Image-Quality Assessment for Several Positron Emitters Using the NEMA NU 4-2008 Standards in the Siemens Inveon Small-Animal PET Scanner
Open accessAbstract
UNLABELLED: The positron emitters (18)F, (68)Ga, (124)I, and (89)Zr are all relevant in small-animal PET. Each of these radionuclides has different positron energies and ranges and a different fraction of single photons emitted. Average positron ranges larger than the intrinsic spatial resolution of the scanner (for (124)I and (68)Ga) will deteriorate the effective spatial resolution and activity recovery coefficient (RC) for small lesions or phantom structures. The presence of single photons (for (124)I and (89)Zr) could increase image noise and spillover ratios (SORs). METHODS: Image noise, expressed as percentage SD in a uniform region (%SD), RC, and SOR (in air and water) were determined using the NEMA NU 4 small-animal image-quality phantom filled with 3.7 MBq of total activity of (18)F, (68)Ga, (124)I, or (89)Zr. Filtered backprojection (FBP), ordered-subset expectation maximization in 2 dimensions, and maximum a posteriori (MAP) reconstructions were compared. In addition to the NEMA NU 4 image-quality parameters, spatial resolutions were determined using small glass capillaries filled with these radionuclides in a water environment. RESULTS: The %SD for (18)F, (68)Ga, (124)I, and (89)Zr using FBP was 6.27, 6.40, 6.74, and 5.83, respectively. The respective RCs were 0.21, 0.11, 0.12, and 0.19 for the 1-mm-diameter rod and 0.97, 0.65, 0.64, and 0.88 for the 5-mm-diameter rod. SORs in air were 0.01, 0.03, 0.04, and 0.01, respectively, and in water 0.02, 0.10, 0.13, and 0.02. Other reconstruction algorithms gave similar differences between the radionuclides. MAP produced the highest RCs. For the glass capillaries using FBP, the full widths at half maximum for (18)F, (68)Ga, (124)I, and (89)Zr were 1.81, 2.46, 2.38, and 1.99 mm, respectively. The corresponding full widths at tenth maximum were 3.57, 6.52, 5.87, and 4.01 mm. CONCLUSION: With the intrinsic spatial resolution (approximately 1.5 mm) of this latest-generation small-animal PET scanner, the finite positron range has become the limiting factor for the overall spatial resolution and activity recovery in small structures imaged with (124)I and (68)Ga. The presence of single photons had only a limited effect on the image noise. MAP, as compared with the other reconstruction algorithms, increased RC and decreased %SD and SOR.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Physics of pure and non-pure positron emitters for PET: a review and a discussion
Conti, M., & Eriksson, L. (2016) · EJNMMI Physics · doi:10.1186/s40658-016-0144-5 · Cited by 393
Comparison of Gallium-68 and Fluorine-18 imaging characteristics in positron emission tomography
Sanchez-Crespo, A. (2013) · Applied Radiation and Isotopes · doi:10.1016/j.apradiso.2012.06.034 · Cited by 276
Positron emission tomography and functional characterization of a complete PBR/TSPO knockout
Banati, R. B., Middleton, R. J., Chan, R., Hatty, C. R., Wai-Ying Kam, W., Quin, C., Graeber, M. B., Parmar, A., Zahra, D., Callaghan, P., et al. (2014) · Nature Communications · doi:10.1038/ncomms6452 · Cited by 241
Metadata
- Authors
- Disselhorst, J. A., Brom, M., Laverman, P., Slump, C. H., Boerman, O. C., Oyen, W. J. G., Gotthardt, M., & Visser, E. P.
- Venue
- Journal of Nuclear Medicine
- Publisher
- Society of Nuclear Medicine
- Published
- 2010-03-17
- Volume
- 51
- Issue
- 4
- Pages
- 610-617
- ISSN
- 0161-5505 (ISSN-L)
- 2159-662X
- OpenAlex
- W2128957254
- Cited by
- 164
- References
- 0
- FWCI
- 11.33
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Image-Quality Assessment for Several Positron Emitters Using the NEMA NU 4-2008 Standards in the Siemens Inveon Small-Animal PET Scanner. https://app.cochat.ai/papers/10.2967/jnumed.109.068858
- The paper
- Disselhorst, J. A., Brom, M., Laverman, P., Slump, C. H., Boerman, O. C., Oyen, W. J. G., Gotthardt, M., & Visser, E. P. (2010). Image-Quality Assessment for Several Positron Emitters Using the NEMA NU 4-2008 Standards in the Siemens Inveon Small-Animal PET Scanner. Journal of Nuclear Medicine. https://doi.org/10.2967/jnumed.109.068858
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.