Cannulation strategies in adult veno-arterial and veno-venous extracorporeal membrane oxygenation: Techniques, limitations, and special considerations
Open accessThe DOI matches its Crossref record; 28 of the 28 references it cites were found in a registry or an index.
Abstract
Extracorporeal membrane oxygenation (ECMO) refers to specific mechanical devices used to temporarily support the failing heart and/or lung. Technological advances as well as growing collective knowledge and experience have resulted in increased ECMO use and improved outcomes. Veno-arterial (VA) ECMO is used in selected patients with various etiologies of cardiogenic shock and entails either central or peripheral cannulation. Central cannulation is frequently used in postcardiotomy cardiogenic shock and is associated with improved venous drainage and reduced concern for upper body hypoxemia as compared to peripheral cannulation. These concerns inherent to peripheral VA ECMO may be addressed through so-called triple cannulation approaches. Veno-venous (VV) ECMO is increasingly employed in selected patients with respiratory failure refractory to more conventional measures. Newer dual lumen VV ECMO cannulas may facilitate extubation and mobilization. In summary, the pathology being addressed impacts the ECMO approach that is deployed, and each ECMO implementation has distinct virtues and drawbacks. Understanding these considerations is crucial to safe and effective ECMO use.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Zapol, W. M. (1979) · JAMA · doi:10.1001/jama.1979.03300200023016 · Cited by 1,184 · cited 1×
“[1] Subsequently, interest in using ECMO in adults waned following a multicenter NIH‐sponsored trial showing no mortality benefit in adults with respiratory failure, and it was primarily used to manage respiratory failure in the neonatal and pediatric populations.[2,3] Since then, technical and other advancements have…”
BackgroundInfluentialHill, J. D., O'Brien, T. G., Murray, J. J., Dontigny, L., Bramson, M. L., Osborn, J. J., & Gerbode, F. (1972) · New England Journal of Medicine · doi:10.1056/nejm197203232861204 · Cited by 1,096 · cited 1×
“The first successful application of ECMO in an adult patient was described in 1972.[1] Subsequently, interest in using ECMO in adults waned following a multicenter NIH‐sponsored trial showing no mortality benefit in adults with respiratory failure, and it was primarily used to manage respiratory failure in the neonata…”
MethodCheng, R., Hachamovitch, R., Kittleson, M., Patel, J., Arabia, F., Moriguchi, J., Esmailian, F., & Azarbal, B. (2014) · The Annals of Thoracic Surgery · doi:10.1016/j.athoracsur.2013.09.008 · Cited by 867 · cited 1×
“7%.[18] Some centers routinely place a distally directed 6–10 Fr catheter in the cannulated limb [Figure 2d].”
BackgroundExtracorporeal Membrane Oxygenation for ARDS in Adults
Brodie, D., & Bacchetta, M. (2011) · New England Journal of Medicine · doi:10.1056/nejmct1103720 · Cited by 836
Extracorporeal Membrane Oxygenation (ECMO) in Neonatal Respiratory Failure
BARTLETT, R. H., TOOMASIAN, J., ROLOFF, D., GAZZANIGA, A. B., CORWIN, A. G., & RUCKER, R. (1986) · Annals of Surgery · doi:10.1097/00000658-198609000-00003 · Cited by 392 · cited 1×
“[1] Subsequently, interest in using ECMO in adults waned following a multicenter NIH‐sponsored trial showing no mortality benefit in adults with respiratory failure, and it was primarily used to manage respiratory failure in the neonatal and pediatric populations.[2,3] Since then, technical and other advancements have…”
BackgroundInfluentialBisdas, T., Beutel, G., Warnecke, G., Hoeper, M. M., Kuehn, C., Haverich, A., & Teebken, O. E. (2011) · The Annals of Thoracic Surgery · doi:10.1016/j.athoracsur.2011.02.018 · Cited by 336 · cited 1×
