General review of magnetic resonance elastography

Open access
Low, Gavin·2016·World Journal of Radiology

Abstract

Magnetic resonance elastography (MRE) is an innovative imaging technique for the non-invasive quantification of the biomechanical properties of soft tissues via the direct visualization of propagating shear waves in vivo using a modified phase-contrast magnetic resonance imaging (MRI) sequence. Fundamentally, MRE employs the same physical property that physicians utilize when performing manual palpation - that healthy and diseased tissues can be differentiated on the basis of widely differing mechanical stiffness. By performing "virtual palpation", MRE is able to provide information that is beyond the capabilities of conventional morphologic imaging modalities. In an era of increasing adoption of multi-parametric imaging approaches for solving complex problems, MRE can be seamlessly incorporated into a standard MRI examination to provide a rapid, reliable and comprehensive imaging evaluation at a single patient appointment. Originally described by the Mayo Clinic in 1995, the technique represents the most accurate non-invasive method for the detection and staging of liver fibrosis and is currently performed in more than 100 centers worldwide. In this general review, the mechanical properties of soft tissues, principles of MRE, clinical applications of MRE in the liver and beyond, and limitations and future directions of this discipline -are discussed. Selected diagrams and images are provided for illustration.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.4329/wjr.v8.i1.59

Cites

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • MRE remains constrained by skull attenuation, reflection, diffraction, limited spatial coverage, and wavelength-dependent spatial resolution.

    “Still, whether externally or intrinsically driven, MRE remains constrained by skull attenuation, reflection, diffraction, limited spatial coverage, and wavelength-dependent spatial resolution (Low et al., 2016).”

    From 40 Years of diffusion MRI in the brain: From history to emerging frontiers (2026) · cites this paper as Low et al., 2016

    In its comparison with ultrasound elastography, the review says MRE offers wide-field imaging coverage, contrary to the claim that its spatial coverage is limited. The Brain and Principles of MRE sections do not describe the claimed skull attenuation, reflection, diffraction, or wavelength-dependent spatial resolution; the latter only specifies a half-wavelength boundary for measuring tissue stiffness. Page numbers are unavailable in our copy. (LLM, conf=0.75)

    Checked by a reasoning model — may be wrong.

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

No references could be resolved yet.

