Prospective Assessment of Computer-Aided Detection in Interpretation of Screening Mammography

Ko, Justin M., Nicholas, Michael J., Mendel, Jeffrey B., Slanetz, Priscilla J. · 2006 · American Journal of Roentgenology
Cited by 121

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

OBJECTIVE: The purpose of this study was to prospectively assess the usefulness of computer-aided detection (CAD) in the interpretation of screening mammography and to provide the true sensitivity and specificity of this technique in a clinical setting. SUBJECTS AND METHODS: Over a 26-month period, 5,016 screening mammograms were interpreted without, and subsequently with, the assistance of the iCAD MammoReader detection system. Data collected for actionable findings included dominant feature (calcification, mass, asymmetry, architectural distortion), detection method (radiologist only, CAD only, or both radiologist and CAD), BI-RADS assessment code, associated histopathology for those undergoing biopsy, and tumor stage for malignant lesions. The study population was cross-checked against an independent reference standard to identify false-negative cases. RESULTS: Of the 5,016 cases, the recall rate increased from 12% to 14% with the addition of CAD. Of the 107 (2%) patients who underwent biopsy, 101 (94%) were prompted by the radiologist and six (6%) were prompted by CAD. Of the 124 biopsies performed on actionable findings in the 107 patients, findings in 79 (64%) were benign and in 45 (36%) were in situ or invasive carcinoma. Three study participants who were not recalled by the radiologist with the assistance of CAD developed cancer within 1 year of the screening mammogram and were considered to be false-negative cases. The radiologist detected 43 (90%) of the 48 total malignancies and 45 (94%) of the 48 malignancies with the assistance of CAD. CAD missed eight cancers that were detected by the radiologist, which presented as architectural distortions (n = 3), irregular masses (n = 4), and a circumscribed mass (n = 1). CAD detected two in situ cancers as a faint cluster of calcifications that had not been perceived by the radiologist and one mass that was dismissed by the radiologist, accounting for at least a 4.7% increase in cancer detection rate. Sensitivity of screening mammography with the use of CAD (94%) represented an absolute and relative 4% increase over the sensitivity of the radiologist alone (90%). Specificity of screening mammography with and without the use of CAD was 99%. CONCLUSION: Routine use of CAD while interpreting screening mammograms significantly increases recall rates, has no significant effect on positive predictive value for biopsy, and can increase cancer detection rate by at least 4.7% and sensitivity by at least 4%. This study provides "true" values for sensitivity and specificity for use of CAD in interpretation of screening mammography as measured prospectively in the context of a working clinical setting.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.2214/ajr.05.1582

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 1 references

Cited by

See all 121

Metadata

Authors
Ko, J. M., Nicholas, M. J., Mendel, J. B., & Slanetz, P. J.
Venue
American Journal of Roentgenology
Published
2006-12
Volume
187
Issue
6
Pages
1483-1491
ISSN
  • 0002-9580 (ISSN-L)
  • 0361-803X
  • 1546-3141
OpenAlex
W2029106704
Cited by
121
References
1
FWCI
11.81
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Prospective Assessment of Computer-Aided Detection in Interpretation of Screening Mammography. https://app.cochat.ai/papers/10.2214/ajr.05.1582
The paper
Ko, J. M., Nicholas, M. J., Mendel, J. B., & Slanetz, P. J. (2006). Prospective Assessment of Computer-Aided Detection in Interpretation of Screening Mammography. American Journal of Roentgenology. https://doi.org/10.2214/ajr.05.1582

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.