FACTORS AFFECTING THE DURABILITY OF PRIMARY TOTAL KNEE PROSTHESES
Abstract
BACKGROUND: Failure of total knee arthroplasty is problematic. The purpose of this study was to evaluate the factors that influence the durability of a primary total knee prosthesis. METHODS: A survivorship analysis of 11,606 primary total knee arthroplasties carried out between January 1, 1978, and December 31, 2000, was performed. An analysis of patient and implant-related factors affecting survivorship was done with use of a multivariate Cox model. RESULTS: The survivorship was 91% (95% confidence interval, 90% to 91%) at ten years (2943 knees), 84% (95% confidence interval, 82% to 86%) at fifteen years (595 knees), and 78% (95% confidence interval, 74% to 81%) at twenty years (104 knees) following the surgery. Prosthetic survivorship at ten years was 83% for patients fifty-five years of age or less compared with 94% for those older than seventy years of age (p < 0.0001), 90% for those with a diagnosis of osteoarthritis compared with 95% for those with inflammatory arthritis (p < 0.005), and 91% for those with retention of the posterior cruciate ligament compared with 76% for those with substitution of the posterior cruciate ligament (a posterior stabilized prosthesis) (p < 0.0001). Survivorship at ten years was 92% for nonmodular metal-backed tibial components, 90% for modular metal-backed tibial components, and 97% for all-polyethylene tibial components (p < 0.0001). Survivorship at ten years was 92% for prostheses fixed with cement compared with 61% for those fixed without cement (p < 0.0001). CONCLUSIONS: Significant risk factors for failure of total knee arthroplasty were the type of implant, age and gender of the patient, diagnosis, type of fixation, and design of the patellar component. In the ideal situation-treatment of a woman over the age of seventy years who has inflammatory arthritis with a nonmodular, metal-backed tibial component, cement fixation, an all-polyethylene patellar component, and retention of the posterior cruciate ligament-the ten-year survivorship of the prosthesis was estimated to be 98% (95% confidence interval, 97% to 99%).
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Nonparametric Estimation from Incomplete Observations
Kaplan, E. L., & Meier, P. (1958) · Journal of the American Statistical Association · doi:10.1080/01621459.1958.10501452 · Cited by 39,158
Cited by
Zhang, W., Moskowitz, R. W., Nuki, G., Abramson, S., Altman, R. D., Arden, N., Bierma-Zeinstra, S., Brandt, K. D., Croft, P., Doherty, M., et al. (2008) · Osteoarthritis and Cartilage · doi:10.1016/j.joca.2007.12.013 · Cited by 2,796
Epidemiology of Knee and Hip Arthroplasty: A Systematic Review§
Singh, J. A. (2011) · The Open Orthopaedics Journal · doi:10.2174/1874325001105010080 · Cited by 347
Singh, J. A., & Lewallen, D. G. (2012) · Arthritis Research & Therapy · doi:10.1186/ar4091 · Cited by 84
Metadata
- Authors
- RAND, J. A., TROUSDALE, R. T., ILSTRUP, D. M., & HARMSEN, W. S.
- Venue
- The Journal of Bone and Joint Surgery-American Volume
- Published
- 2003-02
- Volume
- 85
- Issue
- 2
- Pages
- 259-265
- ISSN
- 0021-9355 (ISSN-L)
- OpenAlex
- W113982794
- Cited by
- 471
- References
- 1
- FWCI
- 15.04
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: FACTORS AFFECTING THE DURABILITY OF PRIMARY TOTAL KNEE PROSTHESES. https://app.cochat.ai/papers/10.2106/00004623-200302000-00012
- The paper
- RAND, J. A., TROUSDALE, R. T., ILSTRUP, D. M., & HARMSEN, W. S. (2003). FACTORS AFFECTING THE DURABILITY OF PRIMARY TOTAL KNEE PROSTHESES. The Journal of Bone and Joint Surgery-American Volume. https://doi.org/10.2106/00004623-200302000-00012
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.