CoChat

Pressure Recovery in Bileaflet Heart Valve Prostheses: Localized High Velocities and Gradients in Central and Side Orifices With Implications for Doppler-Catheter Gradient Relation in Aortic and Mitral Position

Vandervoort, Pieter M., Greenberg, Neil L., Pu, Min, Powell, Kimerly A., Cosgrove, Delos M., Thomas, James D. · 1995 · Circulation
Cited by 106

The DOI matches its Crossref record.

Abstract

Background We investigate pressure recovery in central and side orifices of St Jude valves and the effect of mitral versus aortic position on the relation between Doppler- and catheter-derived pressure gradients. Methods and Results Maximum, transvalvular, and net pressure gradients are calculated and compared with Doppler-derived gradients in an in vitro model. Pressure recovery and pressure loss coefficients are calculated. Simultaneous Doppler and catheter gradients are obtained intraoperatively in five patients undergoing mitral valve replacement. Centerline Doppler gradients correspond closely with maximum catheter gradients but are higher than transvalvular and net pressure gradients. Thirty-six percent of the initial pressure drop is recovered between the valve leaflets and is independent of valve size or configuration. A variable amount of postvalvular pressure recovery is observed depending on aortic or mitral configuration. Side orifice velocities are 85±4% of the centerline velocities. Incorporation of the pressure loss coefficient in the simplified Bernoulli equation shows close agreement between centerline Doppler gradients and transvalvular gradients (r=.99,y=1.11 x−0.19). Conclusions Gradients across the St Jude valve measured by Doppler ultrasound are higher than transvalvular or net catheter gradients due to downstream pressure recovery. This is more marked for Doppler gradients based on centerline velocities than side orifice velocities and is more pronounced for valves in an aortic than a mitral configuration. Therefore, to be comparable with invasive transvalvular catheter gradients, either Doppler gradients should be calculated based on side orifice velocity measurements or the Doppler gradient calculation should include the pressure loss coefficient when based on central orifice velocities.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1161/01.cir.92.12.3464

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 106

Metadata

Authors
Vandervoort, P. M., Greenberg, N. L., Pu, M., Powell, K. A., Cosgrove, D. M., & Thomas, J. D.
Venue
Circulation
Published
1995-12-15
Volume
92
Issue
12
Pages
3464-3472
ISSN
  • 0009-7322 (ISSN-L)
  • 1524-4539
OpenAlex
W2000697964
Cited by
106
References
0
FWCI
3.54
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Pressure Recovery in Bileaflet Heart Valve Prostheses: Localized High Velocities and Gradients in Central and Side Orifices With Implications for Doppler-Catheter Gradient Relation in Aortic and Mitral Position. https://app.cochat.ai/papers/10.1161/01.cir.92.12.3464
The paper
Vandervoort, P. M., Greenberg, N. L., Pu, M., Powell, K. A., Cosgrove, D. M., & Thomas, J. D. (1995). Pressure Recovery in Bileaflet Heart Valve Prostheses: Localized High Velocities and Gradients in Central and Side Orifices With Implications for Doppler-Catheter Gradient Relation in Aortic and Mitral Position. Circulation. https://doi.org/10.1161/01.cir.92.12.3464

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.