Research works printed but not linked: 128

  1. 22308105 DOI: 10.2174/157340511798038684] · DOI
  2. Manduca A, Oliphant TE, Dresner MA, Mahowald JL, Kruse SA, Amromin E, Felmlee JP, Greenleaf JF, Ehman RL. Magnetic resonance elastography: non-invasive mapping of tissue elasticity. Med Image Anal 2001; 5: 237-254 [PMID: 11731304 DOI: 10.1016/S1361-8415(00)00039-6] · DOI
  3. Gao L, Parker KJ, Lerner RM, Levinson SF. Imaging of the elastic properties of tissue--a review. Ultrasound Med Biol 1996; 22: 959-977 [PMID: 9004420 DOI: 10.1016/S0301-5629(96)00120-2] · DOI
  4. Parker KJ, Doyley MM, Rubens DJ. Imaging the elastic properties of tissue: the 20 year perspective. Phys Med Biol 2011; 56: R1-R29 [PMID: 21119234 DOI: 10.1088/0031-9155/56/1/r01] · DOI
  5. Vappou J. Magnetic resonance-and ultrasound imaging-based elasticity imaging methods: a review. Crit Rev Biomed Eng 2012; 40: 121-134 [PMID: 22668238 DOI: 10.1615/CritRevBiomedEng. v40.i2. · DOI
  6. Wells PN, Liang HD. Medical ultrasound: imaging of soft tissue strain and elasticity. J R Soc Interface 2011; 8: 1521-1549 [PMID: 21680780 DOI: 10.1098/rsif.2011.0054] · DOI
  7. Goss SA, Johnston RL, Dunn F. Comprehensive compilation of empirical ultrasonic properties of mammalian tissues. J Acoust Soc Am 1978; 64: 423-457 [PMID: 361793 DOI: 10.1121/1.382016] · DOI
  8. Muthupillai R, Lomas DJ, Rossman PJ, Greenleaf JF, Manduca A, Ehman RL. Magnetic resonance elastography by direct visualization of propagating acoustic strain waves. Science 1995; 269: 1854-1857 [PMID: 7569924 DOI: 10.1126/science.7569924] · DOI · Paper page
  9. Chenevert TL, Skovoroda AR, O'Donnell M, Emelianov SY. Elasticity reconstructive imaging by means of stimulated echo MRI. Magn Reson Med 1998; 39: 482-490 [PMID: 9498605 DOI: 10.1002/mrm.1910390319] · DOI
  10. Plewes DB, Betty I, Urchuk SN, Soutar I. Visualizing tissue compliance with MR imaging. J Magn Reson Imaging 1995; 5: 733-738 [PMID: 8748495 DOI: 10.1002/jmri.1880050620] · DOI
  11. Osman NF. Detecting stiff masses using strain-encoded (SENC) imaging. Magn Reson Med 2003; 49: 605-608 [PMID: 12594769 DOI: 10.1002/mrm.10376] · DOI
  12. Venkatesh SK, Yin M, Ehman RL. Magnetic resonance elasto- graphy of liver: technique, analysis, and clinical applications. J Magn Reson Imaging 2013; 37: 544-555 [PMID: 23423795 DOI: 10.1002/jmri.23731] · DOI
  13. Yin M, Talwalkar JA, Glaser KJ, Manduca A, Grimm RC, Rossman PJ, Fidler JL, Ehman RL. Assessment of hepatic fibrosis with magnetic resonance elastography. Clin Gastroenterol Hepatol 2007; 5: 1207-1213.e2 [PMID: 17916548 DOI: 10.1016/ j.cgh.2007.06.012] · DOI
  14. Rouvière O, Yin M, Dresner MA, Rossman PJ, Burgart LJ, Fidler JL, Ehman RL. MR elastography of the liver: preliminary results. Radiology 2006; 240: 440-448 [PMID: 16864671 DOI: 10.1148/ radiol.2402050606] · DOI
  15. Uffmann K, Ladd ME. Actuation systems for MR elastography: design and applications. IEEE Eng Med Biol Mag 2008; 27: 28-34 [PMID: 18519179 DOI: 10.1109/emb.2007.910268] · DOI
  16. Di Ieva A, Grizzi F, Rognone E, Tse ZT, Parittotokkaporn T, Rodriguez Y Baena F, Tschabitscher M, Matula C, Trattnig S, Rodriguez Y Baena R. Magnetic resonance elastography: a general overview of its current and future applications in brain imaging. Neurosurg Rev 2010; 33: 137-145; discussion 145 [PMID: 20195674 DOI: 10.1007/s10143-010-0249-6] · DOI
  17. Yin M, Chen J, Glaser KJ, Talwalkar JA, Ehman RL. Abdominal magnetic resonance elastography. Top Magn Reson Imaging 2009; 20: 79-87 [PMID: 20010062 DOI: 10.1097/RMR.0b013e3181c4737e] · DOI
  18. Huwart L, Peeters F, Sinkus R, Annet L, Salameh N, ter Beek LC, Horsmans Y, Van Beers BE. Liver fibrosis: non-invasive assessment with MR elastography. NMR Biomed 2006; 19: 173-179 [PMID: 16521091 DOI: 10.1002/nbm.1030] · DOI
  19. Huwart L, Sempoux C, Salameh N, Jamart J, Annet L, Sinkus R, Peeters F, ter Beek LC, Horsmans Y, Van Beers BE. Liver fibrosis: noninvasive assessment with MR elastography versus aspartate aminotransferase-to-platelet ratio index. Radiology 2007; 245: 458-466 [PMID: 17940304 DOI: 10.1148/radiol.2452061673] · DOI
  20. Huwart L, Sempoux C, Vicaut E, Salameh N, Annet L, Danse E, Peeters F, ter Beek LC, Rahier J, Sinkus R, Horsmans Y, Van Beers BE. Magnetic resonance elastography for the noninvasive staging of liver fibrosis. Gastroenterology 2008; 135: 32-40 [PMID: 18471441 DOI: 10.1053/j.gastro.2008.03.076] · DOI
  21. Godfrey EM, Patterson AJ, Priest AN, Davies SE, Joubert I, Krishnan AS, Griffin N, Shaw AS, Alexander GJ, Allison ME, Griffiths WJ, Gimson AE, Lomas DJ. A comparison of MR elastography and 31P MR spectroscopy with histological staging of liver fibrosis. Eur Radiol 2012; 22: 2790-2797 [PMID: 22752441 DOI: 10.1007/s00330-012-2527-x] · DOI
  22. Singh S, Venkatesh SK, Wang Z, Miller FH, Motosugi U, Low RN, Hassanein T, Asbach P, Godfrey EM, Yin M, Chen J, Keaveny AP, Bridges M, Bohte A, Murad MH, Lomas DJ, Talwalkar JA, Ehman RL. Diagnostic performance of magnetic resonance elastography in staging liver fibrosis: a systematic review and meta-analysis of individual participant data. Clin Gastroenterol Hepatol 2015; 13: 440-451.e6 [PMID: 25305349 DOI: 10.1016/j.cgh.2014.09.046] · DOI
  23. Su LN, Guo SL, Li BX, Yang P. Diagnostic value of magnetic resonance elastography for detecting and staging of hepatic fibrosis: a meta-analysis. Clin Radiol 2014; 69: e545-e552 [PMID: 25300557 DOI: 10.1016/j.crad.2014.09.001] · DOI
