Do changes in pulse oximeter oxygen saturation predict equivalent changes in arterial oxygen saturation?

Open access
Perkins, Gavin D, McAuley, Daniel F, Giles, Simon, Routledge, Helen, Gao, Fang·2003·Critical Care

Abstract

Introduction: This study investigates the relation between changes in pulse oximeter oxygen saturation (Sp O2) and changes in arterial oxygen saturation (Sa O2) in the critically ill, and the effects of acidosis and anaemia on precision of using pulse oximetry to predict Sa O2. Patients and methods: Forty-one consecutive patients were recruited from a nine-bed general intensive care unit into a 2-month study. Patients with significant jaundice (bilirubin >40 μmol/l) or inadequate pulse oximetry tracing were excluded. Results: A total of 1085 paired readings demonstrated only moderate correlation (r= 0.606; P < 0.01) between changes in Sp O2 and those in Sa O2, and the pulse oximeter tended to overestimate actual changes in Sa O2. Anaemia increased the degree of positive bias whereas acidosis reduced it. However, the magnitude of these changes was small. Conclusion: Changes in Sp O2 do not reliably predict equivalent changes in Sa O2 in the critically ill. Neither anaemia nor acidosis alters the relation between Sp O2 and Sa O2 to any clinically important extent.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1186/cc2339

Cites

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Pulse oximetry should remain an important diagnostic tool in patients with anemia and sickle cell anemia.

    “pulse oximetry should remain an important diagnostic tool in these patients”

    From Investigating the accuracy of blood oxygen saturation measurements in common consumer smartwatches (2023) · cites this paper as [30]

    The Conclusion calls pulse oximetry a valuable tool in ICU care and reports that anaemia did not alter measurement precision to a clinically important extent. But the Discussion says no patients were known to have sickle cell anaemia or trait, so this study does not support extending that finding specifically to those patients. (LLM, conf=0.88)

    Experimental — accuracy not yet measured
  • In patients with anemia and sickle cell anemia, variability in the relationship between SpO2 and true arterial saturation measurements is not clinically significant.

    “although there is some variability in the relationship between SpO2 and the true arterial saturation measurements, this was not a clinically significant difference”

    From Investigating the accuracy of blood oxygen saturation measurements in common consumer smartwatches (2023) · cites this paper as [30]

    The Discussion says anaemia affects the relationship between SpO2 and SaO2, but that the clinical significance of this effect is small. It also says no patients were known to have sickle cell anaemia or trait, so the finding cannot be extended to that group. The paper further cautions that changes in SpO2 do not reliably predict equivalent changes in SaO2 overall. (LLM, conf=0.88)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

