Wireless communication with implanted medical devices using the conductive properties of the body

Open access
Ferguson, John E, Redish, A David·2011·Expert Review of Medical Devices

Abstract

Many medical devices that are implanted in the body use wires or wireless radiofrequency telemetry to communicate with circuitry outside the body. However, the wires are a common source of surgical complications, including breakage, infection and electrical noise. In addition, radiofrequency telemetry requires large amounts of power and results in low-efficiency transmission through biological tissue. As an alternative, the conductive properties of the body can be used to enable wireless communication with implanted devices. In this article, several methods of intrabody communication are described and compared. In addition to reducing the complications that occur with current implantable medical devices, intrabody communication can enable novel types of miniature devices for research and clinical applications.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1586/erd.11.16

Cites

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • An external "reader coil" generates the electromagnetic field, intercepted by a corresponding coil in the implant; the received signal is demodulated to separate power and data.

    “An external “reader coil” generates the electromagnetic field, intercepted by a corresponding coil in the implant. The received signal is demodulated to separate power and data, effectively fueling and interfacing with the neural implant (Ferguson and Redish, 2011; Pavone et al., 2012).”

    From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Ferguson and Redish, 2011

    The Challenges > Power section (page unavailable) discusses wireless RF power delivery requiring aligned antennas and passive implants responding to an external signal. It does not describe an external reader coil coupling to an implant coil or demodulation that separates power from data. (LLM, conf=0.82)

    Experimental — accuracy not yet measured
  • One primary constraint in distance between transmitter and receiver in RF charging systems is the inherent limitation on the size of the receiver's antenna and wavelength of the electromagnetic waves used, which influences the effectiveness of energy transfer.

    “One of the primary constraints in distance between the transmitter and receiver in Radio Frequency (RF) charging systems is the inherent limitation on the size of the receiver’s antenna and wavelength of the electromagnetic waves used which influences the effectiveness of energy transfer (Ferguson and Redish, 2011).”

    From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Ferguson and Redish, 2011

    Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.

    Experimental — accuracy not yet measured
  • Brain implants require minimization of the system to reduce invasiveness, which might conflict with the ideal antenna size needed for efficient RF energy transfer, leading to a trade-off between the physical size of the implant and the efficiency of power reception.

    “However brain implants require minimization of the system to reduce the invasiveness of the procedure which might conflict with the ideal antenna size needed for efficient RF energy transfer, leading to a trade-off between the physical size of the implant and the efficiency of power reception (Ferguson and Redish, 2011; Di Carlo et al., 2018; Fan et al., 2022).”

    From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Ferguson and Redish, 2011

