The DOI matches its Crossref record; the references it cites are not checked yet.
First experience with spatial frequency domain imaging and red-light excitation of protoporphyrin IX fluorescence during tumor resection
Open accessAbstract
Fluorescence-guided surgery (FGS) enhances intraoperative visualization of tumors to maximize safe resection, and quantitative fluorescence imaging (qFI) of protoporphyrin IX (PpIX) has provided additional information for guidance during intracranial tumor surgery. Previous developments in fluorescence quantification have demonstrated that the depth of fluorescence signals can be estimated given known optical properties in a lab setting, and now with the work described here that these optical properties can be determined in vivo in human brain tissue in the operating room (OR) during tumor resection procedures. More specifically, we report the first depth estimation of subsurface tumor intraoperatively, achieved with the combination of spatial frequency domain imaging (SFDI) for optical property measurement and red-light excitation of PpIX. We modified a commercial surgical microscope (Zeiss) with a digital light processing module (DLI Austin, TX) to modulate light from a xenon arc lamp to illuminate the field. White-light excitation and a liquid crystal tunable filter (LCTF Verispec) were used to measure diffuse reflectance at discrete wavelengths of 670 nm and 710 nm on a sCMOS camera. An illumination-side filter wheel allowed excitation of PpIX fluorescence at 405 nm and 635 nm, and the LCTF measured fluorescence emissions at 670 nm and 710 nm. Data acquisition and processing generated wide-field images of the depth of PpIX fluorescence within 1 minute in the OR. The ability of the clinical microscope to perform optical property mapping with SFDI and convert these wide-field estimates into images of the depth of fluorescence was tested in tissue simulating phantoms and in vivo during a craniotomy for brain tumor resection. Results indicate that wide-field optical property estimates with SFDI can be combined with depth sensing algorithms to produce maps of the depth of PpIX when exposed to red-light in the OR.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
PpIX has maximum absorption at around 405 nm in the Soret band and multiple absorption peaks in the visible Q-bands, enabling excitation with longer wavelengths for deeper tissue penetration.
“With the maximum absorption at around 405 nm in the soret band, PpIX also has multiple absorption peaks in the visible Q-bands, enabling its excitation with longer wavelengths of light to achieve deeper tissue penetration [42].”
From 5-Aminolevulinic Acid as a Theranostic Agent for Tumor Fluorescence Imaging and Photodynamic Therapy (2023) · cites this paper as [42]
Following light excitation, PpIX emits a broad spectrum of red fluorescence centered at about 635 nm and extending beyond 700 nm.
“Following light excitation, PpIX emits a broad spectrum of red fluorescence emission that is centered at about 635 nm and extends beyond 700 nm.”
From 5-Aminolevulinic Acid as a Theranostic Agent for Tumor Fluorescence Imaging and Photodynamic Therapy (2023) · cites this paper as [42]
Citations
The citation lists haven’t been fetched yet
References 21 · Cited by 14
Research works printed but not linked: 20
- Lacroix, 2001, J. Neurosurg., 95, 190, 10.3171/jns.2001.95.2.0190 · DOI
- Stummer, 2009, Curr. Opin. Neurol., 22, 645, 10.1097/WCO.0b013e3283320165 · DOI
- Colditz, 2012, J. Clin. Neurosci., 19, 1471, 10.1016/j.jocn.2012.03.009 · DOI
- Stummer, 2006, Lancet Oncol., 7, 392, 10.1016/S1470-2045(06)70665-9 · DOI · Paper page
- Valdés, 2011, J. Biomed. Opt., 16, 116007, 10.1117/1.3646916 · DOI
- Valdés, 2015, J. Neurosurg., 123, 771, 10.3171/2014.12.JNS14391 · DOI
- Roberts, 2018, J. Neurosurg., 128, 1690, 10.3171/2017.1.JNS162061 · DOI
- Kolste, 2015, J. Biomed. Opt., 20, 026002, 10.1117/1.JBO.20.2.026002 · DOI
- Wirth, 2017, Biomed. Opt. Express, 8, 3656, 10.1364/BOE.8.003656 · DOI
- Cuccia, 2009, J. Biomed. Opt., 14, 024012, 10.1117/1.3088140 · DOI
- Erickson, 2010, J. Biomed. Opt., 15, 036013, 10.1117/1.3431728 · DOI
- Gioux, 2011, J. Biomed. Opt., 16, 086015, 10.1117/1.3614566 · DOI
- Sibai, 2017, J. Biomed. Opt., 22, 076007, 10.1117/1.JBO.22.7.076007 · DOI
- Laughney, 2013, J. Biomed. Opt., 18, 036012, 10.1117/1.JBO.18.3.036012 · DOI
- Wirth, 2018, J. Biomed. Opt., 24, 1, 10.1117/1.JBO.24.7.071608 · DOI
- Gioux, 2009, J. Biomed. Opt., 14, 034045, 10.1117/1.3156840 · DOI
- Srinivasan, 1984, Appl. Opt., 23, 3105, 10.1364/AO.23.003105 · DOI
- Zhou, 1994, J. Mod. Opt., 41, 89, 10.1080/09500349414550101 · DOI
- Michels, 2008, Opt. Express, 16, 5907, 10.1364/OE.16.005907 · DOI
- Hayakawa, 2018, J. Biomed. Opt., 24, 1, 10.1117/1.JBO.24.7.071603 · DOI
Metadata
- Authors
- Wirth, D. J., Sibai, M., Wilson, B. C., Roberts, D. W., & Paulsen, K.
- Affiliations
- Dartmouth–Hitchcock Medical Center (US)
- University Health Network (CA)
- Dartmouth College (US)
- Venue
- Biomedical Optics Express
- Publisher
- Optica Publishing Group
- Published
- 2020-07-15
- Volume
- 11
- Issue
- 8
- Pages
- 4306
- ISSN
- 2156-7085 (ISSN-L)
- PMC
- PMC7449712
- OpenAlex
- W3037464161
- Cited by
- 14
- References
- 21
- Citations by year
- 2024: 2 · 2023: 3 · 2022: 4 · 2021: 4 · 2020: 1
- FWCI
- 1.46
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- Biomedical Optics Express (publishedVersion)
- PubMed Central (publishedVersion, other-oa)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health (R01NS052274-01A2)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Cancer Research › Photodynamic Therapy Research Studies
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Biophysics › Spectroscopy Techniques in Biomedical and Chemical Research
- Keywords
- fluorescence-guided surgery, protoporphyrin IX, spatial frequency domain imaging, red light excitation, depth estimation, brain tumor resection, craniotomy, diffuse reflectance, tissue optical properties, liquid crystal tunable filter, digital light processing, tissue phantoms
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review