“[19] Risk factors for distal ischemia include the use of larger bore cannulas as well as peripheral vascular disease.[20] Of note, weaning VA ECMO flow will also decrease flow through the distal perfusion cannula such that ischemia may result from a prolonged duration of low ECMO flow despite the presence of a distal…”
BackgroundAbrams, D., Javidfar, J., Farrand, E., Mongero, L. B., Agerstrand, C. L., Ryan, P., Zemmel, D., Galuskin, K., Morrone, T. M., Boerem, P., et al. (2014) · Critical Care · doi:10.1186/cc13746 · Cited by 322 · cited 2×
“A further advantage is that upper body cannulation, achieved using the right IJ vein and the right axillary or subclavian artery [Figure 2c], may permit patient mobilization.[11] Potential complications of this approach are highlighted in a cases series from Columbia University Medical Center (New York, USA), in which…”
BackgroundMethodCannulation strategies for percutaneous extracorporeal membrane oxygenation in adults
Napp, L. C., Kühn, C., Hoeper, M. M., Vogel-Claussen, J., Haverich, A., Schäfer, A., & Bauersachs, J. (2016) · Clinical Research in Cardiology · doi:10.1007/s00392-015-0941-1 · Cited by 251 · cited 7×
“preload, afterload, oxygenation, and location of the watershed.[15]”
BackgroundInfluentialInitial Experience With Single Cannulation for Venovenous Extracorporeal Oxygenation in Adults
Bermudez, C. A., Rocha, R. V., Sappington, P. L., Toyoda, Y., Murray, H. N., & Boujoukos, A. J. (2010) · The Annals of Thoracic Surgery · doi:10.1016/j.athoracsur.2010.06.017 · Cited by 187 · cited 1×
“Obviously, mobilization must be done with great care as reported complications include caudad or cephalad cannula migration as well as rotation of the return port away from the tricuspid valve.[35,38] From a mechanical standpoint, the design of the Avalon Elite® allows for bicaval drainage of deoxygenated blood at a c…”
BackgroundSquiers, J. J., Lima, B., & DiMaio, J. M. (2016) · Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/j.jtcvs.2016.02.067 · Cited by 171 · cited 1×
“Alternatively, arterial blood can also be returned using an end‐to‐side graft to the right subclavian or axillary artery.[6,7]”
BackgroundShekar, K., Mullany, D. V., Thomson, B., Ziegenfuss, M., Platts, D. G., & Fraser, J. F. (2014) · Critical Care · doi:10.1186/cc13865 · Cited by 170 · cited 1×
“Alternatively, arterial blood can also be returned using an end‐to‐side graft to the right subclavian or axillary artery.[6,7]”
BackgroundEchocardiography for adult patients supported with extracorporeal membrane oxygenation
Douflé, G., Roscoe, A., Billia, F., & Fan, E. (2015) · Critical Care · doi:10.1186/s13054-015-1042-2 · Cited by 168 · cited 4×
“S15 Annals of Cardiac Anaesthesia | Vol 20 | Special Issue 1 | January 2017 junction or within the RA.[15,32] This allows deoxygenated blood to be removed from the IVC and oxygenated blood to be returned to the SVC or RA.”
BackgroundInfluentialUse of Bicaval Dual-Lumen Catheter for Adult Venovenous Extracorporeal Membrane Oxygenation
Javidfar, J., Brodie, D., Wang, D., Ibrahimiye, A. N., Yang, J., Zwischenberger, J. B., Sonett, J., & Bacchetta, M. (2011) · The Annals of Thoracic Surgery · doi:10.1016/j.athoracsur.2011.03.002 · Cited by 160 · cited 1×
“In the optimally positioned Avalon Elite® cannula, this allows for very little recirculation of oxygenated blood into the ECMO circuit when compared to two cannulas (femoral‐jugular or femoral‐femoral) VV ECMO strategies.[39]”
MethodCardiac Decompression on Extracorporeal Life Support: A Review and Discussion of the Literature
Rupprecht, L., Flörchinger, B., Schopka, S., Schmid, Ç., Philipp, A., Lunz, D., Müller, T. M., & Camboni, D. (2013) · ASAIO Journal · doi:10.1097/mat.0b013e3182a4b2f6 · Cited by 155 · cited 1×
“Central cannulation is frequently used in postcardiotomy cardiogenic shock.[8] It employs cannulas already in place for typical cardiopulmonary bypass (CPB).”
