Bone Mineral Density Assessment Using Spectral Topograms From a Clinical Photon-Counting Detector CT System: A Phantom Evaluation
Abstract
Bone mineral density assessments in a spine phantom had lower errors by spectral topograms from a clinical photon-counting detector CT scanner than by a DXA system. Bone mineral density assessments by PCD CT also had high reproducibility.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
A previous experimental study with spectral localizer radiographs from PCD-CT using the European Spine Phantom showed that spectral localizer radiographs were more accurate than DXA at higher bone density.
“A previous experimental study with spectral localizer radiographs from PCD-CT using the established European Spine Phantom21 showed that spectral localizer radiographs were more accurate that DXA at higher bone density.13”
From In Vivo Bone Mineral Density Assessment With Spectral Localizer Radiographs From Photon-Counting Detector CT (2025) · cites this paper as [13]
Two phantom studies using a clinically approved PCD-CT scanner confirmed these initial observations and showed accurate results in regard to aBMD quantification as compared with phantom specifications and with the clinical reference standard DXA.
“More recently, 2 phantom studies using a clinically approved PCD-CT scanner confirmed these initial observations and showed accurate results in regard to aBMD quantification as compared with phantom specifications and with the clinical reference standard DXA.12,13”
From In Vivo Bone Mineral Density Assessment With Spectral Localizer Radiographs From Photon-Counting Detector CT (2025) · cites this paper as [13]
Cited by
Sorted by citations
Moser, L. J., Klambauer, K., Diaz Machicado, M. C., Frey, D., Mergen, V., Eberhard, M., Nowak, T., Schmidt, B., Flohr, T., Distler, O., et al. (2025) · Investigative Radiology · doi:10.1097/rli.0000000000001159 · Cited by 8 · see citing works
Self-citation
- Authors
- Moser, L. J., Pitteloud, J., Mergen, V., Frey, D., Nowak, T., Distler, O., Eberhard, M., & Alkadhi, H.
- Affiliations
- University Hospital Zurich (CH)
- Siemens Healthineers (Germany) (DE)
- ORCID
- Venue
- American Journal of Roentgenology
- Publisher
- American Roentgen Ray Society
- Published
- 2024-02
- Volume
- 222
- Issue
- 2
- Pages
- e2330347-e2330347
- ISSN
- 0002-9580 (ISSN-L)
- 0361-803X
- 1546-3141
- OpenAlex
- W4388491191
- Cited by
- 5
- References
- 0
- Citations by year
- 2025: 2 · 2024: 3
- FWCI
- 0.39
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- American Journal of Roentgenology (publishedVersion)
- PubMed (publishedVersion)
- Zurich Open Repository and Archive (University of Zurich) (submittedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced X-ray and CT Imaging
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Bone health and osteoporosis research
- Physical Sciences › Engineering › Biomedical Engineering › Medical Imaging and Analysis
- Keywords
- bone mineral density, photon-counting CT, spine phantom, dual-energy X-ray absorptiometry comparison, measurement reproducibility, bone densitometry
- MeSH terms
- Humans, Tomography, X-Ray Computed, Bone Density, Photons, Phantoms, Imaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review