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

The citation lists haven’t been fetched yet

References 42 · Cited by 112

Research works printed but not linked: 36

  1. [2]Moritz CT Perlmutter SI Fetz EE Direct control of paralysed muscles by cortical neurons Nature 456 7222 639 642 2008 18923392 PMC3159518
  2. [3]Benabid AL Chabardes S Mitrofanis J Pollak P Deep brain stimulation of the subthalamic nucleus for the treatment of Parkinson’s disease Lancet Neurol 8 1 67 81 2009 19081516
  3. [4]Hauser RG Katsiyiannis WT Gornick CC Almquist AK Kallinen LM Deaths and cardiovascular injuries due to device-assisted implantable cardioverter-defibrillator and pacemaker lead extraction Europace 12 3 395 401 2010 19946113 PMC2825385
  4. [5]Loeb GE Richmond FJR Baker LL The BION devices: injectable interfaces with peripheral nerves and muscles Neurosurg Focus 20 5 1 9 2006 16711659
  5. [6]Orlov MV Szombathy T Chaudhry GM Haffajee CI Remote surveillance of implantable cardiac devices PACE 32 7 928 939 2009 19572872
  6. [7]Murakawa K Kobayashi M Nakamura O Kawata S A wireless near-infrared energy system for medical implants IEEE Eng Med Biol 18 6 70 72 1999 10576076
  7. [8]Echt DS Cowan MW Riley RE Brisken AF Feasibility and safety of a novel technology for pacing without leads Heart Rhythm 3 10 1202 1206 2006 17018352
  8. [9]Zimmerman TG Personal area networks (PAN): near-field intra-body communication PhD thesis Massachusetts Institute of Technology MA, USA 1995
  9. [10]Meoli D May-Plumlee T Interactive electronic textile development: a review of technologies J Text Apparel Technol Manag 2 2 1 12 2002
  10. [11]Hyoung CH Sung JB Hwang JH Kim JK Park DG Kang SW A novel system for intrabody communication: touch-and-play Proceedings of the IEEE International Symposium on Circuits and Systems 1343 1346 2006
  11. Lucev V Krois I Cifrek M Intrabody communication in biotelemetry Wearable and Autonomous Biomedical Devices and Systems for Smart Environment Springer Berlin, Germany 351 368 2010 Review of capacitive intrabody communication
  12. Lindsey DP McKee EL Hull ML Howell SM A new technique for transmission of signals from implantable transducers IEEE Trans Biomed Eng 45 5 614 619 1998 First report of galvanic intrabody communication from an implanted device 9581060
  13. Sun M Mickle M Liang W Liu Q Sclabassi RJ Data communication between brain implants and computer IEEE Trans Neural Syst Rehabil Eng 11 2 189 192 2003 Demonstration of how changing the current flow can improve the efficiency of intrabody communication 12899271
  14. Wegmueller MS Huclova S Froehlich J Galvanic coupling enabling wireless implant communications IEEE Trans Instrum Meas 58 8 2618 2625 2009 Report of galvanic intrabody communication between implants
  15. Al-Ashmouny KM Boldt C Ferguson JE Erdman AG Redish AD Yoon E IBCOM (intra-brain communication) microsystem: wireless transmission of neural signals within the brain Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society 2054 2057 2009 First report of an in vivo test of galvanic communication through the brain 19964776
  16. [18]Song Y Hao Q Zhang K Wang M Chu Y Kang B The simulation method of the galvanic coupling intrabody communication with different signal transmission paths IEEE Trans Instrum Meas 1 10 2010
  17. [19]Wegmueller MS Oberle M Felber N Kuster N Fichtner W Signal transmission by galvanic coupling through the human body IEEE Trans Instrum Meas 59 4 963 969 2010
  18. [20]Pun SH Gao YM Mou Pa Mak PU Vai MI Du M Multilayer limb quasi-static electromagnetic modeling with experiments for galvanic coupling type IBC Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society 1 378 381 2010 21097189
  19. Sarpeshkar R Ultra Low Power Bioelectronics: Fundamentals, Biomedical Applications, and Bio-Inspired Systems Cambridge University Press Cambridge, UK 2010 Comprehensive textbook covering low-power electronics, biotelemetry and miniature circuit design
  20. [22]Kurs A Karalis A Moffatt R Joannopoulos JD Fisher P Soljacic M Wireless power transfer via strongly coupled magnetic resonances Science 317 5834 83 86 2007 17556549
  21. [23]Zhang F Liu X Hackworth SA Sclabassi RJ Sun M In vitro and in vivo studies on wireless powering of medical sensors and implantable devices Proceedings of the IEEE/NIH Life Science Systems and Applications Workshop 84 87 2009
  22. [24]Lee KL Lau CP Tse HF First human demonstration of cardiac stimulation with transcutaneous ultrasound energy delivery: implications for wireless pacing with implantable devices J Am Coll Cardiol 50 9 877 883 2007 17719475
  23. [25]Justin GA Zhang Y Sun M Sclabassi R Biofuel cells: a possible power source for implantable electronic devices Proceedings of the 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society 2 4096 4099 2004 17271201
  24. [26]Towe BC Passive backscatter biotelemetry for neural interfacing Proceedings of the 3rd International IEEE/EMBS Conference on Neural Engineering 144 147 2007
  25. [28]Papale A Ferguson JE Boldt C Redish AD Steps toward decoupling ensemble recording sites from the headstage Presented at: Society for Neuroscience Annual Conference CA, USA 13–17 November 2010
  26. [29]Troyk PR Injectable electronic identification, monitoring, and stimulation systems Annu Rev Biomed Eng 1 177 209 1999 11701487
  27. [30]Grady MS Howard MA Dacey RG Experimental study of the magnetic stereotaxis system for catheter manipulation within the brain J Neurosurg 93 2 282 288 2000 10930015
  28. [32]Kim DH Viventi J Amsden JJ Dissolvable films of silk fibroin for ultrathin conformal bio-integrated electronics Nat Mater 9 6 511 517 2010 20400953 PMC3034223
  29. [33]Turner JN Shain W Szarowski DH Cerebral astrocyte response to micromachined silicon implants Exp Neurol 156 1 33 49 1999 10192775
  30. [34]Polikov VS Block ML Fellous JM Hong JS Reichert WM In vitro model of glial scarring around neuroelectrodes chronically implanted in the CNS Biomaterials 27 31 5368 5376 2006 16842846
  31. [35]Williams DF On the mechanisms of biocompatibility Biomaterials 29 20 2941 2953 2008 18440630
  32. [36]Onuki Y Bhardwaj U Papadimitrakopoulos F Burgess DJ A review of the biocompatibility of implantable devices: current challenges to overcome foreign body response J Diabetes Sci Technol 2 6 1003 1015 2008 19885290 PMC2769826
  33. [37]Gabriel C Gabriel S Corthout E The dielectric properties of biological tissues: I. Literature survey Phys Med Biol 41 11 2231 2249 1996 8938024
  34. [38]Gabriel S Lau RW Gabriel C The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHz Phys Med Biol 41 11 2251 2269 1996 8938025
  35. [39]Gabriel S Lau RW Gabriel C The dielectric properties of biological tissues: III. Parametric models for the dielectric spectrum of tissues Phys Med Biol 41 11 2271 2293 1996 8938026
  36. [40]ICNIRP Guidelines for limiting exposure to time-varying electric, magnetic, and electromagnetic fields (up to 300 GHz) Health Phys 74 4 494 522 1998 9525427

Authors
Ferguson, J. E., & Redish, A. D.
Affiliations
University of Minnesota (US)
Venue
Expert Review of Medical Devices
Published
2011-07
Volume
8
Issue
4
Pages
427-433
ISSN
  • 1743-4440 (ISSN-L)
  • 1745-2422
OpenAlex
W2111908558
Cited by
112
References
42
Citations by year
2026: 6 · 2025: 5 · 2024: 7 · 2023: 10 · 2022: 9 · 2021: 11 · 2020: 11 · 2019: 7 · 2018: 12 · 2017: 11 · 2016: 7 · 2015: 5 · 2014: 2 · 2013: 5 · 2012: 4
FWCI
0.28
Type
journal article
Language
en
Open access
Locations
  • Expert Review of Medical Devices (publishedVersion)
  • PubMed (publishedVersion)
  • Europe PMC (PubMed Central) (acceptedVersion)
  • PubMed Central (acceptedVersion)
Funders
National Institute of Biomedical Imaging and Bioengineering (T32EB008389)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
  • Physical Sciences › Engineering › Biomedical Engineering › Molecular Communication and Nanonetworks
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Wireless Power Transfer Systems
Keywords
intra-body communication, biomedical implants, wireless communication, surgical complications
MeSH terms
Electric Conductivity, Equipment and Supplies, Humans, Surface Properties, Telemetry, Prostheses and Implants, Wireless Technology

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review