Uncertainty in measurement and total error - are they so incompatible?
Open accessThe DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
There appears to be a growing debate with regard to the use of “Westgard style” total error and “GUM style” uncertainty in measurement. Some may argue that the two approaches are irreconcilable. The recent appearance of an article “
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“However, methods for calculating measurement uncertainty are not prescribed in the standards so accreditation authorities accept both error and uncertainty approaches.47”
From Uncertainty in Measurement and Total Error (2017) · cites this paper as [47]
Citations
Cites
Tate, J. R., Sikaris, K. A., Jones, G. R. D., Yen, T., Koerbin, G., Ryan, J. E., Reed, M., Gill, J., Koumantakis, G., Hickman, P. E., et al. (2014) · Europe PMC (PubMed Central) · Cited by 87
Self-citationUseful measures and models for analytical quality management in medical laboratories
Westgard, J. O. (2016) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2015-0710 · Cited by 64 · cited 1×
“It is with some dismay we have observed a growing debate with regard to the use of “Westgard style” total error and “GUM style” uncertainty in measurement [1, 2].”
MethodEvidence-based approach to harmonised reference intervals
Koerbin, G., Sikaris, K. A., Jones, G. R. D., Ryan, J. E., Reed, M., & Tate, J. R. (2013) · Clinica Chimica Acta · doi:10.1016/j.cca.2013.10.021 · Cited by 57
Self-citationSelection and Analytical Evaluations of Methods-With Statistical Techniques
Línnet, K., & Boyd, J. C. (2006) · Research at the University of Copenhagen (University of Copenhagen) · Cited by 43
Bias Assessment of General Chemistry Analytes using Commutable Samples.
Koerbin, G., Tate, J. R., Ryan, J. E., Jones, G. R. D., Sikaris, K. A., Kanowski, D., Reed, M., Gill, J., Koumantakis, G., Yen, T., et al. (2014) · Europe PMC (PubMed Central) · Cited by 36
Self-citationTotal error vs. measurement uncertainty: the match continues
Panteghini, M., & Sandberg, S. (2015) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2015-1036 · Cited by 31 · cited 1×
“It is with some dismay we have observed a growing debate with regard to the use of “Westgard style” total error and “GUM style” uncertainty in measurement [1, 2].”
Method“Harmonization of laboratory testing — A global activity”
Tate, J. R., Johnson, R. N., Barth, J. H., & Panteghini, M. (2014) · Clinica Chimica Acta · doi:10.1016/j.cca.2014.02.006 · Cited by 19
Research works printed but not linked: 12
- 2001, Biological variation: from principles to practice, 29
- 2014, Harmonising adult and paediatric reference intervals in Australia and New Zealand: An evidence-based approach for establishing a first panel of chemistry analytes, Clin Biochem Rev, 35, 213
- Quality requirements and standards – quality goals at the crossroads: growing, going or gone?
- 2016, Total error vs. measurement uncertainty: the match continues, Clin Chem Lab Med, 54, 195
- 2014, Harmonization of laboratory testing – A global activity, Clin Chim Acta, 432, 1, 10.1016/j.cca.2014.02.006 · DOI
- 2008, Tietz, Fundamentals of Clinical Chemistry, 201
- 2014, Bias assessment of general chemistry analytes using commutable samples, Clin Biochem Rev, 35, 203
- 2001, Biological variation: from principles to practice, 29
- 2016, Useful measures and models for analytical quality management in medical laboratories, Clin Chem Lab Med, 54, 223
- 2014, Evidence-based approach to harmonized reference intervals, Clin Chim Acta, 432, 99, 10.1016/j.cca.2013.10.021 · DOI
- Quality requirements and standards – quality goals at the crossroads: growing, going or gone?
