CoChat

Safety and Functional Integrity of Continuous Glucose Monitoring Components After Simulated Radiologic Procedures

Open access
Thomas, Christopher, Welsh, John B., Lu, Spencer, Gray, J. Michael · 2021 · Journal of Diabetes Science and Technology
Cited by 37

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

Background: We investigated wearable components of the Dexcom G6 continuous glucose monitoring (CGM) System in simulated therapeutic and diagnostic radiologic procedures. Methods: G6 transmitters were loaded with simulated glucose data and attached to sensors. Sets of sensor/transmitter pairs were exposed to x-rays to simulate a radiotherapeutic procedure and to radiofrequency (RF) and magnetic fields to simulate diagnostic magnetic resonance imaging (MRI). The x-ray simulation provided a cumulative dose of 80 Gy. The MRI simulation used RF fields oscillating at 64 or 128 MHz and magnetic fields of 1.5 or 3 T. During the MRI simulation, displacement force, induced heating, and induced currents were measured. After the simulations, bench tests were used to assess data integrity on the transmitters and responsiveness of sensors to various concentrations of aqueous glucose. Results: Glucose concentrations reported by sensor/transmitter pairs after undergoing x-irradiation or a simulated MRI exam were similar to those from control (unexposed) devices. During the 3 T MRI simulation, the devices experienced a displacement force of 306 g, which was insufficient to dislodge the sensor/transmitter from the substrate, RF-induced heating of <2°C, and an induced current of <16 pA. Data stored on the transmitters prior to the MRI simulation remained intact. Conclusion: Wearable components of the G6 CGM System retain basic functionality and data integrity after exposure to simulated therapeutic and diagnostic radiologic procedures. The devices are unlikely to be affected by x-irradiation used in typical imaging studies. Simulated MRI procedures create displacement force, minimal heating, and current in sensor/transmitter pairs.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1177/1932296820920948

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 37

Metadata

Authors
Thomas, C., Welsh, J. B., Lu, S., & Gray, J. M.
Venue
Journal of Diabetes Science and Technology
Published
2020-04-22
Volume
15
Issue
4
Pages
781-785
ISSN
1932-2968 (ISSN-L)
OpenAlex
W3016327970
Cited by
37
References
2
FWCI
3.98
Type
journal article
Language
en
Open access
hybrid

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.