  24. Lee VS, Miller FH, Omary RA, Wang Y, Ganger DR, Wang E, Rao S, Levitsky J. Magnetic resonance elastography and biomarkers to assess fibrosis from recurrent hepatitis C in liver transplant recipients. Transplantation 2011; 92: 581-586 [PMID: 21822174 DOI: 10.1097/TP.0b013e31822805fa] · DOI
  25. Wang Y, Ganger DR, Levitsky J, Sternick LA, McCarthy RJ, Chen ZE, Fasanati CW, Bolster B, Shah S, Zuehlsdorff S, Omary RA, Ehman RL, Miller FH. Assessment of chronic hepatitis and fibrosis: comparison of MR elastography and diffusion-weighted imaging. AJR Am J Roentgenol 2011; 196: 553-561 [PMID: 21343496 DOI: 10.2214/ajr.10.4580] · DOI
  26. Rustogi R, Horowitz J, Harmath C, Wang Y, Chalian H, Ganger DR, Chen ZE, Bolster BD, Shah S, Miller FH. Accuracy of MR elastography and anatomic MR imaging features in the diagnosis of severe hepatic fibrosis and cirrhosis. J Magn Reson Imaging 2012; 35: 1356-1364 [PMID: 22246952 DOI: 10.1002/jmri.23585] · DOI
  27. Park HS, Kim YJ, Yu MH, Choe WH, Jung SI, Jeon HJ. Three- Tesla magnetic resonance elastography for hepatic fibrosis: comparison with diffusion-weighted imaging and gadoxetic acid- enhanced magnetic resonance imaging. World J Gastroenterol 2014; 20: 17558-17567 [PMID: 25516671 DOI: 10.3748/wjg.v20. i46.17558] · DOI
  28. Chen J, Talwalkar JA, Yin M, Glaser KJ, Sanderson SO, Ehman RL. Early detection of nonalcoholic steatohepatitis in patients with nonalcoholic fatty liver disease by using MR elastography. Radiology 2011; 259: 749-756 [PMID: 21460032 DOI: 10.1148/ radiol.11101942] · DOI
  29. Loomba R, Wolfson T, Ang B, Hooker J, Behling C, Peterson M, Valasek M, Lin G, Brenner D, Gamst A, Ehman R, Sirlin C. Magnetic resonance elastography predicts advanced fibrosis in patients with nonalcoholic fatty liver disease: a prospective study. Hepatology 2014; 60: 1920-1928 [PMID: 25103310 DOI: 10.1002/hep.27362] · DOI
  30. Loomba R, Sirlin CB, Ang B, Bettencourt R, Jain R, Salotti J, Soaft L, Hooker J, Kono Y, Bhatt A, Hernandez L, Nguyen P, Noureddin M, Haufe W, Hooker C, Yin M, Ehman R, Lin GY, Valasek MA, Brenner DA, Richards L. Ezetimibe for the treatment of nonalcoholic steatohepatitis: assessment by novel magnetic resonance imaging and magnetic resonance elastography in a randomized trial (MOZART trial). Hepatology 2015; 61: 1239-1250 [PMID: 25482832 DOI: 10.1002/hep.27647] · DOI
  31. Hines CD, Bley TA, Lindstrom MJ, Reeder SB. Repeatability of magnetic resonance elastography for quantification of hepatic stiffness. J Magn Reson Imaging 2010; 31: 725-731 [PMID: 20187219 DOI: 10.1002/jmri.22066] · DOI
  32. Shire NJ, Yin M, Chen J, Railkar RA, Fox-Bosetti S, Johnson SM, Beals CR, Dardzinski BJ, Sanderson SO, Talwalkar JA, Ehman RL. Test-retest repeatability of MR elastography for noninvasive liver fibrosis assessment in hepatitis C. J Magn Reson Imaging 2011; 34: 947-955 [PMID: 21751289 DOI: 10.1002/jmri.22716] · DOI
  33. Hines CD, Lindstrom MJ, Varma AK, Reeder SB. Effects of postprandial state and mesenteric blood flow on the repeatability of MR elastography in asymptomatic subjects. J Magn Reson Imaging 2011; 33: 239-244 [PMID: 21182146 DOI: 10.1002/jmri.22354] · DOI
  34. Motosugi U, Ichikawa T, Sano K, Sou H, Muhi A, Koshiishi T, Ehman RL, Araki T. Magnetic resonance elastography of the liver: preliminary results and estimation of inter-rater reliability. Jpn J Radiol 2010; 28: 623-627 [PMID: 20972864 DOI: 10.1007/ s11604-010-0478-1] · DOI
  35. Shi Y, Guo Q, Xia F, Sun J, Gao Y. Short-and midterm repea- tability of magnetic resonance elastography in healthy volunteers at 3.0 T. Magn Reson Imaging 2014; 32: 665-670 [PMID: 24650683 DOI: 10.1016/j.mri.2014.02.018] · DOI
  36. Venkatesh SK, Wang G, Teo LL, Ang BW. Magnetic resonance elastography of liver in healthy Asians: normal liver stiffness quantification and reproducibility assessment. J Magn Reson Imaging 2014; 39: 1-8 [PMID: 24123300 DOI: 10.1002/jmri.24084] · DOI
  37. Serai SD, Yin M, Wang H, Ehman RL, Podberesky DJ. Cross- vendor validation of liver magnetic resonance elastography. Abdom Imaging 2015; 40: 789-794 [PMID: 25476489 DOI: 10.1007/ s00261-014-0282-y] · DOI
  38. Hallinan JT, Alsaif HS, Wee A, Venkatesh SK. Magnetic reso- nance elastography of liver: influence of intravenous gadolinium administration on measured liver stiffness. Abdom Imaging 2015; 40: 783-788 [PMID: 25331568 DOI: 10.1007/s00261-014-0275-x] · DOI
  39. Yin M, Talwalkar JA, Glaser KJ, Venkatesh SK, Chen J, Manduca A, Ehman RL. Dynamic postprandial hepatic stiffness augmentation assessed with MR elastography in patients with chronic liver disease. AJR Am J Roentgenol 2011; 197: 64-70 [PMID: 21701012 DOI: 10.2214/ajr.10.5989] · DOI
  40. Jajamovich GH, Dyvorne H, Donnerhack C, Taouli B. Quan- titative liver MRI combining phase contrast imaging, elastography, and DWI: assessment of reproducibility and postprandial effect at 3.0 T. PLoS One 2014; 9: e97355 [PMID: 24840288 DOI: 10.1371/ journal.pone.0097355] · DOI
  41. Perumpail RB, Levitsky J, Wang Y, Lee VS, Karp J, Jin N, Yang GY, Bolster BD, Shah S, Zuehlsdorff S, Nemcek AA, Larson AC, Miller FH, Omary RA. MRI-guided biopsy to correlate tissue specimens with MR elastography stiffness readings in liver transplants. Acad Radiol 2012; 19: 1121-1126 [PMID: 22877987 DOI: 10.1016/j.acra.2012.05.011] · DOI
  42. Venkatesh SK, Yin M, Glockner JF, Takahashi N, Araoz PA, Talwalkar JA, Ehman RL. MR elastography of liver tumors: preliminary results. AJR Am J Roentgenol 2008; 190: 1534-1540 [PMID: 18492904 DOI: 10.2214/ajr.07.3123] · DOI
