Patient safety and interactive medical devices: Realigning work as imagined and work as done
Open accessThe DOI matches its Crossref record; 13 of the 13 references it cites were found in a registry or an index.
Abstract
Medical devices are essential tools for modern healthcare delivery. However, significant issues can arise if medical devices are designed for ‘work as imagined’ when this is misaligned with ‘work as done’. This problem can be compounded as the details of device design, in terms of usability and the way a device supports or changes working practices, often receives limited attention. The ways devices are designed and used affect patient safety and quality of care: inappropriate design can provoke user error, create system vulnerabilities and divert attention from other aspects of patient care. Current regulation involves a series of pre-market checks relating to device usability, but this assumes that devices are always used under the conditions and for the purposes intended (i.e. work as imagined); there are many reasons for devices being used in ways other than those assumed at development time. Greater attention needs to be paid to learning points in actual use and user experience (i.e. work as done). This needs to inform manufacturers’ designs, management procurement decisions and local decisions about how devices are used in practice to achieve co-adaptation; without these, we foster risks and inefficiencies in healthcare.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“There are well documented cases where patient safety has been compromised by poorly designed user interfaces and systems (e.g., Leveson and Turner, 1993;Cook et al, 2008;Blandford et al, 2014).”
From HCI for health and wellbeing: Challenges and opportunities (2019) · cites this paper as Blandford et al, 2014
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Koppel, R., Wetterneck, T., Telles, J. L., & Karsh, B.-T. (2008) · Journal of the American Medical Informatics Association · doi:10.1197/jamia.m2616 · Cited by 685 · cited 2×
“For example, Koppel et al.(14) identified various workarounds in the use of a barcode medication administration system, such as placing barcodes on objects in the environment rather than attaching them to the patient, while Furniss et al.”
BackgroundUnderstanding infusion administration in the ICU through Distributed Cognition
Rajkomar, A., & BLANDFORD, A. E. (2012) · Journal of Biomedical Informatics · doi:10.1016/j.jbi.2012.02.003 · Cited by 76
Self-citationMedical devices. Application of usability engineering to medical devices
(2008) · BSI Standards · doi:10.3403/30112973 · Cited by 65
Medical device and equipment design : usability engineering and ergonomics
Wiklund, M. (1995) · Medical Entomology and Zoology · Cited by 48
Furniss, D. J., Masci, P. M., Curzon, P., Mayer, A., & BLANDFORD, A. E. (2014) · Applied Ergonomics · doi:10.1016/j.apergo.2014.05.012 · Cited by 43
Self-citationHazards With Medical Devices: The Role of Design
Fairbanks, R. J., & Wears, R. L. (2008) · Annals of Emergency Medicine · doi:10.1016/j.annemergmed.2008.07.008 · Cited by 33 · cited 1×
“Many interactive medical devices pose usability challenges, and such difficulties can contribute to incidents involving serious harm.(5) For example, Denise Melanson died following an incident in which an infusion pump was programmed to deliver over 4 h medication that should have been delivered over four days.”
BackgroundUnremarkable errors: low-level disturbances in infusion pump use
Furniss, D. J., BLANDFORD, A. E., & Mayer, A. (2011) · Electronic workshops in computing · doi:10.14236/ewic/hci2011.47 · Cited by 26 · cited 3×
“Another kind of difficulty emerges in the ‘‘unremarkable errors’’ observed by Furniss et al.(12): these are little, every-day errors that clinicians barely regard as errors because they are minor and are quickly recovered from.”
BackgroundSelf-citationBetween Choice and Chance: The Role of Human Factors in Acute Care Equipment Decisions
Nemeth, C., Nunnally, M., Bitan, Y., Nunnally, S., & Cook, R. I. (2009) · Journal of Patient Safety · doi:10.1097/pts.0b013e3181a974d9 · Cited by 20 · cited 1×
“Good human factors can save money(22) as well as improving safety.(23) However, the story does not end with purchasing.”
BackgroundThe Mixed Blessings of Smart Infusion Devices and Health Care IT
Nemeth, C., Brown, J., Crandall, B. W., & Fallon, C. K. (2014) · Military Medicine · doi:10.7205/milmed-d-13-00505 · Cited by 19 · cited 1×
“increasing the risk of patient misidentification or confusability of devices, or circumventing safety checks that were designed to catch errors).(18) As Debono et al.”
BackgroundGandillon, R. (2013) · Journal of Clinical Engineering · doi:10.1097/jce.0b013e31827c0fba · Cited by 7
Research works printed but not linked: 3
- Francis R. Report of the Mid Staffordshire NHS foundation trust public inquiry: executive summary. Vol. 947. London: The Stationery Office, 2013.
- AAMI. HE75: Human factors engineering – design of medical devices, Arlington, VA: AAMI, 2009.
- Debono D, Greenfield D, Black D, et al. Workarounds: straddling or widening gaps in the safe delivery of healthcare. In Proceedings of the 7th international conference in Organisational Behaviour in Health Care (OBHC), Birmingham, UK, 11–14 April 2010.
