T1-mapping in the heart: accuracy and precision
Open accessAbstract
The longitudinal relaxation time constant (T1) of the myocardium is altered in various disease states due to increased water content or other changes to the local molecular environment. Changes in both native T1 and T1 following administration of gadolinium (Gd) based contrast agents are considered important biomarkers and multiple methods have been suggested for quantifying myocardial T1 in vivo. Characterization of the native T1 of myocardial tissue may be used to detect and assess various cardiomyopathies while measurement of T1 with extracellular Gd based contrast agents provides additional information about the extracellular volume (ECV) fraction. The latter is particularly valuable for more diffuse diseases that are more challenging to detect using conventional late gadolinium enhancement (LGE). Both T1 and ECV measures have been shown to have important prognostic significance. T1-mapping has the potential to detect and quantify diffuse fibrosis at an early stage provided that the measurements have adequate reproducibility. Inversion recovery methods such as MOLLI have excellent precision and are highly reproducible when using tightly controlled protocols. The MOLLI method is widely available and is relatively mature. The accuracy of inversion recovery techniques is affected significantly by magnetization transfer (MT). Despite this, the estimate of apparent T1 using inversion recovery is a sensitive measure, which has been demonstrated to be a useful tool in characterizing tissue and discriminating disease. Saturation recovery methods have the potential to provide a more accurate measurement of T1 that is less sensitive to MT as well as other factors. Saturation recovery techniques are, however, noisier and somewhat more artifact prone and have not demonstrated the same level of reproducibility at this point in time.This review article focuses on the technical aspects of key T1-mapping methods and imaging protocols and describes their limitations including the factors that influence their accuracy, precision, and reproducibility.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Messroghli, D. R., Moon, J. C., Ferreira, V. M., Grosse-Wortmann, L., He, T., Kellman, P., Mascherbauer, J., Nezafat, R., Salerno, M., Schelbert, E. B., et al. (2016) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-017-0389-8 · Cited by 1,727
Self-citationCardiac T1 Mapping and Extracellular Volume (ECV) in clinical practice: a comprehensive review
Haaf, P., Garg, P., Messroghli, D. R., Broadbent, D. A., Greenwood, J. P., & Plein, S. (2016) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-016-0308-4 · Cited by 959
T1-Mapping and Outcome in Nonischemic Cardiomyopathy
Puntmann, V. O., Carr-White, G., Jabbour, A., Yu, C.-Y., Gebker, R., Kelle, S., Hinojar, R., Doltra, A., Varma, N., Child, N., et al. (2016) · JACC: Cardiovascular Imaging · doi:10.1016/j.jcmg.2015.12.001 · Cited by 478
Akçakaya, M., Moeller, S., Weingärtner, S., & Uğurbil, K. (2019) · Magnetic Resonance in Medicine · doi:10.1002/mrm.27420 · Cited by 391
Comprehensive Cardiac Magnetic Resonance Imaging in Patients With Suspected Myocarditis
Lurz, P., Luecke, C., Eitel, I., Föhrenbach, F., Frank, C., Grothoff, M., de Waha, S., Rommel, K.-P., Lurz, J. A., Klingel, K., et al. (2016) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2016.02.013 · Cited by 389
Metadata
- Authors
- Kellman, P., & Hansen, M. S.
- Venue
- Journal of Cardiovascular Magnetic Resonance
- Publisher
- Elsevier BV
- Published
- 2014-12
- Volume
- 16
- Issue
- 1
- Pages
- 2
- ISSN
- 1097-6647 (ISSN-L)
- OpenAlex
- W2159285706
- Cited by
- 764
- References
- 0
- FWCI
- 50.27
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: T1-mapping in the heart: accuracy and precision. https://app.cochat.ai/papers/10.1186/1532-429x-16-2
- The paper
- Kellman, P., & Hansen, M. S. (2014). T1-mapping in the heart: accuracy and precision. Journal of Cardiovascular Magnetic Resonance. https://doi.org/10.1186/1532-429x-16-2
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.