  43. Hennedige TP, Hallinan JT, Leung FP, Teo LL, Iyer S, Wang G, Chang S, Madhavan KK, Wee A, Venkatesh SK.Comparison of magnetic resonance elastography and diffusion-weighted imaging for differentiating benign and malignant liver lesions. Eur Radiol 2015; Epub ahead of print [PMID: 26032879 DOI: 10.1007/ s00330-015-3835-8] · DOI
  44. Ichikawa S, Motosugi U, Morisaka H, Sano K, Ichikawa T, Tatsumi A, Enomoto N, Matsuda M, Fujii H, Onishi H. Comparison of the diagnostic accuracies of magnetic resonance elastography and transient elastography for hepatic fibrosis. Magn Reson Imaging 2015; 33: 26-30 [PMID: 25308096 DOI: 10.1016/ j.mri.2014.10.003] · DOI
  45. Yoon JH, Lee JM, Joo I, Lee ES, Sohn JY, Jang SK, Lee KB, Han JK, Choi BI. Hepatic fibrosis: prospective comparison of MR elastography and US shear-wave elastography for evaluation. Radiology 2014; 273: 772-782 [PMID: 25007047 DOI: 10.1148/ radiol.14132000] · DOI
  46. Mannelli L, Godfrey E, Joubert I, Patterson AJ, Graves MJ, Gallagher FA, Lomas DJ. MR elastography: Spleen stiffness measurements in healthy volunteers--preliminary experience. AJR Am J Roentgenol 2010; 195: 387-392 [PMID: 20651194 DOI: 10.2214/ajr.09.3390] · DOI
  47. Talwalkar JA, Yin M, Venkatesh S, Rossman PJ, Grimm RC, Manduca A, Romano A, Kamath PS, Ehman RL. Feasibility of in vivo MR elastographic splenic stiffness measurements in the assessment of portal hypertension. AJR Am J Roentgenol 2009; 193: 122-127 [PMID: 19542403 DOI: 10.2214/ajr.07.3504] · DOI
  48. Shin SU, Lee JM, Yu MH, Yoon JH, Han JK, Choi BI, Glaser KJ, Ehman RL. Prediction of esophageal varices in patients with cirrhosis: usefulness of three-dimensional MR elastography with echo-planar imaging technique. Radiology 2014; 272: 143-153 [PMID: 24620910 DOI: 10.1148/radiol.14130916] · DOI
  49. Guo J, Büning C, Schott E, Kröncke T, Braun J, Sack I, Althoff C. In vivo abdominal magnetic resonance elastography for the assessment of portal hypertension before and after transjugular intrahepatic portosystemic shunt implantation. Invest Radiol 2015; 50: 347-351 [PMID: 25599282 DOI: 10.1097/rli.0000000000000136] · DOI
  50. Ronot M, Lambert S, Elkrief L, Doblas S, Rautou PE, Castera L, Vilgrain V, Sinkus R, Van Beers BE, Garteiser P. Assessment of portal hypertension and high-risk oesophageal varices with liver and spleen three-dimensional multifrequency MR elastography in liver cirrhosis. Eur Radiol 2014; 24: 1394-1402 [PMID: 24626745 DOI: 10.1007/s00330-014-3124-y] · DOI
  51. Low G, Owen NE, Joubert I, Patterson AJ, Graves MJ, Glaser KJ, Alexander GJ, Lomas DJ. Reliability of magnetic resonance elastography using multislice two-dimensional spin-echo echo- planar imaging (SE-EPI) and three-dimensional inversion reconstruction for assessing renal stiffness. J Magn Reson Imaging 2015; 42: 844-850 [PMID: 25537823 DOI: 10.1002/jmri.24826] · DOI
  52. Rouvière O, Souchon R, Pagnoux G, Ménager JM, Chapelon JY. Magnetic resonance elastography of the kidneys: feasibility and reproducibility in young healthy adults. J Magn Reson Imaging 2011; 34: 880-886 [PMID: 21769970 DOI: 10.1002/jmri.22670] · DOI
  53. Lee CU, Glockner JF, Glaser KJ, Yin M, Chen J, Kawashima A, Kim B, Kremers WK, Ehman RL, Gloor JM. MR elastography in renal transplant patients and correlation with renal allograft biopsy: a feasibility study. Acad Radiol 2012; 19: 834-841 [PMID: 22503893 DOI: 10.1016/j.acra.2012.03.003] · DOI
  54. Low G, Owen NE, Joubert I, Patterson AJ, Graves MJ, Alexander GJ, Lomas DJ. Magnetic resonance elastography in the detection of hepatorenal syndrome in patients with cirrhosis and ascites. Eur Radiol 2015; 25: 2851-2858 [PMID: 25903705 DOI: 10.1007/ s00330-015-3723-2] · DOI
  55. Lorenzen J, Sinkus R, Lorenzen M, Dargatz M, Leussler C, Röschmann P, Adam G. MR elastography of the breast: preliminary clinical results. Rofo 2002; 174: 830-834 [PMID: 12101471 DOI: 10.1055/s-2002-32690] · DOI
  56. Sinkus R, Siegmann K, Xydeas T, Tanter M, Claussen C, Fink M. MR elastography of breast lesions: understanding the solid/liquid duality can improve the specificity of contrast-enhanced MR mammography. Magn Reson Med 2007; 58: 1135-1144 [PMID: 17969009 DOI: 10.1002/mrm.21404] · DOI
  57. Xydeas T, Siegmann K, Sinkus R, Krainick-Strobel U, Miller S, Claussen CD. Magnetic resonance elastography of the breast: correlation of signal intensity data with viscoelastic properties. Invest Radiol 2005; 40: 412-420 [PMID: 15973132 DOI: 10.1097/01.rli.0000166940.72971.4a · DOI
  58. Sinkus R, Tanter M, Xydeas T, Catheline S, Bercoff J, Fink M. Viscoelastic shear properties of in vivo breast lesions measured by MR elastography. Magn Reson Imaging 2005; 23: 159-165 [PMID: 15833607 DOI: 10.1016/j.mri.2004.11.060] · DOI
  59. Sinkus R, Tanter M, Catheline S, Lorenzen J, Kuhl C, Sondermann E, Fink M. Imaging anisotropic and viscous properties of breast tissue by magnetic resonance-elastography. Magn Reson Med 2005; 53: 372-387 [PMID: 15678538 DOI: 10.1002/mrm.20355] · DOI
  60. Lorenzen J, Sinkus R, Biesterfeldt M, Adam G. Menstrual- cycle dependence of breast parenchyma elasticity: estimation with magnetic resonance elastography of breast tissue during the menstrual cycle. Invest Radiol 2003; 38: 236-240 [PMID: 12649648 DOI: 10.1097/01.rli.0000059544.18910.bd] · DOI
  61. McKnight AL, Kugel JL, Rossman PJ, Manduca A, Hartmann LC, Ehman RL. MR elastography of breast cancer: preliminary results. AJR Am J Roentgenol 2002; 178: 1411-1417 [PMID: 12034608 DOI: 10.2214/ajr.178.6.1781411] · DOI
  62. Lorenzen J, Sinkus R, Schrader D, Lorenzen M, Leussler C, Dargatz M, Röschmann P. [Imaging of breast tumors using MR elastography]. Rofo 2001; 173: 12-17 [PMID: 11225411 DOI: 10.1055/s-2001-10233] · DOI
