High-sensitivity troponin assays for the early rule-out or diagnosis of acute myocardial infarction in people with acute chest pain: a systematic review and cost-effectiveness analysis
Open accessAbstract
Background: Early diagnosis of acute myocardial infarction (AMI) can ensure quick and effective treatment but only 20% of adults with emergency admissions for chest pain have an AMI. High-sensitivity cardiac troponin (hs-cTn) assays may allow rapid rule-out of AMI and avoidance of unnecessary hospital admissions and anxiety. Objective: To assess the clinical effectiveness and cost-effectiveness of hs-cTn assays for the early (within 4 hours of presentation) rule-out of AMI in adults with acute chest pain. Methods: Sixteen databases, including MEDLINE and EMBASE, research registers and conference proceedings, were searched to October 2013. Study quality was assessed using QUADAS-2. The bivariate model was used to estimate summary sensitivity and specificity for meta-analyses involving four or more studies, otherwise random-effects logistic regression was used. The health-economic analysis considered the long-term costs and quality-adjusted life-years (QALYs) associated with different troponin (Tn) testing methods. The de novo model consisted of a decision tree and Markov model. A lifetime time horizon (60 years) was used. Results: Eighteen studies were included in the clinical effectiveness review. The optimum strategy, based on the Roche assay, used a limit of blank (LoB) threshold in a presentation sample to rule out AMI [negative likelihood ratio (LR–) 0.10, 95% confidence interval (CI) 0.05 to 0.18]. Patients testing positive could then have a further test at 2 hours; a result above the 99th centile on either sample and a delta (Δ) of ≥ 20% has some potential for ruling in an AMI [positive likelihood ratio (LR+) 8.42, 95% CI 6.11 to 11.60], whereas a result below the 99th centile on both samples and a Δ of < 20% can be used to rule out an AMI (LR– 0.04, 95% CI 0.02 to 0.10). The optimum strategy, based on the Abbott assay, used a limit of detection (LoD) threshold in a presentation sample to rule out AMI (LR– 0.01, 95% CI 0.00 to 0.08). Patients testing positive could then have a further test at 3 hours; a result above the 99th centile on this sample has some potential for ruling in an AMI (LR+ 10.16, 95% CI 8.38 to 12.31), whereas a result below the 99th centile can be used to rule out an AMI (LR– 0.02, 95% CI 0.01 to 0.05). In the base-case analysis, standard Tn testing was both most effective and most costly. Strategies considered cost-effective depending upon incremental cost-effectiveness ratio thresholds were Abbott 99th centile (thresholds of < £6597), Beckman 99th centile (thresholds between £6597 and £30,042), Abbott optimal strategy (LoD threshold at presentation, followed by 99th centile threshold at 3 hours) (thresholds between £30,042 and £103,194) and the standard Tn test (thresholds over £103,194). The Roche 99th centile and the Roche optimal strategy [LoB threshold at presentation followed by 99th centile threshold and/or Δ20% (compared with presentation test) at 1–3 hours] were extendedly dominated in this analysis. Conclusions: There is some evidence to suggest that hs-CTn testing may provide an effective and cost-effective approach to early rule-out of AMI. Further research is needed to clarify optimal diagnostic thresholds and testing strategies. Study registration: This study is registered as PROSPERO CRD42013005939. Funding: The National Institute for Health Research Health Technology Assessment programme.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies
Whiting, P. F., Rutjes, A. W. S., Westwood, M. E., Mallett, S., Deeks, J. J., Reitsma, J. B., Leeflang, M. M. G., Sterne, J. A. C., Bossuyt, P. M. M., & the QUADAS-2 Group* (2011) · Annals of Internal Medicine · doi:10.7326/0003-4819-155-8-201110180-00009 · Cited by 14,199
