Accuracy of an Electrochemical Sensor for Measuring Capillary Blood Ketones by Fingerstick Samples During Metabolic Deterioration After Continuous Subcutaneous Insulin Infusion Interruption in Type 1 Diabetic Patients
The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
OBJECTIVE—This study was designed to test the accuracy of capillary ketonemia for diagnosis of ketosis after interruption of insulin infusion. RESEARCH DESIGN AND METHODS—A total of 18 patients with type 1 diabetes treated by external pump were studied during pump stop for 5 h. Plasma and capillary ketonemia and ketonuria were determined every hour from 7:00 a.m. (time 0 min = T0) to 12:00 p.m. (time 300 min = T300). Plasma β-hydroxybutyrate (β-OHB) levels were measured by an enzymatic end point spectrophotometric method, and capillary β-OHB levels were measured by an electrochemical method (MediSense Optium meter). Ketonuria was measured by a semiquantitative test (Ketodiastix). Positive ketosis was defined by a value of ≥0.5 mmol/l for ketonemia and ≥4 mmol/l (moderate) for ketonuria. RESULTS—After stopping the pump, concentrations of β-OHB in both plasma and capillary blood increased significantly at time 60 min (T60) compared with T0 (P < 0.001), reaching maximum levels at T300 (1.30 ± 0.49 and 1.23 ± 0.78 mmol/l, respectively). Plasma and capillary β-OHB values were highly correlated (r = 0.94, P < 0.0001). For diagnosis of ketosis, capillary ketonemia has a higher sensitivity and negative predictive value (80.4 and 82.5%, respectively) than ketonuria (63 and 71.8%, respectively). For plasma glucose levels ≥250 mg/dl, plasma and capillary ketonemia were found to be more frequently positive (85 and 78%, respectively) than ketonuria (59%) (P = 0.017). The time delay to diagnosis of ketosis was significantly higher for ketonuria than for plasma ketonemia (212 ± 67 vs. 140 ± 54 min, P = 0.0023), whereas no difference was noted between plasma and capillary ketonemia. CONCLUSIONS—The frequency of screening for ketosis and the efficiency of detection of ketosis definitely may be improved by the use of capillary blood ketone determination in clinical practice.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Bode, B. W., Steed, R. D., & Davidson, P. C. (1996) · Diabetes Care · doi:10.2337/diacare.19.4.324 · Cited by 257
Automated kinetic method for D-3-hydroxybutyrate in plasma or serum
McMurray, C. H., Blanchflower, W. J., & Rice, D. A. (1984) · Clinical Chemistry · doi:10.1093/clinchem/30.3.421 · Cited by 188
Umpierrez, G. E., Watts, N. B., & Phillips, L. S. (1995) · Diabetes Care · doi:10.2337/diacare.18.1.137 · Cited by 54
Causes, consequences, and recognition of false-positive reactions for ketones
Csako, G. (1990) · Clinical Chemistry · doi:10.1093/clinchem/36.7.1388a · Cited by 17
Cited by
Hyperglycemic crises in adult patients with diabetes
Kitabchi, A. E., Umpierrez, G. E., Miles, J. M., & Fisher, J. N. (2009) · Diabetes Care · doi:10.2337/dc09-9032 · Cited by 2,062
Hyperglycemic Crises in Adult Patients With Diabetes
Kitabchi, A. E., Umpierrez, G. E., Murphy, M. B., & Kreisberg, R. A. (2006) · Diabetes Care · doi:10.2337/dc06-9916 · Cited by 512
Dietary ketosis enhances memory in mild cognitive impairment
Krikorian, R., Shidler, M. D., Dangelo, K., Couch, S. C., Benoit, S. C., & Clegg, D. J. (2012) · Neurobiology of Aging · doi:10.1016/j.neurobiolaging.2010.10.006 · Cited by 361
Assessment and monitoring of glycemic control in children and adolescents with diabetes
Rewers, M., Pihoker, C., Donaghue, K., Hanas, R., Swift, P., & Klingensmith, G. J. (2009) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2009.00582.x · Cited by 342
Teymourian, H., Moonla, C., Tehrani, F., Vargas, E., Aghavali, R., Barfidokht, A., Tangkuaram, T., Mercier, P. P., Dassau, E., & Wang, J. (2020) · Analytical Chemistry · doi:10.1021/acs.analchem.9b05109 · Cited by 278
Metadata
- Authors
- Guerci, B., Benichou, M., Floriot, M., Bohme, P., Fougnot, S., Franck, P., & Drouin, P.
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2003-04-01
- Volume
- 26
- Issue
- 4
- Pages
- 1137-1141
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2100295319
- Cited by
- 130
- References
- 4
- FWCI
- 4.17
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Accuracy of an Electrochemical Sensor for Measuring Capillary Blood Ketones by Fingerstick Samples During Metabolic Deterioration After Continuous Subcutaneous Insulin Infusion Interruption in Type 1 Diabetic Patients. https://app.cochat.ai/papers/10.2337/diacare.26.4.1137
- The paper
- Guerci, B., Benichou, M., Floriot, M., Bohme, P., Fougnot, S., Franck, P., & Drouin, P. (2003). Accuracy of an Electrochemical Sensor for Measuring Capillary Blood Ketones by Fingerstick Samples During Metabolic Deterioration After Continuous Subcutaneous Insulin Infusion Interruption in Type 1 Diabetic Patients. Diabetes Care. https://doi.org/10.2337/diacare.26.4.1137
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.