Electromagnetic interference of GSM mobile phones with the implantable deep brain stimulator, ITREL-III

Open access
Kainz, Wolfgang, Alesch, François, Chan, Dulciana Dias·2003·BioMedical Engineering OnLine

Abstract

BACKGROUND: The purpose was to investigate mobile phone interference with implantable deep brain stimulators by means of 10 different 900 Mega Hertz (MHz) and 10 different 1800 MHz GSM (Global System for Mobile Communications) mobile phones. METHODS: All tests were performed in vitro using a phantom especially developed for testing with deep brain stimulators. The phantom was filled with liquid phantom materials simulating brain and muscle tissue. All examinations were carried out inside an anechoic chamber on two implants of the same type of deep brain stimulator: ITREL-III from Medtronic Inc., USA. RESULTS: Despite a maximum transmitted peak power of mobile phones of 1 Watt (W) at 1800 MHz and 2 W at 900 MHz respectively, no influence on the ITREL-III was found. Neither the shape of the pulse form changed nor did single pulses fail. Tests with increased transmitted power using CW signals and broadband dipoles have shown that inhibition of the ITREL-III occurs at frequency dependent power levels which are below the emissions of GSM mobile phones. The ITREL-III is essentially more sensitive at 1800 MHz than at 900 MHz. Particularly the frequency range around 1500 MHz shows a very low interference threshold. CONCLUSION: These investigations do not indicate a direct risk for ITREL-III patients using the tested GSM phones. Based on the interference levels found with CW signals, which are below the mobile phone emissions, we recommend similar precautions as for patients with cardiac pacemakers: 1. The phone should be used at the ear at the opposite side of the implant and 2. The patient should avoid carrying the phone close to the implant.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1186/1475-925x-2-11

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

The citation lists haven’t been fetched yet

References 22 · Cited by 16

Research works printed but not linked: 26

  1. Altamura G, Toscano S, Gentilucci G, Ammirati F: Influence of digital and analogue cellular telephones on implanted pacemakers. Eur Heart 1997, 18: 1632–1641. · DOI
  2. Barbaro V, Bartolini P, Donato A, Militello C: Electromagnetic interference of analog cellular telephones with pacemakers: In vitro and in vivo studies. Pace 1999, 22: 626–634. · DOI
  3. Carillo RG, Williams DB, Traad EA, Schor JS: Electromagnetic filters impede adverse interference of pacemakers by digital cellular telephones [Abstract]. Journal of the American College of Cardiology 1996,27(6):15A. 10.1016/S0735-1097(96)80243-9 · DOI
  4. Chen WH, Lau CP, Leung S-K, Sai-Wah Ho D: Interference of cellular phones with implanted permanent pacemakers. Clin Cardiol 1996, 19: 881–886. · DOI
  5. Ellenbogen KA, Wood MA: Cellular telephones and pacemakers: Urgent call or wrong number? Journal of the American College of Cardiology 1996,27(6):1478–1479. 10.1016/0735-1097(96)00086-1 · DOI
  6. Hayes D, Carrillo R, Findlay G, Embrey M: State of the science: Pacemaker and defibrillator interference from wireless communication devices. Pace 1996, 19: 1419–1430. · DOI
  7. Hayes D, Wang P, Reynolds D, Estes M: Interference with cardiac pacemakers by cellular telephones. The New England Journal of Medicine 1997,336(21):1473–1479. 10.1056/NEJM199705223362101 · DOI
  8. Irnich W, Batz L, Müller R, Tobisch R: Electromagnetic interference of pacemakers by mobile phones. Pace 1996, 19: 1431–1446. · DOI
  9. Naegeli B, Osswald S, Deola M, Burkart F: Intermittent pacemaker dysfunction caused by digital mobile phones. Journal of the American College of Cardiology 1996,27(6):1471–1477. 10.1016/0735-1097(96)00031-9 · DOI
  10. Ruggera P, Witters D, Bassen H: In vitro testing of pacemakers for digital cellular phone electromagnetic interference. Biomedical Instrumentation & Technology 1997,31(4):358–371.
  11. Sparks P, Mond H, Joyner K, Wood M: The safety of digital mobile cellular telephones with minute ventilation rate adaptive pacemakers. Pace 1996, 19: 1451–455. · DOI
  12. Tan K, Hinberg I: Laboratory investigation of wireless digital communication system interference with implantable cardiac pacemakers. WTR Symposium Washington DC 1996.
  13. Trigano A, Azoulay A, Rochdi M, Campillo A: Electromagnetic interference of external pacemakers by walkie-talkies and digital cellular phones: experimental study. Pace 1999, 22: 588–593. · DOI
  14. Wilke A, Grimm W, Funk R, Maisch B: Influence of D-Net (European GSM-Standard) cellular phones on pacemaker function in 50 patients with permanent pacemakers. Pace 1996, 19: 1456–1458. · DOI
  15. Nowak B, Rosocha S, Zellerhoff C, Liebrich A: Is there a risk for interaction between mobile phones and single lead VDD pacemakers? Pace 1996, 19: 1447–1450. · DOI
  16. Ravindran A, Schlegel R, Grant H, Raman S: In vitro study of the interaction between wireless phones and cardiac pacemakers. WTR Symposium Washington DC 1996.
  17. Raychowdhury G, Biswas A, Mitra K, Dasbiswas A: Interference of pacemaker by mobile phones [Abstract]. Pace 1997,20(II):2337.
  18. Ruggera P, Ravindran A, Tan K, Hayes L: Evaluation of the potential for hand held wireless telephones to interfere with implanted cardiac pacemakers. Wireless Technology Research Scientific Symposium Washington DC 1996.
  19. Barbaro V, Bartolini P, Bellocci F, Donato A: Electromagnetic interference of digital and analog cellular telephones with implantable car-dioverter defibrillators: In vitro and in vivo studies. Pace 1999, 22: 626–634. · DOI
  20. Bassen H, Moore H, Ruggera P: Cellular phone interference testing of implantable cardiac defibrillator, in-vitro. Circulation 1995, 92: I-738.
  21. Kainz W, Neubauer G: Electromagnetic compatibility (EMC) of electronic implants. Austrian Research Centers Seibersdorf Report, OEFZS–E-0004 1999.
  22. Kainz W, Neubauer G, Alesch F, Schmid G, Jahn O: Electromagnetic Compatibility of Electronic Implants – Review of the Literature. Wiener Klinische Wochenschrift 2001, 113: 23–24.
  23. Alesch F, Pinter MM, Helscher RJ, Fertl L, Benabid AL, Koos WT: Stimulation of the ventral intermediate thalamic nucleus in tremor dominated Parkinson's disease and essential tremor. Acta Neurochir 1995,136(1–2):75–81. · DOI
  24. Limousin P, Pollak P, Benazzouz A, Hoffmann D, Le Bas JF, Broussolle E, Perret JE, Benabid AL: Effect of parkinsonian signs and symptoms of bilateral subthalamic nucleus stimulation. Lancet 1995,345(8942):91–95. 10.1016/S0140-6736(95)90062-4 · DOI
  25. Gabriel S, Lau R, Gabriel C: The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHz. Phys Med Biol 1996, 41: 2251–2269. 10.1088/0031-9155/41/11/002 · DOI
  26. Institute of Electrical and Electronics Engineers (IEEE) SCC34 Draft Standard: Recommended Practice for Determining the Peak Spatial-Average Specific Absorption Rate (SAR) in the Human Body Due to Wireless Communications Devices: Experimental Techniques. IEEE Std 1528–2002.