  63. Sinkus R, Lorenzen J, Schrader D, Lorenzen M, Dargatz M, Holz D. High-resolution tensor MR elastography for breast tumour detection. Phys Med Biol 2000; 45: 1649-1664 [PMID: 10870716 DOI: 10.1088/0031-9155/45/6/317] · DOI
  64. Plewes DB, Bishop J, Samani A, Sciarretta J. Visualization and quantification of breast cancer biomechanical properties with magnetic resonance elastography. Phys Med Biol 2000; 45: 1591-1610 [PMID: 10870713 DOI: 10.1088/0031-9155/45/6/314] · DOI
  65. Siegmann KC, Xydeas T, Sinkus R, Kraemer B, Vogel U, Claussen CD. Diagnostic value of MR elastography in addition to contrast-enhanced MR imaging of the breast-initial clinical results. Eur Radiol 2010; 20: 318-325 [PMID: 19727753 DOI: 10.1007/ s00330-009-1566-4] · DOI
  66. Zhang J, Green MA, Sinkus R, Bilston LE. Viscoelastic properties of human cerebellum using magnetic resonance elastography. J Biomech 2011; 44: 1909-1913 [PMID: 21565346 DOI: 10.1016/ j.jbiomech.2011.04.034] · DOI
  67. Green MA, Bilston LE, Sinkus R. In vivo brain viscoelastic properties measured by magnetic resonance elastography. NMR Biomed 2008; 21: 755-764 [PMID: 18457350 DOI: 10.1002/ nbm.1254] · DOI
  68. Kruse SA, Rose GH, Glaser KJ, Manduca A, Felmlee JP, Jack CR, Ehman RL. Magnetic resonance elastography of the brain. Neuroimage 2008; 39: 231-237 [PMID: 17913514 DOI: 10.1016/ j.neuroimage.2007.08.030] · DOI
  69. Xu L, Lin Y, Xi ZN, Shen H, Gao PY. Magnetic resonance elasto- graphy of the human brain: a preliminary study. Acta Radiol 2007; 48: 112-115 [PMID: 17325935 DOI: 10.1080/02841850601026401] · DOI
  70. Wuerfel J, Paul F, Beierbach B, Hamhaber U, Klatt D, Papazoglou S, Zipp F, Martus P, Braun J, Sack I. MR-elastography reveals degradation of tissue integrity in multiple sclerosis. Neuroimage 2010; 49: 2520-2525 [PMID: 19539039 DOI: 10.1016/j.neuroimag e.2009.06.018] · DOI
  71. Streitberger KJ, Sack I, Krefting D, Pfüller C, Braun J, Paul F, Wuerfel J. Brain viscoelasticity alteration in chronic-progressive multiple sclerosis. PLoS One 2012; 7: e29888 [PMID: 22276134 DOI: 10.1371/journal.pone.0029888] · DOI
  72. Schregel K, Wuerfel E, Garteiser P, Gemeinhardt I, Prozorovski T, Aktas O, Merz H, Petersen D, Wuerfel J, Sinkus R. Demyelination reduces brain parenchymal stiffness quantified in vivo by magnetic resonance elastography. Proc Natl Acad Sci USA 2012; 109: 6650-6655 [PMID: 22492966 DOI: 10.1073/pnas.1200151109] · DOI
  73. Murphy MC, Huston J, Jack CR, Glaser KJ, Manduca A, Felmlee JP, Ehman RL. Decreased brain stiffness in Alzheimer's disease determined by magnetic resonance elastography. J Magn Reson Imaging 2011; 34: 494-498 [PMID: 21751286 DOI: 10.1002/ jmri.22707] · DOI
  74. Sack I, Streitberger KJ, Krefting D, Paul F, Braun J. The influence of physiological aging and atrophy on brain viscoelastic properties in humans. PLoS One 2011; 6: e23451 [PMID: 21931599 DOI: 10.1371/journal.pone.0023451] · DOI
  75. Boulet T, Kelso ML, Othman SF. Microscopic magnetic reso- nance elastography of traumatic brain injury model. J Neurosci Methods 2011; 201: 296-306 [PMID: 21871490 DOI: 10.1016/ j.jneumeth.2011.08.019] · DOI
  76. Freimann FB, Streitberger KJ, Klatt D, Lin K, McLaughlin J, Braun J, Sprung C, Sack I. Alteration of brain viscoelasticity after shunt treatment in normal pressure hydrocephalus. Neuroradiology 2012; 54: 189-196 [PMID: 21538046 DOI: 10.1007/s00234-011-0871-1] · DOI
  77. Murphy MC, Huston J, Glaser KJ, Manduca A, Meyer FB, Lanzino G, Morris JM, Felmlee JP, Ehman RL. Preoperative assessment of meningioma stiffness using magnetic resonance elastography. J Neurosurg 2013; 118: 643-648 [PMID: 23082888 DOI: 10.3171/2012.9.jns12519] · DOI
  78. Xu L, Lin Y, Han JC, Xi ZN, Shen H, Gao PY. Magnetic resonance elastography of brain tumors: preliminary results. Acta Radiol 2007; 48: 327-330 [PMID: 17453505 DOI: 10.1080/02841850701 199967] · DOI
  79. Reiss-Zimmermann M, Streitberger KJ, Sack I, Braun J, Arlt F, Fritzsch D, Hoffmann KT. High Resolution Imaging of Viscoelastic Properties of Intracranial Tumours by Multi-Frequency Magnetic Resonance Elastography. Clin Neuroradiol 2014; Epub ahead of print [PMID: 24916129 DOI: 10.1007/s00062-014-0311-9] · DOI
  80. Jenkyn TR, Ehman RL, An KN. Noninvasive muscle tension measurement using the novel technique of magnetic resonance elastography (MRE). J Biomech 2003; 36: 1917-1921 [PMID: 14614945 DOI: 10.1016/S0021-9290(03)00005-8] · DOI
  81. Ringleb SI, Bensamoun SF, Chen Q, Manduca A, An KN, Ehman RL. Applications of magnetic resonance elastography to healthy and pathologic skeletal muscle. J Magn Reson Imaging 2007; 25: 301-309 [PMID: 17260391 DOI: 10.1002/jmri.20817] · DOI
  82. Debernard L, Robert L, Charleux F, Bensamoun SF. A possible clinical tool to depict muscle elasticity mapping using magnetic resonance elastography. Muscle Nerve 2013; 47: 903-908 [PMID: 23512304 DOI: 10.1002/mus.23678] · DOI
  83. Bensamoun SF, Ringleb SI, Chen Q, Ehman RL, An KN, Brennan M. Thigh muscle stiffness assessed with magnetic resonance elastography in hyperthyroid patients before and after medical treatment. J Magn Reson Imaging 2007; 26: 708-713 [PMID: 17729336 DOI: 10.1002/jmri.21073] · DOI
  84. Brauck K, Galbán CJ, Maderwald S, Herrmann BL, Ladd ME. Changes in calf muscle elasticity in hypogonadal males before and after testosterone substitution as monitored by magnetic resonance elastography. Eur J Endocrinol 2007; 156: 673-678 [PMID: 17535867 DOI: 10.1530/eje-06-0694] · DOI
