Fluid Dynamics of a Novel Micro-Fistula Implant for the Surgical Treatment of Glaucoma
The DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.
Abstract
PURPOSE: The purpose of this study was to describe the fluidics of a novel non-valved glaucoma implant designed to prevent hypotony and compare the fluidics of this device with two commonly used non-valved glaucoma devices. METHODS: The XEN 45 micro-fistula implant was designed to limit hypotony by virtue of its length and width according to the Hagen-Poiseuille equation. Flow testing was performed using a syringe pump and pressure transducer at multiple flow rates. The pressure differentials across the XEN implant, the Ex-Press implant, and 10 mm of silicone tubing from a Baerveldt implant at a physiologic flow rate (2.5 μL/min) were extrapolated. RESULTS: The XEN 45 achieved a steady-state pressure calculated at 7.56 mm Hg at 2.5 μL/min. At the same flow rate, the Ex-Press device and Baerveldt tubing reached steady-state pressures of 0.09 and 0.01 mm Hg, respectively. CONCLUSIONS: Under flow testing, the XEN micro-fistula implant was able to maintain backpressure above numerical hypotony levels without the use of complex valve systems. This is due to the XEN implant's design, derived from the principles that dictate Newtonian fluids.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Global Prevalence of Glaucoma and Projections of Glaucoma Burden through 2040
Tham, Y.-C., Li, X., Wong, T. Y., Quigley, H. A., Aung, T., & Cheng, C.-Y. (2014) · Ophthalmology · doi:10.1016/j.ophtha.2014.05.013 · Cited by 7,159
Treatment Outcomes in the Tube Versus Trabeculectomy (TVT) Study After Five Years of Follow-up
Gedde, S. J., Schiffman, J. C., Feuer, W. J., Herndon, L. W., Brandt, J. D., & Budenz, D. L. (2012) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2011.10.026 · Cited by 1,110
The Ahmed Versus Baerveldt Study
Christakis, P. G., Kalenak, J. W., Zurakowski, D., Tsai, J. C., Kammer, J. A., Harasymowycz, P. J., & Ahmed, I. I. K. (2011) · Ophthalmology · doi:10.1016/j.ophtha.2011.05.004 · Cited by 379
Self-citation3D simulation of the aqueous flow in the human eye
Villamarin, A., Roy, S., Hasballa, R., Vardoulis, O., Reymond, P., & Stergiopulos, N. (2012) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2012.02.007 · Cited by 85
A New Adjustable Glaucoma Drainage Device
Villamarin, A., Roy, S., Bigler, S., & Stergiopulos, N. (2014) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.13-12626 · Cited by 33
Cited by
Performance and Safety of a New Ab Interno Gelatin Stent in Refractory Glaucoma at 12 Months
Grover, D. S., Flynn, W. J., Bashford, K. P., Lewis, R. A., Duh, Y.-J., Nangia, R. S., & Niksch, B. (2017) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2017.07.023 · Cited by 280
XEN Glaucoma Implant with Mitomycin C 1-Year Follow-Up: Result and Complications
Galal, A., Bilgic, A., Eltanamly, R., & Osman, A. (2017) · Journal of Ophthalmology · doi:10.1155/2017/5457246 · Cited by 174
Metadata
- Authors
- Sheybani, A., Reitsamer, H., & Ahmed, I. I. K.
- Venue
- Investigative Opthalmology & Visual Science
- Published
- 2015-07-25
- Volume
- 56
- Issue
- 8
- Pages
- 4789
- ISSN
- 0146-0404 (ISSN-L)
- 1552-5783
- OpenAlex
- W1572642458
- Cited by
- 196
- References
- 5
- FWCI
- 13.32
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Fluid Dynamics of a Novel Micro-Fistula Implant for the Surgical Treatment of Glaucoma. https://app.cochat.ai/papers/10.1167/iovs.15-16625
- The paper
- Sheybani, A., Reitsamer, H., & Ahmed, I. I. K. (2015). Fluid Dynamics of a Novel Micro-Fistula Implant for the Surgical Treatment of Glaucoma. Investigative Opthalmology & Visual Science. https://doi.org/10.1167/iovs.15-16625
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.