BackgroundSubclavian Artery Cannulation for Venoarterial Extracorporeal Membrane Oxygenation
Javidfar, J., Brodie, D., Costa, J. A., Miller, J. C., Jurrado, J., LaVelle, M., Newmark, A., Takayama, H., Sonett, J. R., & Bacchetta, M. D. (2012) · ASAIO Journal · doi:10.1097/mat.0b013e318268ea15 · Cited by 116 · cited 1×
“[11] Potential complications of this approach are highlighted in a cases series from Columbia University Medical Center (New York, USA), in which three out of twenty VA ECMO patients required conversion from subclavian to femoral artery outflow due to one case each of hematoma, infection, and arm swelling.[17]”
MethodSidebotham, D. A., Allen, S. J., McGeorge, A., Ibbott, N., & Willcox, T. W. (2012) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2012.02.001 · Cited by 110 · cited 1×
“This arrangement is reportedly associated with a greater incidence of recirculation than femoral‐jugular cannulation.[34]”
BackgroundExtracorporeal Membrane Oxygenation Watershed
Hoeper, M. M., Tudorache, I., Kühn, C., Marsch, G., Hartung, D., Wiesner, O., Boenisch, O., Haverich, A., & Hinrichs, J. B. (2014) · Circulation · doi:10.1161/circulationaha.114.011677 · Cited by 108 · cited 1×
“Depending on native lung function, volume status, and cardiac function, blood ejected from the LV may be poorly oxygenated, and this blood may mal‐perfuse the coronary arteries and cerebral circulation.[12]”
BackgroundFeasibility of smaller arterial cannulas in venoarterial extracorporeal membrane oxygenation
Takayama, H., Landes, E., Truby, L., Fujita, K., Kirtane, A. J., Mongero, L. B., Yuzefpolskaya, M., Colombo, P. C., Jorde, U. P., Kurlansky, P. A., et al. (2015) · Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/j.jtcvs.2015.01.042 · Cited by 107
Kohler, K., Valchanov, K. P., Nias, G., & Vuylsteke, A. (2012) · Perfusion · doi:10.1177/0267659112468014 · Cited by 92 · cited 1×
“provide a thorough review of the physics underlying cannula selection.[24] As noted in their review, the intent of appropriate cannula choice is to provide full flow without damaging cells, particularly erythrocytes, from excessive shear stress and turbulence.”
BackgroundTruby, L., Mundy, L. R., Kalesan, B., Kirtane, A. J., Colombo, P. C., Takeda, K., Fukuhara, S., Naka, Y., & Takayama, H. (2015) · ASAIO Journal · doi:10.1097/mat.0000000000000225 · Cited by 85 · cited 1×
“Less commonly, the right internal jugular (IJ) vein may also be used, with either a single lumen or a two‐stage bicaval cannula.[4,5] Typically, the arterial cannula is a short 17–21 Fr catheter placed in the femoral artery with the tip in the common iliac artery.”