Self-citationMethods for the Economic Evaluation of Health Care Programmes
Drummond, M. E., Sculpher, M. J., Torrance, G. W., O’Brien, B. J., & Stoddart, G. L. (2005) · doi:10.1093/oso/9780198529446.001.0001 · Cited by 13,648
Reitsma, J. B., Glas, A. S., Rutjes, A. W. S., Scholten, R. J. P. M., Bossuyt, P. M., & Zwinderman, A. H. (2005) · Journal of Clinical Epidemiology · doi:10.1016/j.jclinepi.2005.02.022 · Cited by 3,369
Early versus delayed invasive intervention in acute coronary syndromes
Mehta, S. R., Granger, C. B., Boden, W. E., Steg, P. G., Bassand, J.-P., Faxon, D. P., Afzal, R., Chrolavicius, S., Jolly, S. S., Widimsky, P., et al. (2009) · New England Journal of Medicine · doi:10.1056/nejmoa0807986 · Cited by 925
Analytical Characteristics of High-Sensitivity Cardiac Troponin Assays
Apple, F. S., Collinson, P. O., & for the IFCC Task Force on Clinical Applications of Cardiac Biomarkers (2012) · Clinical Chemistry · doi:10.1373/clinchem.2011.165795 · Cited by 861
Goodacre, S., Bradburn, M., Fitzgerald, P., Cross, E., Collinson, P., Gray, A., & Hall, A. (2011) · Health Technology Assessment · doi:10.3310/hta15230 · Cited by 64
Cited by
Hotchkiss, R. S., Moldawer, L. L., Opal, S. M., Reinhart, K., Turnbull, I. R., & Vincent, J.-L. (2016) · Nature Reviews Disease Primers · doi:10.1038/nrdp.2016.45 · Cited by 1,620
Rossaint, R., Bouillon, B., Cerny, V., Coats, T. J., Duranteau, J., Fernández-Mondéjar, E., Filipescu, D., Hunt, B. J., Komadina, R., Nardi, G., et al. (2016) · Critical Care · doi:10.1186/s13054-016-1265-x · Cited by 1,484
Management of severe perioperative bleeding: Guidelines from the European Society of Anaesthesiology
Kozek-Langenecker, S. A., Afshari, A., Albaladejo, P., Santullano, C. A. A., De Robertis, E., Filipescu, D. C., Fries, D., Görlinger, K., Haas, T., Imberger, G., et al. (2013) · European Journal of Anaesthesiology · doi:10.1097/eja.0b013e32835f4d5b · Cited by 909
2021 AHA/ACC/ASE/CHEST/SAEM/SCCT/SCMR Guideline for the Evaluation and Diagnosis of Chest Pain
Gulati, M., Levy, P. D., Mukherjee, D., Amsterdam, E., Bhatt, D. L., Birtcher, K. K., Blankstein, R., Boyd, J., Bullock-Palmer, R. P., Conejo, T., et al. (2021) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2021.07.053 · Cited by 824
Kozek-Langenecker, S. A., Ahmed, A. B., Afshari, A., Albaladejo, P., Aldecoa, C., Barauskas, G., De Robertis, E., Faraoni, D., Filipescu, D. C., Fries, D., et al. (2017) · European Journal of Anaesthesiology · doi:10.1097/eja.0000000000000630 · Cited by 800
Metadata
- Authors
- Westwood, M., van Asselt, T., Ramaekers, B., Whiting, P., Thokala, P., Joore, M., Armstrong, N., Ross, J., Severens, J., & Kleijnen, J.
- Venue
- Health Technology Assessment
- Published
- 2015-06
- Volume
- 19
- Issue
- 44
- Pages
- 1-234
- ISSN
- 1366-5278 (ISSN-L)
- 2046-4924
- OpenAlex
- W1854312764
- Cited by
- 161
- References
- 6
- FWCI
- 5.95
- Type
- journal article
- Language
- en
- Open access
- diamond
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: High-sensitivity troponin assays for the early rule-out or diagnosis of acute myocardial infarction in people with acute chest pain: a systematic review and cost-effectiveness analysis. https://app.cochat.ai/papers/10.3310/hta19440
- The paper
- Westwood, M., van Asselt, T., Ramaekers, B., Whiting, P., Thokala, P., Joore, M., Armstrong, N., Ross, J., Severens, J., & Kleijnen, J. (2015). High-sensitivity troponin assays for the early rule-out or diagnosis of acute myocardial infarction in people with acute chest pain: a systematic review and cost-effectiveness analysis. Health Technology Assessment. https://doi.org/10.3310/hta19440
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.