False Positive Rate of Carbon Monoxide Saturation by Pulse Oximetry of Emergency Department Patients
The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Symptoms of carbon monoxide (CO) poisoning are non-specific. Diagnosis requires suspicion of exposure, confirmed by measuring ambient CO levels or carboxyhemoglobin (COHb). An FDA-approved pulse oximeter (Rad-57) can measure CO saturation (S pCO ). The device accuracy has implications for clinical decision-making. METHODS: From April 1 to August 15, 2008, study personnel measured S pCO and documented demographic factors at time of clinical blood draw, in a convenience sample of 1,363 subjects presenting to the emergency department at Intermountain Medical Center, Murray, Utah. The technician then assayed COHb. COHb and S pCO values were compared by subject; false positive or negative values were defined as S pCO at least 3 percentage points greater or less than COHb level, reported by the manufacturer to be ± 1 SD in performance. RESULTS: In 1,363 subjects, 613 (45%) were male, 1,141 (84%) were light-skinned, 14 in shock, 4 with CO poisoning, and 122 (9%) met the criteria for a false positive value (range 3–19 percentage points), while 247 (18%) met the criteria for a false negative value (−13 to −3 percentage points). Risks for a false positive S pCO reading included being female and having a lower perfusion index. Methemoglobin, body temperature, and blood pressure also appear to influence the S pCO accuracy. There was variability among monitors, possibly related to technician technique, as rotation of monitors among technicians was not enforced. CONCLUSIONS: While the Rad-57 pulse oximeter functioned within the manufacturer's specifications, clinicians using the Rad-57 should expect some S pCO readings to be significantly higher or lower than COHb measurements, and should not use S pCO to direct triage or patient management. An elevated S pCO could broaden the diagnosis of CO poisoning in patients with non-specific symptoms. However, a negative S pCO level in patients suspected of having CO poisoning should never rule out CO poisoning, and should always be confirmed by COHb.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Unfortunately, it is still unclear if the accuracy of pulse CO oximetry alone compared with COHb measured by spectrophotometry from laboratory CO oximeter is adequate, and thus, pulse CO oximetry levels should be con fi rmed with laboratory measurements (12, 46) (47).”
From Carbon monoxide poisoning: pathogenesis, management, and future directions of therapy (2017) · cites this paper as [47]
In the citing paper
“A prospective study on the performance of pulse CO oximetry showed that normal COHb pulse CO oximetry values cannot rule out CO poisoning (47).”
From Carbon monoxide poisoning: pathogenesis, management, and future directions of therapy (2017) · cites this paper as [47]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Feiner, J. R., Severinghaus, J. W., & Bickler, P. E. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000285988.35174.d9 · Cited by 400
Measurement of Carboxyhemoglobin and Methemoglobin by Pulse Oximetry: A Human Volunteer Study
Barker, S. J., Curry, J., Redford, D., & Morgan, S. (2006) · Anesthesiology · doi:10.1097/00000542-200611000-00008 · Cited by 238
Research works printed but not linked: 17
- Hampson NB, , Weaver LK. Carbon monoxide poisoning: a new incidence for an old disease. Undersea Hyperb Med, 2007; 34(3):163-168.
- 10.1056/NEJMcp0808891 · DOI · Paper page
- 10.1016/S0196-0644(95)70262-8 · DOI
- 10.1016/S0379-0738(03)00011-2 · doi:10.1016/s0379-0738(03)00011-2
- 10.1016/S0196-0644(94)70137-7 · DOI
- 10.1016/j.jemermed.2007.12.004 · DOI
- 10.1080/15563650701725110 · DOI
- 10.1017/S1049023X00008128 · DOI
- Hampson NB, , Weaver LK. Noninvasive CO measurement by first responders. JEMS, 2006; 31(5):S10-S12.
- 10.1016/j.annemergmed.2010.08.016 · DOI
- 10.1016/j.annemergmed.2010.12.024 · DOI
- 10.3109/08958370902839754 · DOI
- Coulange M, , Barthelemy A, , Hug F, , Thierry AL, , De Haro L. Reliability of new pulse CO-oximeter in victims of carbon monoxide poisoning. Undersea Hyperb Med, 2008; 35(2):107-111.
- 10.1016/j.annemergmed.2010.03.041 · DOI
- 10.1164/ajrccm.160.5.9806006 · DOI · Paper page
- 10.1093/clinchem/41.10.1434 · doi:10.1093/clinchem/41.10.1434
- 10.1016/j.ajem.2007.04.028 · DOI
Cited by
Carbon monoxide poisoning: pathogenesis, management, and future directions of therapy
Rose, J. J., Wang, L., Xu, Q., McTiernan, C. F., Shiva, S., Tejero, J., & Gladwin, M. T. (2017) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.201606-1275ci · Cited by 965
Metadata
- Authors
- Weaver, L. K., Churchill, S. K., Deru, K., & Cooney, D.
- Affiliations
- Division of Hyperbaric Medicine, LDS Hospital, Salt Lake City, Utah, and with Intermountain Medical Center, Murray, Utah
- SciMetrika, Durham, North Carolina.
- Venue
- Respiratory Care
- Publisher
- SAGE Publications
- Published
- 2012-07-12
- Volume
- 58
- Issue
- 2
- Pages
- 232-240
- ISSN
- 0020-1324
- Cited by
- 45
- References
- 2
- Language
- en
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: False Positive Rate of Carbon Monoxide Saturation by Pulse Oximetry of Emergency Department Patients. https://app.cochat.ai/papers/10.4187/respcare.01744
- The paper
- Weaver, L. K., Churchill, S. K., Deru, K., & Cooney, D. (2013). False Positive Rate of Carbon Monoxide Saturation by Pulse Oximetry of Emergency Department Patients. Respiratory Care. https://doi.org/10.4187/respcare.01744
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.