MethodHou, X., Yang, X., Du, Z., Xing, J., Li, H., Jiang, C., Wang, J., Xing, Z., Li, S., Li, X., et al. (2015) · Critical Care · doi:10.1186/s13054-015-0791-2 · Cited by 81 · cited 5×
“An animal model of dual circuit circulation, in which femoral VA ECMO was applied to sheep in respiratory failure, showed that oxygen saturation in the SVC, pulmonary artery, and ascending aorta ranged from 33% to 40% while it was 84% in the IVC.[13,14]”
BackgroundInfluentialApplication of veno-arterial-venous extracorporeal membrane oxygenation in differential hypoxia
Choi, J. Y., Kim, S., Kim, S., Kim, K.-S., Kim, K.-S., Joo, S., & Lee, J. I. (2014) · Multidisciplinary Respiratory Medicine · doi:10.1186/2049-6958-9-55 · Cited by 79 · cited 1×
“has been used in peripherally cannulated patients with significant heart and lung dysfunction.[16] Here, part of”
MethodRamakrishna, H., Sen, A., Callisen, H. E., Alwardt, C. M., Larson, J. S., Lowell, A. A., Libricz, S. L., Tarwade, P., & Patel, B. (2016) · Annals of Cardiac Anaesthesia · doi:10.4103/0971-9784.173027 · Cited by 60 · cited 2×
“As such, VV ECMO provides no cardiac support, and all organs are exposed to approximately the same oxygen content, which is determined by the relative contributions of the VV ECMO device, native lung function and oxygen extraction, and blood returning from the coronary sinus.[31]”
BackgroundMethodSelf-citationRubino, A., Vuylsteke, A., Jenkins, D. P., Fowles, J.-A., Hockings, L. E., & Valchanov, K. P. (2014) · The International Journal of Artificial Organs · doi:10.5301/ijao.5000357 · Cited by 49 · cited 3×
“Proper function of this cannula requires that the return lumen orifice is situated superior (cranial) to the tricuspid valve, with the orifice oriented toward the tricuspid valve.[35] Positioning the cannula in this manner allows drainage of mainly deoxygenated blood from the SVC and IVC as well as return of mainly ox…”
BackgroundNoly, P., Kirsch, M. E. W., Quessard, A., Léger, P., Pavie, A., Amour, J., & Leprince, P. N. (2014) · Interactive Cardiovascular and Thoracic Surgery · doi:10.1093/icvts/ivu072 · Cited by 48
Shafii, A. E., & McCurry, K. R. (2012) · The Annals of Thoracic Surgery · doi:10.1016/j.athoracsur.2012.02.024 · Cited by 41 · cited 1×
“The cannula is designed for right IJ vein insertion and so may not be appropriate for patients in whom this vessel cannot accommodate a large bore catheter although there are reports of using the left IJ and subclavian veins.[40,41] Unsurprisingly, for a dual lumen catheter, cannula size limitations may limit ECMO flo…”
BackgroundAbrams, D. C., Brodie, D., Javidfar, J., BRENNER, K. R., Wang, D., Zwischenberger, J. B., Sonett, J. R., & Bacchetta, M. D. (2012) · ASAIO Journal · doi:10.1097/mat.0b013e31826feda5 · Cited by 27 · cited 1×
“The cannula is designed for right IJ vein insertion and so may not be appropriate for patients in whom this vessel cannot accommodate a large bore catheter although there are reports of using the left IJ and subclavian veins.[40,41] Unsurprisingly, for a dual lumen catheter, cannula size limitations may limit ECMO flo…”
BackgroundExtracorporeal membrane oxygenation (ECMO) in neonatal respiratory failure: 100 cases
Ford, E. G. (1987) · Journal of Pediatric Surgery · doi:10.1016/s0022-3468(87)80135-5 · Cited by 20
Cited by
Bonicolini, E., Martucci, G., Simons, J., Raffa, G. M., Spina, C., Lo Coco, V., Arcadipane, A., Pilato, M., & Lorusso, R. (2019) · Critical Care · doi:10.1186/s13054-019-2541-3 · Cited by 239
Metadata
- Authors
- Jayaraman, A., Cormican, D., Shah, P., & Ramakrishna, H.
- Venue
- Annals of Cardiac Anaesthesia
- Publisher
- Medknow
- Published
- 2017
- Volume
- 20
- Issue
- 5
- Pages
- 11
- ISSN
- 0971-9784 (ISSN-L)
- OpenAlex
- W2571377300
- Cited by
- 199
- References
- 28
- FWCI
- 8.26
- Type
- journal article
- Language
- en
- Open access
- diamond
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›