Effect of Hematocrit Concentration on Blood Glucose Value Determined on Glucometer II
Abstract
Nineteen patient blood samples each with modified hematocrit concentrations of approximately 20, 30, 40, 50, and 60%, were assayed for their glucose concentration by the Glucometer II. Blood removal from the test strip was by the one- and two-blot techniques. The reference method was the Yellow Springs Instruments (YSI) blood glucose analyzer. Glucometer II results were falsely high for the single blot (13-59%, mean 33%) and double blot (12-41%, mean 19%) at 20% hematocrit and falsely low at 60% hematocrit for the single blot (22-44%, mean 37%) and the double blot (26-49%, mean 43%). At 40-50% hematocrit, Glucometer II and YSI results agreed only for the one-blot technique.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“In strip-based glucose assays, high hematocrit values, due to increased blood viscosity, can reduce the diffusion of plasma and determine underestimated results, whereas low hematocrit levels, like in anemia, are associated with overestimated readings 101,102.”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [101]
Citations
The citation lists haven’t been fetched yet
Cited by 38
Metadata
- Authors
- Barreau, P. B., & Buttery, J. E.
- Affiliations
- Queen Elizabeth Hospital (AU)
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 1988-02-01
- Volume
- 11
- Issue
- 2
- Pages
- 116-118
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W1974416649
- Cited by
- 38
- References
- 0
- Citations by year
- 2021: 1 · 2020: 1 · 2019: 1 · 2018: 1 · 2015: 2 · 2014: 1 · 2013: 2 · 2012: 3
- FWCI
- 0.61
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Diabetes Care (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Keywords
- hematocrit, blood glucose measurement, blood glucose meters, blood sampling technique, point-of-care glucose testing
- MeSH terms
- Blood Glucose, Colorimetry, Diabetes Mellitus, False Negative Reactions, False Positive Reactions, Hematocrit, Humans, Reagent Strips
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Effect of Hematocrit Concentration on Blood Glucose Value Determined on Glucometer II. https://app.cochat.ai/papers/10.2337/diacare.11.2.116
- The paper
- Barreau, P. B., & Buttery, J. E. (1988). Effect of Hematocrit Concentration on Blood Glucose Value Determined on Glucometer II. Diabetes Care. https://doi.org/10.2337/diacare.11.2.116