A Novel Method for Nonfluoroscopic Catheter-Based Electroanatomical Mapping of the Heart: In Vitro and In Vivo Accuracy Results

Gepstein, Lior, Hayam, Gal, Ben-Haim, Shlomo A.·1997·Circulation

Abstract

Background Cardiac mapping is essential for understanding the mechanisms of arrhythmias and for directing curative procedures. A major limitation of the current methods is the inability to accurately relate local electrograms to their spatial orientation. The objective of this study was to present and test the accuracy of a new method for nonfluoroscopic, catheter-based, endocardial mapping. Methods and Results The method is based on using a new locatable catheter connected to an endocardial mapping and navigating system. The system uses magnetic technology to accurately determine the location and orientation of the catheter and simultaneously records the local electrogram from its tip. By sampling a plurality of endocardial sites, the system reconstructs the three-dimensional geometry of the chamber, with the electrophysiological information color-coded and superimposed on the anatomy. The accuracy of the system was tested in both in vitro and in vivo studies and was found to be highly reproducible (SD, 0.16±0.02 [mean±SEM] and 0.74±0.13 mm) and accurate (mean errors, 0.42±0.05 and 0.73±0.03 mm). In further studies, electroanatomical mapping of the cardiac chambers was performed in 34 pigs. Both the geometry and activation sequence were repeatable in all pigs. Conclusions The new mapping method is highly accurate and reproducible. The ability to combine electrophysiological and spatial information provides a unique tool for both research and clinical electrophysiology. Consequently, the main shortcomings of conventional mapping—namely, prolonged x-ray exposure, low spatial resolution, and the inability to accurately navigate to a predefined site—can all be overcome with this new method.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1161/01.cir.95.6.1611

Cites

11 of 11 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 11 references

Cited by

See all 759

Metadata

Authors
Gepstein, L., Hayam, G., & Ben-Haim, S. A.
Venue
Circulation
Published
1997-03-18
Volume
95
Issue
6
Pages
1611-1622
ISSN
  • 0009-7322 (ISSN-L)
  • 1524-4539
OpenAlex
W2063884817
Cited by
759
References
11
FWCI
24.64
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: A Novel Method for Nonfluoroscopic Catheter-Based Electroanatomical Mapping of the Heart: In Vitro and In Vivo Accuracy Results. https://app.cochat.ai/papers/10.1161/01.cir.95.6.1611
The paper
Gepstein, L., Hayam, G., & Ben-Haim, S. A. (1997). A Novel Method for Nonfluoroscopic Catheter-Based Electroanatomical Mapping of the Heart: In Vitro and In Vivo Accuracy Results. Circulation. https://doi.org/10.1161/01.cir.95.6.1611

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.