Computer-aided detection of malignant lung nodules on chest radiographs: effect on observers’ performance
Open accessAbstract
OBJECTIVE: To evaluate the effect of computer-aided detection (CAD) system on observer performance in the detection of malignant lung nodules on chest radiograph. MATERIALS AND METHODS: Two hundred chest radiographs (100 normal and 100 abnormal with malignant solitary lung nodules) were evaluated. With CT and histological confirmation serving as a reference, the mean nodule size was 15.4 mm (range, 7-20 mm). Five chest radiologists and five radiology residents independently interpreted both the original radiographs and CAD output images using the sequential testing method. The performances of the observers for the detection of malignant nodules with and without CAD were compared using the jackknife free-response receiver operating characteristic analysis. RESULTS: Fifty-nine nodules were detected by the CAD system with a false positive rate of 1.9 nodules per case. The detection of malignant lung nodules significantly increased from 0.90 to 0.92 for a group of observers, excluding one first-year resident (p = 0.04). When lowering the confidence score was not allowed, the average figure of merit also increased from 0.90 to 0.91 (p = 0.04) for all observers after a CAD review. On average, the sensitivities with and without CAD were 87% and 84%, respectively; the false positive rates per case with and without CAD were 0.19 and 0.17, respectively. The number of additional malignancies detected following true positive CAD marks ranged from zero to seven for the various observers. CONCLUSION: The CAD system may help improve observer performance in detecting malignant lung nodules on chest radiographs and contribute to a decrease in missed lung cancer.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Overreliance may lead to blind trust (overtrust) or overutilization.
“whereas overreliance may lead to blind trust (overtrust) or overutilization. 11,12”
From Appropriate trust in artificial intelligence for the optical diagnosis of colorectal polyps: the role of human/artificial intelligence interaction (2024) · cites this paper as [12]
The Results section (page unavailable) reports that one resident rejected many true nodules CAD did not detect and that readers accepted false-positive CAD marks. These findings support a concern about trusting CAD too much, but the paper does not address overutilization as a consequence of overreliance. (LLM, conf=0.78)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
No references could be resolved yet.
Research works printed but not linked: 19
- [1]Muhm JR Miller WE Fontana RS Sanderson DR Uhlenhopp MA Lung cancer detected during a screening program using four-month chest radiographs Radiology 1983 148 609 615 6308709
- [2]Austin JH Romney BM Goldsmith LS Missed bronchogenic carcinoma: radiographic findings in 27 patients with a potentially resectable lesion evident in retrospect Radiology 1992 182 115 122 1727272
- [3]Monnier-Cholley L Arrivé L Porcel A Shehata K Dahan H Urban T Characteristics of missed lung cancer on chest radiographs: a French experience Eur Radiol 2001 11 597 605 11354754
- [4]Quekel LG Kessels AG Goei R van Engelshoven JM Miss rate of lung cancer on the chest radiograph in clinical practice Chest 1999 115 720 724 10084482
- [5]Choi EJ Jin GY Han YM Lee YS Kweon KS Solitary pulmonary nodule on helical dynamic CT scans: analysis of the enhancement patterns using a computer-aided diagnosis (CAD) system Korean J Radiol 2008 9 401 408 18838848 PMC2627214
- [6]Song KD Chung MJ Kim HC Jeong SY Lee KS Usefulness of the CAD system for detecting pulmonary nodule in real clinical practice Korean J Radiol 2011 12 163 168 21430932 PMC3052606
- [7]Goo JM A computer-aided diagnosis for evaluating lung nodules on chest CT: the current status and perspective Korean J Radiol 2011 12 145 155 21430930 PMC3052604
- [8]Kakeda S Moriya J Sato H Aoki T Watanabe H Nakata H Improved detection of lung nodules on chest radiographs using a commercial computer-aided diagnosis system AJR Am J Roentgenol 2004 182 505 510 14736690
- [9]Shiraishi J Li F Doi K Computer-aided diagnosis for improved detection of lung nodules by use of posterior-anterior and lateral chest radiographs Acad Radiol 2007 14 28 37 17178363 PMC1892186
- [10]Bley TA Baumann T Saueressig U Pache G Treier M Schaefer O Comparison of radiologist and CAD performance in the detection of CT-confirmed subtle pulmonary nodules on digital chest radiographs Invest Radiol 2008 43 343 348 18496038
