEvaluating fracture healing using digital X-ray image analysis
Whiley, S. P. (2011) · CME/Continuing medical education · Cited by 13
Electrical System for Bioelectric Impedance using AD5933 Impedance Converter
Garcia, A., & Sabuncu, A. C. (2019) · SMU Journal of Undergraduate Research · doi:10.25172/jour.4.1.3 · Cited by 13 · cited 5×
“As a result, the AD5933 evaluation board was employed to evaluate impedance in this study.”
BackgroundInfluentialQuantitative monitoring of bone healing process using ultrasound
Hijazy, A., Smoudi, H. A. –., Swedan, M., Qaddoum, N., Al-Nashash, H. A., & Ramesh, K. G. (2006) · Journal of the Franklin Institute · doi:10.1016/j.jfranklin.2006.03.014 · Cited by 12 · cited 1×
“-ere are various quantitative techniques for evaluating bone fractures healing process, such as ultrasound [11], bending stiffness test [12], measurement of bone mineral density [13], quantitative computed tomography (QCT) [14], acoustic emission [15], and vibratory devices [16].”
BackgroundImpedance spectroscopy to monitor fracture healing
Lin, M. C., Herfat, S. T., Bahney, C. S., Marmor, M. T., & Maharbiz, M. M. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7319548 · Cited by 11 · cited 1×
“As a result, impedance values in tissues with less water, such as bone, are greater than impedance values in tissues with a lot of water, such as blood and muscle [29, 30].”
BackgroundBioimpedance and bone fracture detection: A state of the art
Dell'Osa, A., Felice, C. J., & Símini, F. (2019) · Journal of Physics Conference Series · doi:10.1088/1742-6596/1272/1/012010 · Cited by 11 · cited 1×
“Several studies in this field using diverse experiments such as histology, radiography, and micro CT revealed a direct association between bone healing and impedance [21, 26, 32].”
BackgroundMonitoring of bone healing by piezoelectric-EMI method
Mazlina, M., Naimullah, B. S. S., Tawie, R., Daho, C. D., & Annuar, I. (2016) · AIP conference proceedings · doi:10.1063/1.4940256 · Cited by 9 · cited 8×
“As a result, the AD5933 evaluation board was employed to evaluate impedance in this study.”
BackgroundInfluentialUltrasound for Detection of Fracture Healing
Maffulli, N. (1999) · Journal of Orthopaedic Trauma · doi:10.1097/00005131-199906000-00029 · Cited by 8 · cited 1×
“-ere are various quantitative techniques for evaluating bone fractures healing process, such as ultrasound [11], bending stiffness test [12], measurement of bone mineral density [13], quantitative computed tomography (QCT) [14], acoustic emission [15], and vibratory devices [16].”
BackgroundLow-cost impedance approach using AD5933 for sensing and monitoring applications
Tawie, R., & Da’ud, E. (2018) · IOP Conference Series Materials Science and Engineering · doi:10.1088/1757-899x/429/1/012104 · Cited by 8 · cited 2×
“However, because it is difficult to determine when to remove the fixator from the bone [28], decisions based on radiography and clinical examination have poor interobserver and intraobserver reliability.”
Background
Cited by
Methods to accelerate fracture healing – a narrative review from a clinical perspective
Ganse, B. (2024) · Frontiers in Immunology · doi:10.3389/fimmu.2024.1384783 · Cited by 33
Mechanoresponsive regulation of tissue regeneration during distraction osteogenesis
Hu, W., Guo, Z., Tang, W., & Long, J. (2024) · The FASEB Journal · doi:10.1096/fj.202401303r · Cited by 25
“115 Indeed, the consolidation phase is the most time-consuming period, with an average estimate suggesting that 1 month is required for every 1 cm of extension.”
Sensor‐enabled Orthopedic Implants for Musculoskeletal Monitoring
Huang, J., Dong, H., Zhang, W., Cui, J., Li, Q., Wang, J., Zhao, Z., & Zang, X. (2025) · Advanced Sensor Research · doi:10.1002/adsr.202400138 · Cited by 24
Heidari, N., Vosough, M., Bagherifard, A., Sami, S. H., Sarabi, P. A., Behmanesh, A., & Shams, R. (2024) · Pathology - Research and Practice · doi:10.1016/j.prp.2024.155615 · Cited by 3
Self-citationFrost, M. W., Tirta, M., Rahbek, O., Rytoft, L. A., Ding, M. L., Shen, M., Duch, K. S., & Kold, S. (2024) · Journal of Experimental Orthopaedics · doi:10.1002/jeo2.12048 · Cited by 2
“Some studies placed the electrodes within the fracture site [13], which is clinically unsound, while others used the metal pins from the external fi xator as electrodes giving way to noise from surrounding tissues [12, 14, 21, 25 – 27].”
Metadata
- Authors
- Sadoughi, F., Behmanesh, A., Najd Mazhar, F., Joghataei, M. T., Yazdani, S., Shams, R., Morovvati, H., Najaf Asaadi, S., & Vosough, A.
- Affiliations
- Iran University of Medical Sciences (IR)
- Bone and Joint Reconstruction Research Center, Department of Orthopedics, School of Medicine, Iran University of Medical Sciences, Tehran, Iran
- Shahid Beheshti University of Medical Sciences (IR)
- University of Tehran (IR)
- Department of Basic Science, Faculty of Veterinary Medicine, University of Tehran, Tehran, Iran
- Islamic Azad University of Garmsar (IR)
- Corresponding author
- Behmanesh, Ali
- ORCID
- Venue
- Journal of Healthcare Engineering
- Publisher
- Wiley
- Published
- 2022-04-07
- Volume
- 2022
- Pages
- 1-13
- ISSN
- 2040-2295 (ISSN-L)
- 2040-2309
- PMC
- PMC9010194
- OpenAlex
- W4226197551
- Cited by
- 10
- References
- 35
- Citations by year
- 2026: 3 · 2025: 2 · 2024: 4 · 2022: 1
- FWCI
- 1.26
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- Journal of Healthcare Engineering (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
- Funders
- Iran University of Medical Sciences (IUMS/SHMIS-97-3-37-12579)
- Topics
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Bone fractures and treatments
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Bone health and osteoporosis research
- Physical Sciences › Engineering › Biomedical Engineering › Bone Tissue Engineering Materials
- Keywords
- bone healing monitoring, distraction osteogenesis, bioimpedance, tibial osteotomy, external fixator, mini external fixator, rabbit model
- MeSH terms
- Animals, Humans, Male, Osteotomy, Rabbits, Tibia, External Fixators, Fracture Healing, Osteogenesis, Distraction
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.