Metadata

Authors
Kainz, W., Alesch, F., & Chan, D. D.
Affiliations
  • Seibersdorf Laboratories (Austria) (AT)
  • Vienna General Hospital (AT)
  • Johns Hopkins University (US)
Corresponding author
Kainz, Wolfgang
Venue
BioMedical Engineering OnLine
Published
2003-05-07
Volume
2
Issue
1
Pages
11-11
ISSN
1475-925X (ISSN-L)
OpenAlex
W2126300430
Cited by
16
References
22
Citations by year
2025: 1 · 2024: 1 · 2021: 2 · 2017: 1 · 2016: 2 · 2015: 1 · 2013: 1 · 2012: 1
FWCI
0.00
Type
journal article
Language
en
Open access
Locations
  • BioMedical Engineering OnLine (publishedVersion)
  • PubMed Central (publishedVersion, other-oa)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Topics
  • Health Sciences › Medicine › Clinical Neurology › Deep Brain Stimulation and Movement Disorders
  • Life Sciences › Biochemistry, Genetics and Molecular Biology › Biophysics › Electromagnetic Fields and Biological Effects
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
Keywords
electromagnetic interference, deep brain stimulator, GSM mobile phones, 900 MHz, 1800 MHz, biomedical implants, interference threshold, continuous wave signals, tissue phantoms
MeSH terms
Brain Diseases, Electrodes, Implanted, Electromagnetic Fields, Electric Stimulation Therapy, Humans, Models, Anatomic, Pacemaker, Artificial, Risk Assessment, Phantoms, Imaging, Prostheses and Implants, Cell Phone

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Electromagnetic interference of GSM mobile phones with the implantable deep brain stimulator, ITREL-III. https://app.cochat.ai/papers/10.1186/1475-925x-2-11
The paper
Kainz, W., Alesch, F., & Chan, D. D. (2003). Electromagnetic interference of GSM mobile phones with the implantable deep brain stimulator, ITREL-III. BioMedical Engineering OnLine. https://doi.org/10.1186/1475-925x-2-11