Machine learning approach for ambient-light-corrected parameters and the Pupil Reactivity (PuRe) score in smartphone-based pupillometry

Bogucki, Aleksander, John, Ivo, Zinkiewicz, Łukasz, Jachura, Michał, Jaworski, Damian, Suwała, Karolina, Chrost, Hugo, Wlodarski, Michal, Kałużny, Jakub, Campbell, Doug, Bakken, Paul, Pandya, Shawna, Chrapkiewicz, Radosław, Manohar, Sanjay G. · 2024 · Frontiers in Neurology

Published version of a preprint

Cited by 16

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

Introduction: The pupillary light reflex (PLR) is the constriction of the pupil in response to light. The PLR in response to a pulse of light follows a complex waveform that can be characterized by several parameters. It is a sensitive marker of acute neurological deterioration, but is also sensitive to the background illumination in the environment in which it is measured. To detect a pathological change in the PLR, it is therefore necessary to separate the contributions of neuro-ophthalmic factors from ambient illumination. Illumination varies over several orders of magnitude and is difficult to control due to diurnal, seasonal, and location variations. Methods and results: We assessed the sensitivity of seven PLR parameters to differences in ambient light, using a smartphone-based pupillometer (AI Pupillometer, Solvemed Inc.). Nine subjects underwent 345 measurements in ambient conditions ranging from complete darkness (<5 lx) to bright lighting (≲10,000 lx). Lighting most strongly affected the initial pupil size, constriction amplitude, and velocity. Nonlinear models were fitted to find the correction function that maximally stabilized PLR parameters across different ambient light levels. Next, we demonstrated that the lighting-corrected parameters still discriminated reactive from unreactive pupils. Ten patients underwent PLR testing in an ophthalmology outpatient clinic setting following the administration of tropicamide eye drops, which rendered the pupils unreactive. The parameters corrected for lighting were combined as predictors in a machine learning model to produce a scalar value, the Pupil Reactivity (PuRe) score, which quantifies Pupil Reactivity on a scale 0–5 (0, non-reactive pupil; 0–3, abnormal/“sluggish” response; 3–5, normal/brisk response). The score discriminated unreactive pupils with 100% accuracy and was stable under changes in ambient illumination across four orders of magnitude. Discussion: This is the first time that a correction method has been proposed to effectively mitigate the confounding influence of ambient light on PLR measurements, which could improve the reliability of pupillometric parameters both in pre-hospital and inpatient care settings. In particular, the PuRe score offers a robust measure of Pupil Reactivity directly applicable to clinical practice. Importantly, the formulae behind the score are openly available for the benefit of the clinical research community.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.3389/fneur.2024.1363190

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Research works printed but not linked: 31

  1. Kardon, 1995, Pupillary light reflex, Curr Opin Ophthalmol, 6, 20, 10.1097/00055735-199512000-00004 · DOI
  2. Ellis, 1981, The pupillary light reflex in normal subjects, Br J Ophthalmol, 65, 754, 10.1136/bjo.65.11.754 · DOI · Paper page
  3. Meeker, 2005, Pupil examination: validity and clinical utility of an automated pupillometer, J Neurosci Nurs, 37, 34, 10.1097/01376517-200502000-00006 · doi:10.1097/01376517-200502000-00006
  4. Sharma, 2016, Factors influencing the pupillary light reflex in healthy individuals, Graefe's Archiv Clin Exp Ophthalmol, 254, 1353, 10.1007/s00417-016-3311-4 · DOI · Paper page
  5. Ong, 2019, The effect of ambient light conditions on quantitative pupillometry, Neurocrit Care, 30, 316, 10.1007/s12028-018-0607-8 · DOI · Paper page
  6. Couret, 2019, The effect of ambient-light conditions on quantitative pupillometry: a history of rubber cup, Neurocrit Care, 30, 492, 10.1007/s12028-018-0664-z · doi:10.1007/s12028-018-0664-z
  7. Greer, 2020, Determination of brain death/death by neurologic criteria: the world brain death project, J Am Med Assoc, 324, 1078, 10.1001/jama.2020.11586 · DOI · Paper page
  8. Couret, 2016, Reliability of standard pupillometry practice in neurocritical care: an observational, double-blinded study, Crit Care, 20, 99, 10.1186/s13054-016-1239-z · DOI · Paper page
  9. Aderman, 2023, Normative values for pupillary light reflex metrics among healthy service academy cadets, Milit Med, 2023, usad271, 10.1093/milmed/usad271 · doi:10.1093/milmed/usad271
  10. Shah, 2020, Establishing a normative database for quantitative pupillometry in the pediatric population, BMC Ophthalmol, 20, 121, 10.1186/s12886-020-01389-x · doi:10.1186/s12886-020-01389-x
  11. Lussier, 2019, Distributions and reference ranges for automated pupillometer values in neurocritical care patients, J Neurosci Nurs, 51, 335, 10.1097/JNN.0000000000000478 · doi:10.1097/jnn.0000000000000478
  12. Phillips, 2019, A systematic review assessing the current state of automated pupillometry in the NeuroICU, Neurocrit Care, 31, 142, 10.1007/s12028-018-0645-2 · doi:10.1007/s12028-018-0645-2
  13. Kim, 2022, Quantitative assessments of pupillary light reflexes in neurocritically ill patients, J Neurocrit Care, 15, 79, 10.18700/jnc.220085 · doi:10.18700/jnc.220085
  14. Boulter, 2021, The expanding role of quantitative pupillometry in the evaluation and management of traumatic brain injury, Front Neurol, 12, 685313, 10.3389/fneur.2021.685313 · doi:10.3389/fneur.2021.685313
  15. Bower, 2021, Quantitative pupillometry in the intensive care unit, J Intens Care Med, 36, 383, 10.1177/0885066619881124 · doi:10.1177/0885066619881124
  16. Chen, 2011, Pupillary reactivity as an early indicator of increased intracranial pressure: the introduction of the neurological pupil index, Surg Neurol Int, 2, 82, 10.4103/2152-7806.82248 · DOI · Paper page
  17. Ong, 2023, Quantitative pupillometry: not a clear predictor of delayed cerebral ischemia after subarachnoid hemorrhage, Neurocritical Care, 39, 34, 10.1007/s12028-023-01709-1 · doi:10.1007/s12028-023-01709-1
  18. Prescott, 2022, Anisocoria and poor pupil reactivity by quantitative pupillometry in patients with intracranial pathology, Crit Care Med, 50, e143, 10.1097/CCM.0000000000005272 · DOI
  19. Olson, 2016, The use of automated pupillometry in critical care, Crit Care Nurs Clin N Am, 28, 101, 10.1016/j.cnc.2015.09.003 · DOI
  20. Ong, 2023, The prognostic potential of pupillometry in patients with acute brain injury, Lancet Neurol, 22, 876, 10.1016/S1474-4422(23)00314-9 · doi:10.1016/s1474-4422(23)00314-9
  21. Zafar, 2014, Automated pupillometer for monitoring the critically ill patient: a critical appraisal, J Crit Care, 29, 599, 10.1016/j.jcrc.2014.01.012 · DOI
  22. Leinonen, 2018, Raised intracranial pressure and brain edema, Handb Clin Neurol, 145, 25, 10.1016/B978-0-12-802395-2.00004-3 · DOI
  23. Changa, 2019, Management of elevated intracranial pressure: a review, Current Neurol Neurosci Rep, 19, 99, 10.1007/s11910-019-1010-3 · DOI
  24. Olson, 2016, Interrater reliability of pupillary assessments, Neurocrit Care, 24, 251, 10.1007/s12028-015-0182-1 · DOI
  25. Omburo, 2017, High variance in pupillary examination findings among postanesthesia care unit nurses, J PeriAnesth Nurs, 32, 219, 10.1016/j.jopan.2015.05.117 · DOI · Paper page
  26. Chougule, 2019, Light-induced pupillary responses in Alzheimer's disease, Front Neurol, 10, 360, 10.3389/fneur.2019.00360 · DOI · Paper page
  27. Packiasabapathy, 2021, Pupillometry in perioperative medicine: a narrative review, Can J Anesth, 68, 566, 10.1007/s12630-020-01905-z · doi:10.1007/s12630-020-01905-z
  28. Alexis, 2023, Assessment of altered mental status in the emergency department using quantitative pupillometry: a prospective cohort study (P7-4, Neurology, 10.1212/WNL.0000000000204165 · doi:10.1212/wnl.0000000000204165
  29. Fink, 2023, From pre-processing to advanced dynamic modeling of pupil data, Behav Res Methods, 23, 2098, 10.3758/s13428-023-02098-1 · DOI
  30. Reilly, 2023, The human task-evoked pupillary response function is linear: Implications for baseline response scaling in pupillometry, Behav Res Methods, 59, 865, 10.3758/s13428-018-1134-4 · DOI
  31. Sherman, 2018, Acute and chronic effects of hypobaric exposure upon the brain, 10.5772/intechopen.74231 · DOI

Cited by

See all 16

Metadata

Authors
Bogucki, A., John, I., Zinkiewicz, Ł., Jachura, M., Jaworski, D., Suwała, K., Chrost, H., Wlodarski, M., Kałużny, J., Campbell, D., Bakken, P., Pandya, S., Chrapkiewicz, R., & Manohar, S. G.
Venue
Frontiers in Neurology
Published
2024-04-09
Volume
15
ISSN
1664-2295
Cited by
16
References
2

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Machine learning approach for ambient-light-corrected parameters and the Pupil Reactivity (PuRe) score in smartphone-based pupillometry. https://app.cochat.ai/papers/10.3389/fneur.2024.1363190
The paper
Bogucki, A., John, I., Zinkiewicz, Ł., Jachura, M., Jaworski, D., Suwała, K., Chrost, H., Wlodarski, M., Kałużny, J., Campbell, D., Bakken, P., Pandya, S., Chrapkiewicz, R., & Manohar, S. G. (2024). Machine learning approach for ambient-light-corrected parameters and the Pupil Reactivity (PuRe) score in smartphone-based pupillometry. Frontiers in Neurology. https://doi.org/10.3389/fneur.2024.1363190

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.