- [11]He Q He W Wang K Ma D Effect of multiscale processing in digital chest radiography on automated detection of lung nodule with a computer assistance system J Digit Imaging 2008 21 Suppl 1 S164 S170 18239963 PMC3043876
- [12]Kasai S Li F Shiraishi J Doi K Usefulness of computer-aided diagnosis schemes for vertebral fractures and lung nodules on chest radiographs AJR Am J Roentgenol 2008 191 260 265 18562756
- [13]Li F Engelmann R Metz CE Doi K MacMahon H Lung cancers missed on chest radiographs: results obtained with a commercial computer-aided detection program Radiology 2008 246 273 280 18096539
- [14]van Beek EJ Mullan B Thompson B Evaluation of a real-time interactive pulmonary nodule analysis system on chest digital radiographic images: a prospective study Acad Radiol 2008 15 571 575 18423313
- [15]White CS Flukinger T Jeudy J Chen JJ Use of a computer-aided detection system to detect missed lung cancer at chest radiography Radiology 2009 252 273 281 19561261
- [16]de Hoop B De Boo DW Gietema HA van Hoorn F Mearadji B Schijf L Computer-aided detection of lung cancer on chest radiographs: effect on observer performance Radiology 2010 257 532 540 20807851
- [17]Chakraborty DP Analysis of location specific observer performance data: validated extensions of the jackknife free-response (JAFROC) method Acad Radiol 2006 13 1187 1193 16979067
- [18]Giger ML Chan HP Boone J Anniversary paper: History and status of CAD and quantitative image analysis: the role of Medical Physics and AAPM Med Phys 2008 35 5799 5820 19175137 PMC2673617
- [19]Li Q Li F Shiraishi J Katsuragawa S Sone S Doi K Investigation of new psychophysical measures for evaluation of similar images on thoracic computed tomography for distinction between benign and malignant nodules Med Phys 2003 30 2584 2593 14596294
Cited by
Sorted by citations
Hwang, E. J., Park, S., Jin, K.-N., Kim, J. I., Choi, S. Y., Lee, J. H., Goo, J. M., Aum, J., Yim, J.-J., Cohen, J. G., et al. (2019) · JAMA Network Open · doi:10.1001/jamanetworkopen.2019.1095 · Cited by 447 · see citing works
Self-citationSim, Y., Chung, M. J., Kotter, E., Yune, S., Kim, M., Do, S., Han, K., Kim, H., Yang, S., Lee, D.-J., et al. (2020) · Radiology · doi:10.1148/radiol.2019182465 · Cited by 248 · see citing works
Li, D., Pehrson, L. M., Lauridsen, C. A., Tøttrup, L., Fraccaro, M., Elliott, D., Zając, H. D., Darkner, S., Carlsen, J. F., & Nielsen, M. B. (2021) · Diagnostics · doi:10.3390/diagnostics11122206 · Cited by 35 · see citing works
van der Zander, Q. E. W., Roumans, R., Kusters, C. H. J., Dehghani, N., Masclee, A. A. M., de With, P. H. N., van der Sommen, F., Snijders, C. C. P., & Schoon, E. J. (2024) · Gastrointestinal Endoscopy · doi:10.1016/j.gie.2024.06.029 · Cited by 27 · see citing works
XAI-based Clinical Decision Support System: A Systematic Review
Kim, S., Kim, D., Kim, M., Ko, H., & Jeong, O. (2024) · doi:10.20944/preprints202406.0721.v1 · Cited by 72 · see citing works
- Authors
- Lee, K. H., Goo, J. M., Park, C. M., Lee, H. J., & Jin, K. N.
- Affiliations
- Seoul National University (KR)
- Corresponding author
- Goo, Jin Mo
- Venue
- Korean Journal of Radiology
- Publisher
- The Korean Society of Radiology
- Published
- 2012
- Volume
- 13
- Issue
- 5
- Pages
- 564
- ISSN
- 1229-6929 (ISSN-L)
- 2005-8330
- PMC
- PMC3435853
- OpenAlex
- W2139150365
- Cited by
- 37
- References
- 0
- Citations by year
- 2024: 3 · 2023: 4 · 2022: 2 · 2021: 10 · 2020: 4 · 2019: 3 · 2018: 1 · 2017: 2 · 2016: 2 · 2015: 2 · 2014: 3 · 2013: 1
- FWCI
- 1.50
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Korean Journal of Radiology (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- Seoul National University Open Repository (Seoul National University) (submittedVersion)
- Funders
- Ministry of Knowledge Economy
- Topics
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Lung Cancer Diagnosis and Treatment
- Physical Sciences › Computer Science › Artificial Intelligence › AI in cancer detection
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiomics and Machine Learning in Medical Imaging
- Keywords
- computer-aided detection, chest radiograph, observer performance, lung cancer detection, false positive rate, sensitivity, radiologist interpretation, sequential testing
- MeSH terms
- Aged, Algorithms, Diagnosis, Computer-Assisted, Diagnosis, Differential, Female, Humans, Image Interpretation, Computer-Assisted, Lung Neoplasms, Male, Middle Aged, ROC Curve, Radiography, Thoracic, Tomography, X-Ray Computed, Reproducibility of Results, Observer Variation
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review