The citation lists haven’t been fetched yet

References 26 · Cited by 136

Research works printed but not linked: 22

  1. Jubran A Advances in respiratory monitoring during mechanical ventilation Chest 1999 116 1416 1425 10559107 10.1378/chest.116.5.1416 · DOI
  2. Paky F Koeck CM Pulse oximetry in ventilated preterm newborns: reliability of detection of hyperoxaemia and hypoxaemia, and feasibility of alarm settings Acta Paediatr 1995 84 613 616 7670240
  3. Inman KJ Sibbald WJ Rutledge FS Speechley M Martin CM Clark BJ Does implementing pulse oximetry in a critical care unit result in substantial arterial blood gas savings? Chest 1993 104 542 546 8339646
  4. Slutsky AS Consensus conference on mechanical ventilation: January 28–30, 1993 at Northbrook, Illinois, USA. Part I. European Society of Intensive Care Medicine, the ACCP and the SCCM Intensive Care Med 1994 20 64 79 8163765
  5. Seguin P Le Rouzo A Tanguy M Guillou YM Feuillu A Malledant Y Evidence for the need of bedside accuracy of pulse oxime-try in an intensive care unit Crit Care Med 2000 28 703 706 10752818
  6. Rotello LC Warren J Jastremski MS Milewski A A nurse-directed protocol using pulse oximetry to wean mechanically ventilated patients from toxic oxygen concentrations Chest 1992 102 1833 1835 1446498
  7. Lee S Tremper KK Barker SJ Effects of anemia on pulse oximetry and continuous mixed venous hemoglobin saturation monitoring in dogs Anesthesiology 1991 75 118 122 2064035
  8. Jay GD Hughes L Renzi FP Pulse oximetry is accurate in acute anemia from hemorrhage Ann Emerg Med 1994 24 32 35 8010546
  9. Ramsing T Rosenberg J Pulse oximetry in severe anaemia Intensive Care Med 1992 18 125 126 1613193
  10. Ralston AC Webb RK Runciman WB Potential errors in pulse oximetry. I. Pulse oximeter evaluation Anaesthesia 1991 46 202 206 2014898
  11. Bland JM Altman DG Comparing methods of measurement: why plotting difference against standard method is misleading Lancet 1995 346 1085 1087 7564793
  12. Van de Louw A Cracco A Cert C Harf A Duvaldesin P Lemaire F Brochard L Accuracy of pulse oximetry in the intensive care unit Intensive Care Med 2001 27 1606 1613 11685301 10.1007/s001340101064 · DOI
  13. Wukitsch MW Petterson MT Tobler DR Pologe JA Pulse oximetry: analysis of theory, technology, and practice J Clin Monit 1988 4 290 301 3057122
  14. Severinghaus JW Koh SO Effect of anemia on pulse oximeter accuracy at low saturation J Clin Monit 1990 6 85 88 2352007
  15. Ortiz FO Aldrich TK Nagel RL Benjamin LJ Accuracy of pulse oximetry in sickle cell disease Am J Respir Crit Care Med 1999 159 447 451 9927356
  16. Palve H Vuori A Accuracy of three pulse oximeters at low cardiac index and peripheral temperature Crit Care Med 1991 19 560 562 2019144
  17. Secker C Spiers P Accuracy of pulse oximetry in patients with low systemic vascular resistance Anaesthesia 1997 52 127 130 9059094
  18. Ibanez J Velasco J Raurich JM The accuracy of the Biox 3700 pulse oximeter in patients receiving vasoactive therapy Intensive Care Med 1991 17 484 486 1797894
  19. Watcha MF Connor MT Hing AV Pulse oximetry in methemoglobinemia Am J Dis Child 1989 143 845 847 2741859
  20. Bozeman WP Myers RA Barish RA Confirmation of the pulse oximetry gap in carbon monoxide poisoning Ann Emerg Med 1997 30 608 611 9360570
  21. Turner JA McNicol MW Sillett RW Distribution of carboxyhaemoglobin concentrations in smokers and non-smokers Thorax 1986 41 25 27 3704963 PMC460247
  22. Veyckemans F Baele P Guillaume JE Willems E Robert A Clerbaux T Hyperbilirubinemia does not interfere with hemoglobin saturation measured by pulse oximetry Anesthesiology 1989 70 118 122 2912292

Authors
Perkins, G. D., McAuley, D. F., Giles, S., Routledge, H., & Gao, F.
Affiliations
Heartlands Hospital (GB)
Corresponding author
Gao, Fang
Venue
Critical Care
Published
2003-06-11
Volume
7
Issue
4
Pages
R67-R67
ISSN
1364-8535 (ISSN-L)
OpenAlex
W1757861364
Cited by
136
References
26
Citations by year
2026: 4 · 2025: 9 · 2024: 7 · 2023: 11 · 2022: 15 · 2021: 15 · 2020: 6 · 2019: 5 · 2018: 9 · 2017: 8 · 2016: 5 · 2015: 10 · 2014: 6 · 2013: 5 · 2012: 5
FWCI
1.58
Type
journal article
Language
en
Open access
Locations
  • Critical Care (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, other-oa)
  • PubMed Central (publishedVersion, other-oa)
  • PubMed (publishedVersion)
  • University of Birmingham Research Portal (University of Birmingham) (submittedVersion, other-oa)
  • Research Portal (Queen's University Belfast) (submittedVersion)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
  • Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Respiratory Support and Mechanisms
Keywords
pulse oximetry, arterial oxygen saturation, critically ill patients, acidosis, anemia, measurement bias, jaundice, intensive care unit
MeSH terms
Acidosis, Aged, Aged, 80 and over, Anemia, Arteries, Female, United Kingdom, Humans, Intensive Care Units, Male, Middle Aged, Oximetry, Oxygen, Oxygen Consumption, Critical Illness

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review