Validation of a Self-Applied Unattended Monitor for Sleep Disordered Breathing
Open accessAbstract
Study objectives To evaluate the validity of the Apnea Risk Evaluation System (ARES) Unicorder, a self-applied, limited-channel portable monitoring device for the evaluation of sleep disordered breathing (SDB). Design Prospective study with blinded analysis. Setting Sleep disorder center, academic institution. Participants Eighty patients with suspected obstructive sleep apnea hypopnea syndrome (OSAHS) and 22 volunteers. Interventions N/A. Measurements and results Subjects used the ARES Unicorder at home for 2 nights using only written instructions. Within 2 weeks, they returned to the laboratory for full nocturnal polysomnography (NPSG) with simultaneous monitoring with the Unicorder. NPSGs were scored manually to obtain an apnea-hypopnea index based on Medicare guidelines (AHI4%) and a respiratory disturbance index (RDI). ARES studies were autoscored and reviewed to obtain indices based on equivalent definitions i.e., AHI4%(ARES) and apnea hypopnea (events with 1% desaturation) index (AHI1%(ARES)). Indices from the NPSG were compared to the in-lab ARES and in-home ARES indices using mean differences and the intraclass correlations (ICC). For the in-lab comparison, there was high concordance between AHI4%(NPSG) and AHI4%(ARES) (ICC = 0. 96, mean difference = 0.5/hour) and RDI(NPSG ) and AHI1%(ARES) (ICC =0.93, mean difference = 3.2/hour). For NPSG versus In-Home ARES comparison, there was good concordance between AHI4%(NPSG) and AHI4%(ARES) (ICC = 0.8, mean difference = 4.1/…
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In a high-risk population using cutoffs of AHI ≥ 5, AHI ≥ 15 and AHI ≥ 30, accuracy of 2-3 channel HSAT devices ranged from 81% to 93%, 72% to 87%, and 71% to 90%, respectively.
“In a high-risk population using cutoffs of AHI ≥ 5,95–97 AHI ≥ 15,95–99 and AHI ≥ 30,96,97 accuracy ranged from 81% to 93%, 72% to 87%, and 71% to 90%, respectively.”
From Clinical Practice Guideline for Diagnostic Testing for Adult Obstructive Sleep Apnea: An American Academy of Sleep Medicine Clinical Practice Guideline (2017) · cites this paper as [95]
Cited by
Sorted by citations
Kapur, V. K., Auckley, D. H., Chowdhuri, S., Kuhlmann, D. C., Mehra, R., Ramar, K., & Harrod, C. G. (2017) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.6506 · Cited by 3,253 · see citing works
Metrics of sleep apnea severity: beyond the apnea-hypopnea index
Malhotra, A., Ayappa, I., Ayas, N., Collop, N., Kirsch, D., Mcardle, N., Mehra, R., Pack, A. I., Punjabi, N., White, D. P., et al. (2021) · SLEEP · doi:10.1093/sleep/zsab030 · Cited by 489 · see citing works
Self-citationObstructive Sleep Apnea Devices for Out-Of-Center (OOC) Testing: Technology Evaluation
Collop, N. A., Tracy, S. L., Kapur, V., Mehra, R., Kuhlmann, D., Fleishman, S. A., & Ojile, J. M. (2011) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.1328 · Cited by 427 · see citing works
International Consensus Statement on Obstructive Sleep Apnea
Chang, J. L., Goldberg, A. N., Alt, J. A., Mohammed, A., