Complexity of MRI induced heating on metallic leads: Experimental measurements of 374 configurations
Open accessThe DOI matches its Crossref record; 9 of the 9 references it cites were found in a registry or an index.
Abstract
BACKGROUND: MRI induced heating on PM leads is a very complex issue. The widely varying results described in literature suggest that there are many factors that influence the degree of heating and that not always are adequately addressed by existing testing methods. METHODS: We present a wide database of experimental measurements of the heating of metallic wires and PM leads in a 1.5 T RF coil. The aim of these measurements is to systematically quantify the contribution of some potential factors involved in the MRI induced heating: the length and the geometric structure of the lead; the implant location within the body and the lead path; the shape of the phantom used to simulate the human trunk and its relative position inside the RF coil. RESULTS: We found that the several factors are the primary influence on heating at the tip. Closer locations of the leads to the edge of the phantom and to the edge of the coil produce maximum heating. The lead length is the other crucial factor, whereas the implant area does not seem to have a major role in the induced temperature increase. Also the lead structure and the geometry of the phantom revealed to be elements that can significantly modify the amount of heating. CONCLUSION: Our findings highlight the factors that have significant effects on MRI induced heating of implanted wires and leads. These factors must be taken into account by those who plan to study or model MRI heating of implants. Also our data should help those who wish to develop guidelines for defining safe medical implants for MRI patients. In addition, our database of the entire set of measurements can help those who wish to validate their numerical models of implants that may be exposed to MRI systems.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Journal of Magnetic Resonance Imaging
(2006) · Journal of Magnetic Resonance Imaging · doi:10.1002/(issn)1522-2586 · Cited by 666
Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla
Martin, E. T., Coman, J. A., Shellock, F. G., Pulling, C. C., Fair, R., & Jenkins, K. (2004) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2003.12.016 · Cited by 422 · cited 1×
“some edema occurred at the tip-lead tissue interface, which subsequently resolved." Martin and Coman [ 7 ] observed 9.4% of their patients undergo "significant" changes to their pacing thresholds after MRI.”
BackgroundInfluentialCurrent Clinical Issues for MRI Scanning of Pacemaker and Defibrillator Patients
KALIN, R., & STANTON, M. S. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.50024.x · Cited by 381 · cited 1×
“The outstanding increases in both MRI usage and cardiac device-based therapy have resulted in an estimated 50–75% probability of a patient being indicated for an MRI over the lifetime of their device [1,2].”
BackgroundRoguin, A., Zviman, M. M., Meininger, G. R., Rodrigues, E. R., Dickfeld, T. M., Bluemke, D. A., Lardo, A., Berger, R. D., Calkins, H., & Halperin, H. R. (2004) · Circulation · doi:10.1161/01.cir.0000137121.28722.33 · Cited by 355
On the heating of linear conductive structures as guide wires and catheters in interventional MRI
Nitz, W. R., Oppelt, A., Renz, W., Manke, C., Lenhart, M., & Link, J. (2001) · Journal of Magnetic Resonance Imaging · doi:10.1002/1522-2586(200101)13:1<105::aid-jmri1016>3.0.co;2-0 · Cited by 321
Effects of magnetic resonance imaging on cardiac pacemakers and electrodes
Achenbach, S. S., Moshage, W. E. L., Diem, B., Bieberlea, T., Schibgilla, V., & Bachmann, K. F. (1997) · American Heart Journal · doi:10.1016/s0002-8703(97)70083-8 · Cited by 254 · cited 1×
“[5] reported pacing lead tip temperature elevation in leads not attached to an implanted pulse generator (IPG) of >63°C during 90 seconds of MRI.”
BackgroundMR Imaging and Cardiac Pacemakers: In Vitro Evaluation and in Vivo Studies in 51 Patients at 0.5 T
Sommer, T., Vahlhaus, C., Lauck, G., Smekal, A. V., Reinke, M., Hofer, U., Block, W., Träber, F., Schneider, C., Gieseke, J., et al. (2000) · Radiology · doi:10.1148/radiology.215.3.r00jn08869 · Cited by 244
Rezai, A. R., Baker, K. B., Tkach, J. A., Phillips, M., Hrdlicka, G., Sharan, A. D., Nyenhuis, J. A., Ruggieri, P. M., Shellock, F. G., & Henderson, J. M. (2005) · Neurosurgery · doi:10.1227/01.neu.0000186935.87971.2a · Cited by 104 · cited 3×
“Rezai et al. [ 10 ] reported deep brain stimulator temperature elevations at the electrode tip of >25°C within 15 minutes of MR imaging.”
BackgroundInfluentialBassen, H., Kainz, W., Mendoza, G., & Kellom, T. (2006) · Minimally Invasive Therapy & Allied Technologies · doi:10.1080/13645700600640931 · Cited by 72 · cited 4×
“Data acquisition was performed using the analog output of the Luxtron thermometer, an A/D converter and a 16-bit acquisition card (National Instrument, DAQCard-AI-16XE-50), installed on a standard notebook computer (Acer 1670, Windows XP OS) In previous studies [9,17] the positions and mounting of the temperature prob…”
BackgroundMethodInfluentialSelf-citation
Cited by
Roy, S., Azad, A. N. M. W., Baidya, S., Alam, M. K., & Khan, F. (2022) · IEEE Transactions on Power Electronics · doi:10.1109/tpel.2022.3164890 · Cited by 255
On the model validation of active implantable medical device for MRI safety assessment
Wang, Z., Zheng, J., Wang, Y., Kainz, W., & Chen, J. (2020) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2019.2957766 · Cited by 22
Self-citation
Metadata
- Authors
- Mattei, E., Triventi, M., Calcagnini, G., Censi, F., Kainz, W., Mendoza, G., Bassen, H. I., & Bartolini, P.
- Venue
- BioMedical Engineering OnLine
- Published
- 2008
- Volume
- 7
- Issue
- 1
- Pages
- 11
- ISSN
- 1475-925X (ISSN-L)
- OpenAlex
- W2093816193
- Cited by
- 181
- References
- 9
- FWCI
- 6.45
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Complexity of MRI induced heating on metallic leads: Experimental measurements of 374 configurations. https://app.cochat.ai/papers/10.1186/1475-925x-7-11
- The paper
- Mattei, E., Triventi, M., Calcagnini, G., Censi, F., Kainz, W., Mendoza, G., Bassen, H. I., & Bartolini, P. (2008). Complexity of MRI induced heating on metallic leads: Experimental measurements of 374 configurations. BioMedical Engineering OnLine. https://doi.org/10.1186/1475-925x-7-11
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.