The Influence of Bilirubin, Haemolysis and Turbidity on 20 Analytical Tests Performed on Automatic Analysers
Open accessThe DOI matches its Crossref record.
Abstract
The director of a laboratory has to be sure to give out reliable results for routine tests on automatic analysers regardless of the clinical context. However, he may find hyperbilirubinaemia in some circumstances, parenteral nutrition causing turbidity in others, and haemolysis occurring if sampling is difficult. For this reason, the Commission for Instrumentation of the Société Française de Biologie Clinique (SFBC) (president Alain Feuillu) decided to look into "visible" interferences--bilirubin, haemolysis and turbidity--and their effect on 20 major tests: 13 substrates/chemistries: albumin, calcium, cholesterol, creatinine, glucose, iron, magnesium, phosphorus, total bilirubin, total proteins, triacylglycerols, uric acid, urea, and 7 enzymatic activities: alkaline phosphatase, alanine aminotransferase, alpha-amylase, aspartate aminotransferase, creatine kinase, gamma-glutamyl transferase and lactate dehydrogenase measured on 15 automatic analysers representative of those found on the French market (Astra 8, AU 510, AU 5010, AU 5000, Chem 1, CX 7, Dax 72, Dimension, Ektachem, Hitachi 717, Hitachi 737, Hitachi 747, Monarch, Open 30, Paramax, Wako 30 R) and to see how much they affect the accuracy of results under routine conditions in the laboratory. The study was carried out following the SFBC protocol for the validation of techniques using spiked plasma pools with bilirubin, ditauro-bilirubin, haemoglobin (from haemolysate) and Intralipid (turbidity). Overall, the following results were obtained: haemolysis affects tests the most often (34.5% of cases); total bilirubin interferes in 21.7% of cases; direct bilirubin and turbidity seem to interfere less at around 17%. The different tests are not affected to the same extent; enzyme activity is hardly affected at all; on the other hand certain major tests are extremely sensitive, increasingly so as we go through the following: creatinine (interference of bilirubin), triacylglycerols (interference of bilirubin and haemoglobin), glucose (interference of bilirubin), cholesterol (interference of bilirubin), phosphorus (interference of bilirubin and haemoglobin), uric acid (interference of turbidity), iron (interference of haemoglobin and turbidity), total proteins (interference of bilirubin, haemoglobin and turbidity) and bilirubin (interference of haemoglobin and turbidity). Three categories of interferences can be found: interference by addition, chemical interference and spectral interference, and the results show that not only the choice of a method is important on the analyser, but also how this has been adapted. By looking into these conditions carefully, it is sometimes possible to find a reason for the problem and thereby a simple solution to correct it.(ABSTRACT TRUNCATED AT 400 WORDS)
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Haemolysis: an overview of the leading cause of unsuitable specimens in clinical laboratories
Lippi, G., Blanckaert, N., Bonini, P., Green, S., Kitchen, S., Palicka, V., Vassault, A. J., & Plebani, M. (2008) · Clinical Chemistry and Laboratory Medicine · doi:10.1515/cclm.2008.170 · Cited by 466
Influence of hemolysis on routine clinical chemistry testing
Lippi, G., Luca Salvagno, G., Montagnana, M., Brocco, G., & Cesare Guidi, G. (2006) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2006.054 · Cited by 344
“The interference of in vitro hemolysis on laboratory testing might be caused by: (i) leakage of hemoglobin and other intracellular components into the surrounding fluid, which induces false elevations of some analytes or dilution effects; (ii) chemical interference by free hemoglobin in the analytical reaction; and (i…”
BackgroundInfluentialNarayanan, S. (2000) · American Journal of Clinical Pathology · doi:10.1309/c0nm-q7r0-ll2e-b3uy · Cited by 198
Woolley, A., Golmard, J. ‐., & Kitchen, S. (2016) · International Journal of Laboratory Hematology · doi:10.1111/ijlh.12498 · Cited by 47
Blood Glucose Determination: Effect of Tube Additives
(2018) · Advances in Clinical Chemistry · doi:10.1016/bs.acc.2017.12.003 · Cited by 28
Metadata
- Authors
- Grafmeyer, D., Bondon, M., Manchon, M., & Levillain, P.
- Venue
- cclm
- Publisher
- Walter de Gruyter GmbH
- Published
- 1995
- Volume
- 33
- Issue
- 1
- Pages
- 31-52
- ISSN
- 1434-6621 (ISSN-L)
- 1437-4331
- OpenAlex
- W2022273366
- Cited by
- 59
- References
- 0
- FWCI
- 1.06
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.