Accuracy of 25-Hydroxyvitamin D Assays: Confronting the Issues
Abstract
Measurement of 25-hydroxyvitamin D (25-OHD) is widely used for assessing vitamin D status. There has been a dramatic increase in 25-OHD requests over recent years prompting many laboratories to consider the use of automated immunoassays. To achieve higher throughput, these methods have abandoned the traditional solvent extraction of samples and are therefore more prone to non-specific interference. The Vitamin D External Quality Assessment Scheme (DEQAS) has revealed method-related differences in 25-OHD results, raising concerns about the comparability and accuracy of different assays. This paper highlights some of the pre-analytical, analytical and post-analytical issues which may influence the accuracy of 25-OHD assays and interpretation of results. Recent attention has focused on reconciling the relatively high results given by liquid chromatography-tandem mass spectrometry (LC-MS/MS) to those of the DiaSorin radioimmunoassay (RIA) on which clinical decision points have previously been based. Data is presented on 20 DEQAS samples which were analysed by an LC-MS/MS assay developed as a candidate reference measurement procedure by the US National Institute of Standards and Technology (NIST). The NIST results were on average 11.2% lower than those given by routine LC-MS/MS methods. If confirmed, these results suggest that most routine LC-MS/MS assays are perhaps overestimating 25-OHD by failing to resolve a molecule having the same mass as 25-OHD3 and a similar fragmentation pattern. All 25-OHD assays should be monitored by a proficiency testing scheme and the results made available to clinicians and editors of scientific journals. Keywords: 25-Hydroxyvitamin D, accuracy, methods, DEQAS, Vitamin D External Quality Assessment Scheme, liquid chromatography-tandem mass spectrometry (LC-MS/MS), radioimmunoassay, NIST, 7-dehydrocholesterol (provitamin D), D binding protein, 25-OHD3, proficiency-testing scheme, competitive protein-binding (CPB) assay, DiaSorin, UV spectrophotometry, enzyme immunoassay (EIA), mass spectrometry, GC-MS, ESA, RMP, liquid chromatography
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Global Consensus Recommendations on Prevention and Management of Nutritional Rickets
Munns, C. F., Shaw, N., Kiely, M., Specker, B. L., Thacher, T. D., Ozono, K., Michigami, T., Tiosano, D., Mughal, M. Z., Mäkitie, O., et al. (2016) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2015-2175 · Cited by 1,150
A systematic review of vitamin D status in populations worldwide
Hilger, J., Friedel, A., Herr, R., Rausch, T., Roos, F., Wahl, D. A., Pierroz, D. D., Weber, P., & Hoffmann, K. (2014) · British Journal of Nutrition · doi:10.1017/s0007114513001840 · Cited by 832
Food allergy as a risk factor for nutritional rickets
Fox, A. T., Du Toit, G., Lang, A., & Lack, G. (2004) · Pediatric Allergy and Immunology · doi:10.1111/j.1399-3038.2004.00158.x · Cited by 576
Global Consensus Recommendations on Prevention and Management of Nutritional Rickets
Munns, C. F., Shaw, N., Kiely, M., Specker, B. L., Thacher, T. D., Ozono, K., Michigami, T., Tiosano, D., Mughal, M. Z., Mäkitie, O., et al. (2016) · Hormone Research in Paediatrics · doi:10.1159/000443136 · Cited by 349
Dietary reference values for vitamin D
EFSA Panel on Dietetic Products, N. A. A. (. (2016) · EFSA Journal · doi:10.2903/j.efsa.2016.4547 · Cited by 304
Metadata
- Authors
- D. Carter, G.
- Venue
- Current Drug Targets
- Publisher
- Bentham Science Publishers Ltd.
- Published
- 2011-01-01
- Volume
- 12
- Issue
- 1
- Pages
- 19-28
- ISSN
- 1389-4501 (ISSN-L)
- OpenAlex
- W2037715054
- Cited by
- 221
- References
- 0
- FWCI
- 11.41
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Accuracy of 25-Hydroxyvitamin D Assays: Confronting the Issues. https://app.cochat.ai/papers/10.2174/138945011793591608
- The paper
- D. Carter, G. (2011). Accuracy of 25-Hydroxyvitamin D Assays: Confronting the Issues. Current Drug Targets. https://doi.org/10.2174/138945011793591608
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.