  85. Heers G, Jenkyn T, Dresner MA, Klein MO, Basford JR, Kaufman KR, Ehman RL, An KN. Measurement of muscle activity with magnetic resonance elastography. Clin Biomech (Bristol, Avon) 2003; 18: 537-542 [PMID: 12828903 DOI: 10.1016/ S0268-0033(03)00070-6] · DOI
  86. Debernard L, Robert L, Charleux F, Bensamoun SF. Analysis of thigh muscle stiffness from childhood to adulthood using magnetic resonance elastography (MRE) technique. Clin Biomech (Bristol, Avon) 2011; 26: 836-840 [PMID: 21571411 DOI: 10.1016/j.clinbi omech.2011.04.004] · DOI
  87. Domire ZJ, McCullough MB, Chen Q, An KN. Feasibility of using magnetic resonance elastography to study the effect of aging on shear modulus of skeletal muscle. J Appl Biomech 2009; 25: 93-97 [PMID: 19299834]
  88. Basford JR, Jenkyn TR, An KN, Ehman RL, Heers G, Kaufman KR. Evaluation of healthy and diseased muscle with magnetic resonance elastography. Arch Phys Med Rehabil 2002; 83: 1530-1536 [PMID: 12422320 DOI: 10.1053/apmr.2002.35472] · DOI
  89. McCullough MB, Domire ZJ, Reed AM, Amin S, Ytterberg SR, Chen Q, An KN. Evaluation of muscles affected by myositis using magnetic resonance elastography. Muscle Nerve 2011; 43: 585-590 [PMID: 21319167 DOI: 10.1002/mus.21923] · DOI
  90. Hong SH, Hong SJ, Yoon JS, Oh CH, Cha JG, Kim HK, Bolster B Jr. Magnetic resonance elastography (MRE) for measurement of muscle stiffness of the shoulder: feasibility with a 3 T MRI system. Acta Radiol 2015; Epub ahead of print [PMID: 25711231 DOI: 10.1177/0284185115571987] · DOI
  91. Kolipaka A, Aggarwal SR, McGee KP, Anavekar N, Manduca A, Ehman RL, Araoz PA. Magnetic resonance elastography as a method to estimate myocardial contractility. J Magn Reson Imaging 2012; 36: 120-127 [PMID: 22334349 DOI: 10.1002/ jmri.23616] · DOI
  92. Kolipaka A, McGee KP, Manduca A, Anavekar N, Ehman RL, Araoz PA. In vivo assessment of MR elastography-derived effective end-diastolic myocardial stiffness under different loading conditions. J Magn Reson Imaging 2011; 33: 1224-1228 [PMID: 21509882 DOI: 10.1002/jmri.22531] · DOI
  93. Elgeti T, Beling M, Hamm B, Braun J, Sack I. Cardiac magnetic resonance elastography: toward the diagnosis of abnormal myo- cardial relaxation. Invest Radiol 2010; 45: 782-787 [PMID: 20829709 DOI: 10.1097/RLI.0b013e3181ec4b63] · DOI
  94. Elgeti T, Laule M, Kaufels N, Schnorr J, Hamm B, Samani A, Braun J, Sack I. Cardiac MR elastography: comparison with left ventricular pressure measurement. J Cardiovasc Magn Reson 2009; 11: 44 [PMID: 19900266 DOI: 10.1186/1532-429x-11-44] · DOI
  95. Elgeti T, Rump J, Hamhaber U, Papazoglou S, Hamm B, Braun J, Sack I. Cardiac magnetic resonance elastography. Initial results. Invest Radiol 2008; 43: 762-772 [PMID: 18923255 DOI: 10.1097/ RLI.0b013e3181822085] · DOI
  96. Sack I, Rump J, Elgeti T, Samani A, Braun J. MR elastography of the human heart: noninvasive assessment of myocardial elasticity changes by shear wave amplitude variations. Magn Reson Med 2009; 61: 668-677 [PMID: 19097236 DOI: 10.1002/mrm.21878] · DOI
  97. Shi Y, Glaser KJ, Venkatesh SK, Ben-Abraham EI, Ehman RL. Feasibility of using 3D MR elastography to determine pancreatic stiffness in healthy volunteers. J Magn Reson Imaging 2015; 41: 369-375 [PMID: 24497052 DOI: 10.1002/jmri.24572] · DOI · Paper page
  98. Jiang X, Asbach P, Streitberger KJ, Thomas A, Hamm B, Braun J, Sack I, Guo J. In vivo high-resolution magnetic resonance elasto- graphy of the uterine corpus and cervix. Eur Radiol 2014; 24: 3025-3033 [PMID: 25038856 DOI: 10.1007/s00330-014-3305-8] · DOI
  99. Arani A, Plewes D, Krieger A, Chopra R. The feasibility of endorectal MR elastography for prostate cancer localization. Magn Reson Med 2011; 66: 1649-1657 [PMID: 21574182 DOI: 10.1002/ mrm.22967] 100 · DOI
  100. Thörmer G, Reiss-Zimmermann M, Otto J, Hoffmann KT, Moche M, Garnov N, Kahn T, Busse H. Novel technique for MR elastography of the prostate using a modified standard endorectal coil as actuator. J Magn Reson Imaging 2013; 37: 1480-1485 [PMID: 23055397 DOI: 10.1002/jmri.23850] 101 · DOI
  101. Sahebjavaher RS, Baghani A, Honarvar M, Sinkus R, Salcudean SE. Transperineal prostate MR elastography: initial in vivo results. Magn Reson Med 2013; 69: 411-420 [PMID: 22505273 DOI: 10.1002/mrm.24268] 102 · DOI
  102. Arani A, Da Rosa M, Ramsay E, Plewes DB, Haider MA, Chopra R. Incorporating endorectal MR elastography into multi-parametric MRI for prostate cancer imaging: Initial feasibility in volunteers. J Magn Reson Imaging 2013; 38: 1251-1260 [PMID: 23408516 DOI: 10.1002/jmri.24028] 103 · DOI
  103. Li S, Chen M, Wang W, Zhao W, Wang J, Zhao X, Zhou C. A feasibility study of MR elastography in the diagnosis of prostate cancer at 3.0T. Acta Radiol 2011; 52: 354-358 [PMID: 21498375 DOI: 10.1258/ar.2010.100276] 104 · DOI
  104. Arani A, Plewes D, Chopra R. Transurethral prostate magnetic resonance elastography: prospective imaging requirements. Magn Reson Med 2011; 65: 340-349 [PMID: 20882673 DOI: 10.1002/ mrm.22633] 105 · DOI
  105. Chopra R, Arani A, Huang Y, Musquera M, Wachsmuth J, Bronskill M, Plewes D. In vivo MR elastography of the prostate gland using a transurethral actuator. Magn Reson Med 2009; 62: 665-671 [PMID: 19572390 DOI: 10.1002/mrm.22038] 106 · DOI
  106. Kemper J, Sinkus R, Lorenzen J, Nolte-Ernsting C, Stork A, Adam G. MR elastography of the prostate: initial in-vivo application. Rofo 2004; 176: 1094-1099 [PMID: 15346284 DOI: 10.1055/s-2004-813279] 107 · DOI
  107. Kolipaka A, Woodrum D, Araoz PA, Ehman RL. MR elastography of the in vivo abdominal aorta: a feasibility study for comparing aortic stiffness between hypertensives and normotensives. J Magn Reson Imaging 2012; 35: 582-586 [PMID: 22045617 DOI: 10.1002/jmri.22866] 108 · DOI