- 2008, Tietz, Fundamentals of Clinical Chemistry, 201
Cited by
X-ray computed tomography: from medical imaging to dimensional metrology
Villarraga-Gómez, H., Herazo, E. L., & Smith, S. T. (2019) · Precision Engineering · doi:10.1016/j.precisioneng.2019.06.007 · Cited by 233
Six Sigma Quality Management System and Design of Risk-based Statistical Quality Control
Westgard, J. O., & Westgard, S. A. (2016) · Clinics in Laboratory Medicine · doi:10.1016/j.cll.2016.09.008 · Cited by 68
Villarraga-Gómez, H., Thousand, J. D., & Smith, S. T. (2020) · Precision Engineering · doi:10.1016/j.precisioneng.2020.03.004 · Cited by 67
Uncertainty of Measurement in Laboratory Medicine
Milinković, N., Ignjatović, S., Šumarac, Z., & Majkić‐Singh, N. (2018) · Journal of Medical Biochemistry · doi:10.2478/jomb-2018-0002 · Cited by 65
“A preferred total allowable analytical error (TEa) in routine laboratory medicine can be calculated from the following formulas: TEa = Bias + 1.65 × imprecision TEa = Bias + 1.65 × SD Unauthentifiziert | Heruntergeladen 10.02.20 01:25 UTC If the value is Z = 1.65, one-sided data distribution with a CI of 95% is observ…”
BackgroundInfluentialUncertainty in Measurement and Total Error
Theodorsson, E. (2017) · Clinics in Laboratory Medicine · doi:10.1016/j.cll.2016.09.002 · Cited by 54
“However, methods for calculating measurement uncertainty are not prescribed in the standards so accreditation authorities accept both error and uncertainty approaches.(47) Top-downerrormethods for calculatingmeasurement uncertainty from internal quality control samplesareprobably themostcommonlyusedapproaches for calc…”
MethodISO 15189 Accreditation: Navigation Between Quality Management and Patient Safety
Plebani, M., & Sciacovelli, L. (2017) · Journal of Medical Biochemistry · doi:10.1515/jomb-2017-0038 · Cited by 44
Padoan, A., Antonelli, G., Aita, A., Sciacovelli, L., & Plebani, M. (2017) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2016-0896 · Cited by 41
“Depending on the test purpose, it might also be important to consider different models for calculating MU [4, 9, 10].”
InfluentialWestgard, J. O., & Westgard, S. A. (2016) · Clinics in Laboratory Medicine · doi:10.1016/j.cll.2016.09.001 · Cited by 39
Martinello, F., Snoj, N., Skitek, M., & Jerin, A. (2020) · Biochemia Medica · doi:10.11613/bm.2020.020101 · Cited by 34
“Other authors even suggest that harmonization of methods could minimize the bias and, essentially, make both total error and MU a result of imprecision (5,6).”
BackgroundInfluentialTran, M. T. C., Hoang, K., & Greaves, R. F. (2016) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2016.08.008 · Cited by 33
Farrance, I., Badrick, T. C., & Frenkel, R. B. (2018) · Practical Laboratory Medicine · doi:10.1016/j.plabm.2018.e00097 · Cited by 27
“When using either the total error approach to medical laboratory quality specifications [31,32], or uncertainty in measurement procedures based on the Guide to the Expression of Uncertainty in Measurement (usually referred to as the GUM) [33], the analytical tolerance with regard to both random result variation (impre…”
MethodSelf-citationSources of uncertainty in annual global horizontal irradiance data
Urraca-Valle, R., Huld, T., Martínez-de-Pisón, F. J., & Sanz-García, A. (2018) · Solar Energy · doi:10.1016/j.solener.2018.06.005 · Cited by 26
“GUM analyzes random and systematic errors separately (Farrance et al., 2016), and it calculates the uncertainty (u) only from random errors because it considers that systematic errors can be detected and corrected.”
BackgroundQuality in laboratory medicine and the Journal: walking together
Plebani, M. (2022) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2022-0755 · Cited by 20
Frenkel, R. B., Farrance, I., & Badrick, T. C. (2019) · Clinica Chimica Acta · doi:10.1016/j.cca.2019.03.1633 · Cited by 20
Self-citationCoşkun, A. (2025) · Annals of Laboratory Medicine · doi:10.3343/alm.2024.0569 · Cited by 14
Uncertainty in measurement and total error: different roads to the same quality destination?
Farrance, I., Badrick, T. C., & Frenkel, R. B. (2018) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2018-0421 · Cited by 14
Self-citationMeasurement uncertainty – a revised understanding of its calculation and use
Tate, J. R., & Plebani, M. (2016) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2016-0327 · Cited by 13
Extra-analytical sources of uncertainty: which ones really matter?
Padoan, A., Sciacovelli, L., Zhou, R., & Plebani, M. (2019) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2019-0197 · Cited by 12
Frenkel, R. B., & Farrance, I. (2018) · Advances in clinical chemistry · doi:10.1016/bs.acc.2018.02.003 · Cited by 11
Self-citationSánchez, J. M. (2024) · Journal of Chemical Education · doi:10.1021/acs.jchemed.4c00583 · Cited by 10
Frenkel, R. B., & Farrance, I. (2017) · Accreditation and Quality Assurance · doi:10.1007/s00769-017-1274-8 · Cited by 7
“However, the use of MU in this environment, with ‘separate procedures for the assessment of imprecision and bias’ has been defended as ‘simple and inherently logical’ [13].”
