Nasal Nitric Oxide: A Comparison of Measurement Techniques

Silkoff, P.E., Chatkin, J., Qian, W., Chakravorty, S., Gutierrez, C., Furlott, H., McClean, P., Rai, S., Zamel, N., Haight, J. · 1999 · American Journal of Rhinology
Cited by 62

The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.

Abstract

Nasal nitric oxide measurement may be a surrogate marker of upper airway inflammation. There is, however, no standardized measurement technique; and this led us to examine measurement techniques for acceptability and reproducibility. In five subjects we examined the flow dependence of nasal NO. In 13 healthy volunteers, nasal NO was measured on-line by five methods: 1) Tidal nasal and oral breathing: NO sampling during exclusive nasal followed by exclusive oral tidal breathing; 2) Fixed flow exhalation: NO sampling during exclusive nasal followed by exclusive oral exhalation at 100 mL/second from total lung capacity; 3) Nasal-oral aspiration: air aspirated from the mouth via both nares at 100 mL/second with glottis closure; 4) Aspiration from one nares: air aspirated from one nares at 3.3 mL/second using nitric oxide analyzer sample line with velum closure; 5) Nasal Insufflation: NO sampled at one nares as air insufflated into the other nares at a flow of 100 mL/second with velum closure. Acceptability of all methods was assessed by subjects and technicians. Nasal NO concentration showed a significant inverse correlation with transnasal flow rate. All methods showed excellent reproducibility as assessed by the intraclass correlation coefficient except tidal breathing, which showed highly variable breath-to-breath NO levels, although mean breath values were reproducible. Mean nasal NO concentrations with methods 1, 2, 3, 4, and 5 were 32.1, 50.2, 62.8, 1381, and 60.0 ppb, respectively. Velum closure was not always achieved in methods 4 and 5, whereas methods 1 and 2 required separate nasal and oral procedures. Method 5 had reduced acceptability. NO concentrations were similar with methods that used the same airflow (2, 3, and 5). Nasal NO can be sampled in different ways with excellent reproducibility. In view of the flow dependence of nasal NO, it is vital to use a constant flow rate, and lower airway NO contribution must be excluded or subtracted. The fixed flow exhalation appears to be the preferred method as it is highly reproducible and acceptable.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.2500/105065899781389803

Cites

3 of 3 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 62

Metadata

Authors
Silkoff, P. E., Chatkin, J., Qian, W., Chakravorty, S., Gutierrez, C., Furlott, H., McClean, P., Rai, S., Zamel, N., & Haight, J.
Venue
American Journal of Rhinology
Published
1999-05
Volume
13
Issue
3
Pages
169-178
ISSN
  • 1050-6586 (ISSN-L)
  • 1539-6290
OpenAlex
W2029864494
Cited by
62
References
3
FWCI
6.49
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.