Cited by
Items 101–119 of 119
Groenier, M., Spijkerboer, K., Venix, L. P. J., Bannink, L., Yperlaan, S., Eyck, Q., van Manen, J. G., & Miedema, H. A. T. (2023) · Figshare · doi:10.6084/m9.figshare.22620490
Zani, C. M. (2021) · LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas)
Owen, T., Jones, M., & Delahaye, B. (2025) · Cronfa (Swansea University)
Jokisalo, E. (2024) · Theseus (Ammattikorkeakoulujen)
Debriefing modes of simulation sessions as organizational probes of safety management
Boulard, C., & Falzon, P. (2024) · HAL
Sanders, R. (2021) · UvA-DARE (University of Amsterdam)
Qing, Opie, J., Brookshaw, T., Erro, E., BLANDFORD, A. E., & Selden, C. (2026) · JMIR Human Factors · doi:10.2196/82405
“The think-aloud study had participants read and react to both the IFU document and videos of the IFU procedures carried out by a developer, capturing participants’ perceptions of “work-as-imag-ined” (WAI) based on simply reading the IFU [36,37].”
Self-citationMiller, A. W. (2025) · TSpace (University of Toronto)
Pawar, A., Dimri, U., Bhowmik, S., Jagani, R., Sharma, A., Reddy, R. M., Ninawe, S. M., & Maaheraa, L. A. (2026) · Medical Journal Armed Forces India · doi:10.1016/j.mjafi.2026.01.017
Mühlenbeck, B., Opie, J., Fernandez, C. S., Jaufuraully, S., Parris, D., Desjardins, A. E., David, A. L., Siassakos, D. M., Tiwari, M. K., & BLANDFORD, A. E. (2026) · JMIR Formative Research · doi:10.2196/78845
“The application of a novel sensor when dealing with instruments prone to cause harm to the patient was also identified as an opportunity, as previous research has indicated that the design of interactive medical devices can provoke user error [40].”
Self-citationGalinsky, E. K., Dangler, M., VanMiddendorp, E., Clark, J., Konwinski, L. M., Meilan, J., & Catchpole, K. (2026) · Journal of Patient Safety · doi:10.1097/pts.0000000000001489
Nacu, A., Constantin, D.-A., & Rogozea, L. M. (2026) · Healthcare · doi:10.3390/healthcare14111542
“The findings align with resilient healthcare scholarship showing that safety is sustained through adaptive work at the boundary between formal procedure and practical contingency [3,4].”
Fukami, T. (2026) · Journal of Patient Safety and Risk Management · doi:10.1177/25160435261458499
Machado, S. M., & Fermam, R. K. S. (2026) · Revista Sustinere · doi:10.12957/sustinere.2026.88221
Wilson, B., Roach, M., Browning, G., Connor, C., & Rawlinson, D. (2026) · arXiv (Cornell University) · doi:10.48550/arxiv.2609.07705
การพัฒนาอุปกรณ์ยึดสายนำเลือดชนิดปรับขนาดได้สำหรับการผ่าตัดหัวใจ
Jiraratkul, S., & Pichaiphanupatt, P. (2022) · Nursing Research and Innovation Journal
Laohavich, C., & Laungdamrongchai, S. (2020) · Nursing Research and Innovation Journal
การพัฒนาคลิปล็อคอุปกรณ์สำาหรับรวบถุงคลุมกล้องผ่าตัดปลอดเื้อและการประเมินประสิทธิผล
Pongsukwetchakul, T., & Chiannilkulchai, N. (2018) · Nursing Research and Innovation Journal
การพัฒนาอุปกรณ์ถ่างขยายลิ้นหัวใจเอออร์ติกแบบเปิดแผลเล็ก
Pichaiphanupatt, P. (2021) · Nursing Research and Innovation Journal
Metadata
- Authors
- Blandford, A., Furniss, D., & Vincent, C.
- Affiliations
- University College London (GB)
- Corresponding author
- Blandford, Ann
- Venue
- Clinical Risk
- Publisher
- SAGE Publications
- Published
- 2014-09
- Volume
- 20
- Issue
- 5
- Pages
- 107-110
- ISSN
- 1356-2622 (ISSN-L)
- 1758-1028
- PMC
- PMC4361486
- OpenAlex
- W2116903837
- Cited by
- 83
- References
- 13
- Citations by year
- 2026: 7 · 2025: 7 · 2024: 7 · 2023: 6 · 2022: 5 · 2021: 9 · 2020: 5 · 2019: 14 · 2018: 10 · 2017: 4 · 2016: 5 · 2015: 2 · 2014: 1
- FWCI
- 3.97
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by-nc
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Clinical Risk (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- UCL Discovery (University College London) (submittedVersion)
- Funders
- Medical Research Council (MR/K006584/1)
- Engineering and Physical Sciences Research Council (EP/G059063/1)
- Topics
- Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Patient Safety and Medication Errors
- Physical Sciences › Computer Science › Human-Computer Interaction › Usability and User Interface Design
- Keywords
- patient safety, work-as-imagined, work-as-done, device usability, user errors, system vulnerabilities, user experience, co-adaptation, health care purchasing, medical device design
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.