  108. Woodrum DA, Herrmann J, Lerman A, Romano AJ, Lerman LO, Ehman RL. Phase-contrast MRI-based elastography technique detects early hypertensive changes in ex vivo porcine aortic wall. J Magn Reson Imaging 2009; 29: 583-587 [PMID: 19243040 DOI: 10.1002/jmri.21702] 109 · DOI
  109. Xu L, Chen J, Glaser KJ, Yin M, Rossman PJ, Ehman RL. MR elastography of the human abdominal aorta: a preliminary study. J Magn Reson Imaging 2013; 38: 1549-1553 [PMID: 23371244 DOI: 10.1002/jmri.24056] 110 · DOI
  110. Damughatla AR, Raterman B, Sharkey-Toppen T, Jin N, Simonetti OP, White RD, Kolipaka A. Quantification of aortic stiffness using MR elastography and its comparison to MRI-based pulse wave velocity. J Magn Reson Imaging 2015; 41: 44-51 [PMID: 24243654 DOI: 10.1002/jmri.24506] 111 · DOI
  111. Mariappan YK, Kolipaka A, Manduca A, Hubmayr RD, Ehman RL, Araoz P, McGee KP. Magnetic resonance elastography of the lung parenchyma in an in situ porcine model with a noninvasive mechanical driver: correlation of shear stiffness with trans- respiratory system pressures. Magn Reson Med 2012; 67: 210-217 [PMID: 21590723 DOI: 10.1002/mrm.22976] 112 · DOI
  112. Mariappan YK, Glaser KJ, Hubmayr RD, Manduca A, Ehman RL, McGee KP. MR elastography of human lung parenchyma: technical development, theoretical modeling and in vivo validation. J Magn Reson Imaging 2011; 33: 1351-1361 [PMID: 21591003 DOI: 10.1002/jmri.22550] 113 · DOI
  113. McGee KP, Hubmayr RD, Levin D, Ehman RL. Feasibility of quantifying the mechanical properties of lung parenchyma in a small-animal model using (1)H magnetic resonance elastography (MRE). J Magn Reson Imaging 2009; 29: 838-845 [PMID: 19306407 DOI: 10.1002/jmri.21720] 114 · DOI
  114. McGee KP, Hubmayr RD, Ehman RL. MR elastography of the lung with hyperpolarized 3He. Magn Reson Med 2008; 59: 14-18 [PMID: 18058936 DOI: 10.1002/mrm.21465] 115 · DOI
  115. Goss BC, McGee KP, Ehman EC, Manduca A, Ehman RL. Magnetic resonance elastography of the lung: technical feasibility. Magn Reson Med 2006; 56: 1060-1066 [PMID: 17036283 DOI: 10.1002/mrm.21053] 116 · DOI
  116. Yeung DK, Bhatia KS, Lee YY, King AD, Garteiser P, Sinkus R, Ahuja AT. MR elastography of the head and neck: driver design and initial results. Magn Reson Imaging 2013; 31: 624-629 [PMID: 23164497 DOI: 10.1016/j.mri.2012.09.008] 117 · DOI
  117. Bahn MM, Brennan MD, Bahn RS, Dean DS, Kugel JL, Ehman RL. Development and application of magnetic resonance elastography of the normal and pathological thyroid gland in vivo. J Magn Reson Imaging 2009; 30: 1151-1154 [PMID: 19856448 DOI: 10.1002/jmri.21963] 118 · DOI
  118. Huwart L, Salameh N, ter Beek L, Vicaut E, Peeters F, Sinkus R, Van Beers BE. MR elastography of liver fibrosis: preliminary results comparing spin-echo and echo-planar imaging. Eur Radiol 2008; 18: 2535-2541 [PMID: 18504591 DOI: 10.1007/ s00330-008-1051-5] 119 · DOI
  119. Herzka DA, Kotys MS, Sinkus R, Pettigrew RI, Gharib AM. Magnetic resonance elastography in the liver at 3 Tesla using a second harmonic approach. Magn Reson Med 2009; 62: 284-291 [PMID: 19449374 DOI: 10.1002/mrm.21956] 120 · DOI
  120. Hamhaber U, Klatt D, Papazoglou S, Hollmann M, Stadler J, Sack I, Bernarding J, Braun J. In vivo magnetic resonance elastography of human brain at 7 T and 1.5 T. J Magn Reson Imaging 2010; 32: 577-583 [PMID: 20815054 DOI: 10.1002/jmri.22294] 121 · DOI
  121. Papazoglou S, Hirsch S, Braun J, Sack I. Multifrequency inversion in magnetic resonance elastography. Phys Med Biol 2012; 57: 2329-2346 [PMID: 22460134 DOI: 10.1088/0031-9155/57/8/2329] 122 · DOI
  122. Asbach P, Klatt D, Hamhaber U, Braun J, Somasundaram R, Hamm B, Sack I. Assessment of liver viscoelasticity using multifrequency MR elastography. Magn Reson Med 2008; 60: 373-379 [PMID: 18666132 DOI: 10.1002/mrm.21636] 123 · DOI
  123. Asbach P, Klatt D, Schlosser B, Biermer M, Muche M, Rieger A, Loddenkemper C, Somasundaram R, Berg T, Hamm B, Braun J, Sack I. Viscoelasticity-based staging of hepatic fibrosis with multifrequency MR elastography. Radiology 2010; 257: 80-86 [PMID: 20679447 DOI: 10.1148/radiol.10092489] 124 · DOI
  124. Van Houten EE, Miga MI, Weaver JB, Kennedy FE, Paulsen KD. Three-dimensional subzone-based reconstruction algorithm for MR elastography. Magn Reson Med 2001; 45: 827-837 [PMID: 11323809 DOI: 10.1002/mrm.1111] 125 · DOI
  125. Millonig G, Friedrich S, Adolf S, Fonouni H, Golriz M, Mehrabi A, Stiefel P, Pöschl G, Büchler MW, Seitz HK, Mueller S. Liver stiffness is directly influenced by central venous pressure. J Hepatol 2010; 52: 206-210 [PMID: 20022130 DOI: 10.1016/ j.jhep.2009.11.018] 126 · DOI
  126. Sagir A, Erhardt A, Schmitt M, Häussinger D. Transient elasto- Low et al . Review of magnetic resonance elastography WJR|www.wjgnet.com graphy is unreliable for detection of cirrhosis in patients with acute liver damage. Hepatology 2008; 47: 592-595 [PMID: 18098325 DOI: 10.1002/hep.22056] 127 · DOI
  127. Lebray P, Varnous S, Charlotte F, Varaut A, Poynard T, Ratziu V. Liver stiffness is an unreliable marker of liver fibrosis in patients with cardiac insufficiency. Hepatology 2008; 48: 2089 [PMID: 19003902 DOI: 10.1002/hep.22594] 128 · DOI
  128. Warner L, Yin M, Glaser KJ, Woollard JA, Carrascal CA, Korsmo MJ, Crane JA, Ehman RL, Lerman LO. Noninvasive In vivo assessment of renal tissue elasticity during graded renal ischemia using MR elastography. Invest Radiol 2011; 46: 509-514 [PMID: 21467945 DOI: 10.1097/RLI.0b013e3182183a95] · DOI