MethodSelf-citationMilinković, N., Jovičić, S. Ž., & Ignjatović, S. (2020) · Critical Reviews in Clinical Laboratory Sciences · doi:10.1080/10408363.2020.1784838 · Cited by 7
“required by ISO 15189 is to document (and understand) your chosen process” [24].”
BackgroundAyan, N. N., Çat, A., & Uçar, K. T. (2024) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365513.2024.2338738 · Cited by 5
“The permissible targets of BV, often preferred for comparing MU results, are quite narrow [10, 20, 28, 29].”
BackgroundGalindo-Méndez, M., Sánchez-López, A., & Cruz-Fuentes, L. (2018) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2018.10.012 · Cited by 4
Balance of the unsuccessful systematization of measurement uncertainty in medical laboratories
Pereira, P., & Seghatchian, J. (2017) · Transfusion and Apheresis Science · doi:10.1016/j.transci.2017.03.002 · Cited by 4
Reporting measurement uncertainties with ethanol results
Erdoğan, S., Sen, O., Bal, C., Öztaş, D., Avcıoğlu, G., Baran, P., & Erel, Ö. (2019) · Turkish Journal of Biochemistry · doi:10.1515/tjb-2018-0508 · Cited by 2
“The laboratory shall define the performance requirements for the measurement uncertainty of each measurement procedure and regularly review estimates of measurement uncertainty [8].”
BackgroundKwon, Y.-I. (2020) · Korean Journal of Clinical Laboratory Science · doi:10.15324/kjcls.2020.52.1.28
“진단검사에서 불확도 추정을 위한 단순 원리는 Monte Carlo 기법[15, 16]을 적용할 수 있는데 이때 검토되는 변이는 생물학적 변이, 분석 전 변이, 분석 변이, 분석 후 변이이다[13].”
Chawla, R., Subberwal, M., & Singhal, A. (2021) · Indian Journal of Clinical Biochemistry · doi:10.1007/s12291-021-01016-6
“As methods for calculating measurement uncertainty are not prescribed in the standards so accreditation authorities accept both error and uncertainty approaches [9].”
MethodDebate on Measurement Uncertainty Verses Total Error: A Take by Tertiary Care Hospital Laboratory
Bansal, A., & Agarwal, R. (2026) · Indian Journal of Clinical Biochemistry · doi:10.1007/s12291-026-01391-y
Elsharkawi, S. M. (2026) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2026.111208
Metadata
- Authors
- Farrance, I., Badrick, T., & Sikaris, K. A.
- Affiliations
- RMIT University (AU)
- Royal College of Pathologists of Australasia (AU)
- Corresponding author
- Farrance, Ian
- Venue
- Clinical Chemistry and Laboratory Medicine (CCLM)
- Publisher
- Walter de Gruyter GmbH
- Published
- 2016-01-01
- Volume
- 54
- Issue
- 8
- Pages
- 1309-11
- ISSN
- 1434-6621 (ISSN-L)
- 1437-4331
- OpenAlex
- W2404333965
- Cited by
- 31
- References
- 7
- Citations by year
- 2026: 2 · 2025: 1 · 2024: 2 · 2022: 1 · 2021: 1 · 2020: 4 · 2019: 4 · 2018: 7 · 2017: 4 · 2016: 5
- FWCI
- 5.37
- Type
- journal article
- Language
- en
- Open access
- green
- hdl.handle.net
- Licence
- other-oa
- Locations
- Clinical Chemistry and Laboratory Medicine (CCLM) (publishedVersion)
- PubMed (publishedVersion)
- RMIT Research Repository (RMIT University Library) (submittedVersion)
- Figshare (submittedVersion)
- Griffith Research Online (submittedVersion, other-oa)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Physics and Astronomy › Instrumentation › Scientific Measurement and Uncertainty Evaluation
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Healthcare Quality and Management
- Keywords
- measurement uncertainty, total error, Westgard rules, Guide to the Expression of Uncertainty in Measurement, ISO 15189, laboratory quality control, clinical laboratory testing
- MeSH terms
- Humans, Quality Control, Internet, Uncertainty
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.