Accuracy of Dose Calibrators for 68Ga PET Imaging: Unexpected Findings in a Multicenter Clinical Pretrial Assessment
Open accessThe DOI matches its Crossref record.
Abstract
We report the discovery of a systematic miscalibration during the work-up process for site validation of a multicenter clinical PET imaging trial using 68Ga, which manifested as a consistent and reproducible underestimation in the quantitative accuracy (assessed by SUV) of a range of PET systems from different manufacturers at several different facilities around Australia. Methods: Sites were asked to follow a strict preparation protocol to create a radioactive phantom with 68Ga to be imaged using a standard clinical protocol before commencing imaging in the trial. All sites had routinely used 68Ga for clinical PET imaging for many years. The reconstructed image data were transferred to an imaging core laboratory for analysis, along with information about ancillary equipment such as the radionuclide dose calibrator. Fourteen PET systems were assessed from 10 nuclear medicine facilities in Australia, with the aim for each PET system being to produce images within 5% of the true SUV. Results: At initial testing, 10 of the 14 PET systems underestimated the SUV by 15% on average (range, 13%–23%). Multiple PET systems at one site, from two different manufacturers, were all similarly affected, suggesting a common cause. We eventually identified an incorrect factory-shipped dose calibrator setting from a single manufacturer as being the cause. The calibrator setting for 68Ga was subsequently adjusted by the users so that the reconstructed images produced accurate values. Conclusion: PET imaging involves a chain of measurements and calibrations to produce accurate quantitative performance. Testing of the entire chain is simple, however, and should form part of any quality assurance program or prequalifying site assessment before commencing a quantitative imaging trial or clinical imaging.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Hofman, M. S., Lawrentschuk, N., Francis, R. J., Tang, C., Vela, I., Thomas, P., Rutherford, N., Martin, J. M., Frydenberg, M., Shakher, R., et al. (2020) · The Lancet · doi:10.1016/s0140-6736(20)30314-7 · Cited by 2,040
Self-citationHofman, M. S., Emmett, L., Sandhu, S., Iravani, A., Joshua, A. M., Goh, J. C., Pattison, D. A., Tan, T. H., Kirkwood, I. D., Ng, S., et al. (2021) · The Lancet · doi:10.1016/s0140-6736(21)00237-3 · Cited by 1,170
Self-citationButeau, J. P., Martin, A. J., Emmett, L., Iravani, A., Sandhu, S., Joshua, A. M., Francis, R. J., Zhang, A. Y., Scott, A. M., Lee, S.-T., et al. (2022) · The Lancet Oncology · doi:10.1016/s1470-2045(22)00605-2 · Cited by 296
Self-citationHofman, M. S., Emmett, L., Sandhu, S., Iravani, A., Buteau, J. P., Joshua, A. M., Goh, J. C., Pattison, D. A., Tan, T. H., Kirkwood, I. D., et al. (2024) · The Lancet Oncology · doi:10.1016/s1470-2045(23)00529-6 · Cited by 230
Self-citationIravani, A., Violet, J., Azad, A., & Hofman, M. S. (2020) · Prostate Cancer and Prostatic Diseases · doi:10.1038/s41391-019-0174-x · Cited by 68
Self-citation
Metadata
- Authors
- Bailey, D. L., Hofman, M. S., Forwood, N. J., O’Keefe, G. J., Scott, A. M., van Wyngaardt, W. M., Howe, B., Kovacev, O., & Francis, R. J.
- Venue
- Journal of Nuclear Medicine
- Publisher
- Society of Nuclear Medicine
- Published
- 2018-01-11
- Volume
- 59
- Issue
- 4
- Pages
- 636-638
- ISSN
- 0161-5505 (ISSN-L)
- 2159-662X
- OpenAlex
- W2782820455
- Cited by
- 55
- References
- 0
- FWCI
- 3.78
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Accuracy of Dose Calibrators for 68Ga PET Imaging: Unexpected Findings in a Multicenter Clinical Pretrial Assessment. https://app.cochat.ai/papers/10.2967/jnumed.117.202861
- The paper
- Bailey, D. L., Hofman, M. S., Forwood, N. J., O’Keefe, G. J., Scott, A. M., van Wyngaardt, W. M., Howe, B., Kovacev, O., & Francis, R. J. (2018). Accuracy of Dose Calibrators for 68Ga PET Imaging: Unexpected Findings in a Multicenter Clinical Pretrial Assessment. Journal of Nuclear Medicine. https://doi.org/10.2967/jnumed.117.202861
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.