Which Monitor? An Analysis of 2000 Incident Reports

Open access
Webb, R. K., Van Der Walt, J. H., Runciman, W. B., Williamson, J. A., Cockings, J., Russell, W. J., Helps, S. · 1993 · Anaesthesia and Intensive Care

Abstract

The role of monitors in patients undergoing general anaesthesia was studied by analysing the first 2000 incidents reported to the Australian Incident Monitoring Study; 1256 (63%) were considered applicable to this study. In 52% of these a monitor detected the incident first; oximetry (27%) and capnography (24%) detected over half of the monitor detected incidents, the electrocardiograph 19%, blood pressure monitors 12%, a low pressure (circuit) alarm 8%, and the oxygen analyser 4%. Of the other monitors used, 5 first detected 1–2% of incidents, and the remaining 8 less than 0.5% each. The oximeter would have detected over 40% of the monitor detected incidents had its more informative modulated pulse tone always been relied upon instead of the “bleep” of the ECG. A theoretical analysis was then carried out to determine which of an array of 17 monitors would reliably have detected each incident had each monitor been used on its own and had the incident been allowed to evolve. To facilitate “scoring” of monitors, the incidents were categorized empirically into 60 clinical situations; 40% of applicable incidents were accounted for by only 5 clinical situations, 60% by 10 and nearly 80% by 20. 98% were accounted for by the 60 situations. A pulse oximeter, used on its own, would theoretically have detected 82% of applicable incidents (nearly 60% before any potential for organ damage). These figures for capnography are 55% and 43 % and for oximetry and capnography combined are 88% and 65%, respectively. With the addition of blood pressure monitoring these become 93% and 65%, and of an oxygen analyser, 95 and 67%. Other monitors, including the ECG, each increase the yield by by less than 0.5%. The international monitoring recommendations and those of the Australian and New Zealand College of Anaesthetists are thoroughly vindicated by the patterns revealed in this study. The priority sequence of monitor acquisition for those with limited resources should be stethoscope, sphygmomanometer, oxygen analyser if nitrous oxide is to be used, pulse oximeter, capnograph, high pressure alarm, and, if patients are to be mechanically ventilated, a low pressure alarm (or spirometer with alarm); an ECG, a defibrillator, a spirometer and a thermometer should be available.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1177/0310057x9302100508

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 193

Metadata

Authors
Webb, R. K., Van Der Walt, J. H., Runciman, W. B., Williamson, J. A., Cockings, J., Russell, W. J., & Helps, S.
Venue
Anaesthesia and Intensive Care
Published
1993-10
Volume
21
Issue
5
Pages
529-542
ISSN
  • 0310-057X (ISSN-L)
  • 1448-0271
OpenAlex
W9697418
Cited by
193
References
0
FWCI
12.86
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Which Monitor? An Analysis of 2000 Incident Reports. https://app.cochat.ai/papers/10.1177/0310057x9302100508
The paper
Webb, R. K., Van Der Walt, J. H., Runciman, W. B., Williamson, J. A., Cockings, J., Russell, W. J., & Helps, S. (1993). Which Monitor? An Analysis of 2000 Incident Reports. Anaesthesia and Intensive Care. https://doi.org/10.1177/0310057x9302100508

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.