PURPOSE: The purpose of this study was to examine the accuracy of the Polar S410 for estimating gross energy expenditure (EE) during exercise when using both predicted and measured VO2max and HRmax versus indirect calorimetry (IC). METHODS: Ten males and 10 females initially had their VO2max and HRmax predicted by the S410, and then performed a maximal treadmill test to determine their actual values. The participants then performed three submaximal exercise tests at RPE of 3, 5, and 7 on a treadmill, cycle, and rowing ergometer for a total of nine submaximal bouts. For all submaximal testing, the participant had two S410 heart rate monitors simultaneously collecting data: one heart rate monitor (PHRM) utilized their predicted VO2max and HRmax, and one heart rate monitor (AHRM) used their actual values. Simultaneously, EE was measured by IC. RESULTS: In males, there were no differences in EE among the mean values for the AHRM, PHRM, and IC for any exercise mode (P > 0.05). In females, the PHRM significantly overestimated mean EE on the treadmill (by 2.4 kcal x min(-1)), cycle (by 2.9 kcal x min(-1)), and rower (by 1.9 kcal x min(-1)) (all P < 0.05). The AHRM for females significantly improved the estimation of mean EE for all exercise modes, but it still overestimated mean EE on the treadmill (by 0.6 kcal x min(-1)) and cycle (by 1.2 kcal x min(-1)) (P < 0.05). CONCLUSION: When the predicted values of VO2max and HRmax are used, the Polar S410 HRM provides a rough estimate of EE during running, rowing, and cycling. Using the actual values for VO2max and HRmax reduced the individual error scores for both genders, but in females the mean EE was still overestimated by 12%.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Eston, R. G., Rowlands, A. V., & Ingledew, D. K. (1998) · Journal of Applied Physiology · doi:10.1152/jappl.1998.84.1.362 · Cited by 643 · cited 1×
“Furthermore, because HR is linearly related to oxygen uptake for dynamic activities involving large muscle groups (6,24), it can provide a reasonable estimate of energy expenditure (EE) during exercise (5,8).”
BackgroundA combined heart rate and movement sensor: proof of concept and preliminary testing study
Rennie, K., Rowsell, T., Jebb, S., Holburn, D., & Wareham, N. (2000) · European Journal of Clinical Nutrition · doi:10.1038/sj.ejcn.1600973 · Cited by 83 · cited 1×
“The present study examined planned bouts of structured exercise whereas Flex HR studies have used much longer time periods, ranging from 6 h (23,25) to 3–4 d (18).”
Background
Cited by
Practical Guide to Measuring Physical Activity
Sylvia, L. G., Bernstein, E. E., Hubbard, J. L., Keating, L., & Anderson, E. J. (2014) · Journal of the Academy of Nutrition and Dietetics · doi:10.1016/j.jand.2013.09.018 · Cited by 637
Accuracy of commercially available heart rate monitors in athletes: a prospective study
Pasadyn, S. R., Soudan, M., Gillinov, M., Houghtaling, P., Phelan, D., Gillinov, N., Bittel, B., & Desai, M. Y. (2019) · Cardiovascular Diagnosis and Therapy · doi:10.21037/cdt.2019.06.05 · Cited by 135
“There have been studies examining wrist worn HR monitor accuracy during aerobic activity, demonstrating their validity, but none at high levels of exertion (3,5-7).”
BackgroundKato, H., Suzuki, K., Bannai, M., & Moore, D. R. (2016) · PLOS ONE · doi:10.1371/journal.pone.0157406 · Cited by 128
Menstrual disruption with exercise is not linked to an energy availability threshold
LIEBERMAN, J. L., DE SOUZA, M. J., WAGSTAFF, D. A., & WILLIAMS, N. I. (2018) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001451 · Cited by 127
Energy availability discriminates clinical menstrual status in exercising women
Reed, J. L., De Souza, M. J., Mallinson, R. J., Scheid, J. L., & Williams, N. I. (2015) · Journal of the International Society of Sports Nutrition · doi:10.1186/s12970-015-0072-0 · Cited by 86
“The OwnCal feature has been validated for calculating EEE from heart rate in University age females running, cycling, and rowing at various submaximal intensities [27].”
Method
Metadata
- Authors
- CROUTER, S. E., ALBRIGHT, C., & BASSETT, D. R.
- Venue
- Medicine & Science in Sports & Exercise
- Published
- 2004-08
- Volume
- 36
- Issue
- 8
- Pages
- 1433-1439
- ISSN
- 0195-9131 (ISSN-L)
- OpenAlex
- W2072898489
- Cited by
- 140
- References
- 3
- FWCI
- 7.99
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Accuracy of Polar S410 Heart Rate Monitor to Estimate Energy Cost of Exercise. https://app.cochat.ai/papers/10.1249/01.mss.0000135794.01507.48
- The paper
- CROUTER, S. E., ALBRIGHT, C., & BASSETT, D. R. (2004). Accuracy of Polar S410 Heart Rate Monitor to Estimate Energy Cost of Exercise. Medicine & Science in Sports & Exercise. https://doi.org/10.1249/01.mss.0000135794.01507.48
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.