Cited by

Sorted by citations

Authors
Low, G.
Affiliations
Alberta Hospital Edmonton (CA)
Corresponding author
Low, Gavin
Venue
World Journal of Radiology
Published
2016
Volume
8
Issue
1
Pages
59
ISSN
1949-8470 (ISSN-L)
OpenAlex
W2411235467
Cited by
123
References
0
Citations by year
2026: 6 · 2025: 12 · 2024: 14 · 2023: 18 · 2022: 6 · 2021: 17 · 2020: 13 · 2019: 10 · 2018: 15 · 2017: 5 · 2016: 5 · 2015: 2
FWCI
1.82
Type
journal article
Language
en
Open access
Licence
cc-by-nc
Locations
  • World Journal of Radiology (publishedVersion, cc-by-nc)
  • PubMed Central (publishedVersion)
  • PubMed (publishedVersion)
Funders
  • National Institutes of Health (EB001981)
  • NIHR Cambridge Biomedical Research Centre
  • National Institute of Biomedical Imaging and Bioengineering (R01EB001981)
Topics
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Ultrasound Imaging and Elastography
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › MRI in cancer diagnosis
Keywords
magnetic resonance elastography, shear wave imaging, phase-contrast MRI, tissue stiffness, soft tissue mechanics, liver fibrosis, noninvasive fibrosis staging, virtual palpation, multiparametric MRI

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review