Effect of Hematocrit Concentration on Blood Glucose Value Determined on Glucometer II

Barreau, Peter B, Buttery, Joseph E·1988·Diabetes Care

Abstract

Nineteen patient blood samples each with modified hematocrit concentrations of approximately 20, 30, 40, 50, and 60%, were assayed for their glucose concentration by the Glucometer II. Blood removal from the test strip was by the one- and two-blot techniques. The reference method was the Yellow Springs Instruments (YSI) blood glucose analyzer. Glucometer II results were falsely high for the single blot (13-59%, mean 33%) and double blot (12-41%, mean 19%) at 20% hematocrit and falsely low at 60% hematocrit for the single blot (22-44%, mean 37%) and the double blot (26-49%, mean 43%). At 40-50% hematocrit, Glucometer II and YSI results agreed only for the one-blot technique.

Overview

Integrity signals

Registry

Crossref · 10.2337/diacare.11.2.116

Cites

Statistics

Recomputing p-values needs the paper’s full text.

Verification

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Citations

The citation lists haven’t been fetched yet

Cited by 38

Metadata

Authors
Barreau, P. B., & Buttery, J. E.
Affiliations
Queen Elizabeth Hospital (AU)
Venue
Diabetes Care
Published
1988-02-01
Volume
11
Issue
2
Pages
116-118
ISSN
  • 0149-5992 (ISSN-L)
  • 1935-5548
OpenAlex
W1974416649
Cited by
38
References
0
Citations by year
2021: 1 · 2020: 1 · 2019: 1 · 2018: 1 · 2015: 2 · 2014: 1 · 2013: 2 · 2012: 3
FWCI
0.61
Type
journal article
Language
en
Open access
closed
Locations
  • Diabetes Care (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
Keywords
hematocrit, blood glucose measurement, blood glucose meters, blood sampling technique, point-of-care glucose testing
MeSH terms
Blood Glucose, Colorimetry, Diabetes Mellitus, False Negative Reactions, False Positive Reactions, Hematocrit, Humans, Reagent Strips

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Effect of Hematocrit Concentration on Blood Glucose Value Determined on Glucometer II. https://app.cochat.ai/papers/10.2337/diacare.11.2.116
The paper
Barreau, P. B., & Buttery, J. E. (1988). Effect of Hematocrit Concentration on Blood Glucose Value Determined on Glucometer II. Diabetes Care. https://doi.org/10.2337/diacare.11.2.116