Accurate Assessment of Patient Effective Radiation Dose and Associated Detriment Risk From Radiofrequency Catheter Ablation Procedures

Open access
Perisinakis, Kostas, Damilakis, John, Theocharopoulos, Nicholas, Manios, Emmanouel, Vardas, Panayiotis, Gourtsoyiannis, Nicholas · 2001 · Circulation

Abstract

Background— Radiofrequency (RF) cardiac catheter ablation procedures may require extended fluoroscopic exposure resulting in elevated radiation risk. The aim of the present study was to accurately establish RF ablation radiation risk levels and to provide means for accurate patient risk estimation from studies performed in any electrophysiology laboratory. Methods and Results— Fluoroscopy required during cardiac ablation was classified into 4 types identified by beam orientation and irradiated tissue: (1) posteroanterior exposure during catheter advancing from the groin to the heart, (2) posteroanterior heart exposure, (3) left anterior oblique heart exposure, and (4) right anterior oblique heart exposure. The duration of each exposure was monitored in 24 patients undergoing RF cardiac ablation. Dose per minute of fluoroscopy was measured at 15 organs/tissues for each projection with the use of anthropomorphic phantom and thermoluminescence dosimetry. The effective dose rate was 219, 144, 136, and 112 μGy/min for groin-to-heart posteroanterior, posteroanterior, left anterior oblique, and right anterior oblique exposure, respectively. A typical ablation procedure results in a total effective dose of 8.3 mSv per hour of fluoroscopy. The average excess of fatal cancers was estimated to be 650 and 480 per million patients undergoing RF ablation requiring 1 hour of fluoroscopy for US and UK populations, respectively. The average risk for genetic defects was determined to be 1 per million births. Conclusions— Radiation risk from RF cardiac ablation is moderate compared with other complications, but it may highly exceed radiation risk from common radiological procedures. Efforts should be made toward minimization of patient radiation risk from RF ablation procedures.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1161/hc2601.091710

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 1 references

Cited by

See all 176

Metadata

Authors
Perisinakis, K., Damilakis, J., Theocharopoulos, N., Manios, E., Vardas, P., & Gourtsoyiannis, N.
Venue
Circulation
Published
2001-07-03
Volume
104
Issue
1
Pages
58-62
ISSN
  • 0009-7322 (ISSN-L)
  • 1524-4539
OpenAlex
W2019306804
Cited by
176
References
1
FWCI
2.52
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Accurate Assessment of Patient Effective Radiation Dose and Associated Detriment Risk From Radiofrequency Catheter Ablation Procedures. https://app.cochat.ai/papers/10.1161/hc2601.091710
The paper
Perisinakis, K., Damilakis, J., Theocharopoulos, N., Manios, E., Vardas, P., & Gourtsoyiannis, N. (2001). Accurate Assessment of Patient Effective Radiation Dose and Associated Detriment Risk From Radiofrequency Catheter Ablation Procedures. Circulation. https://doi.org/10.1161/hc2601.091710

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.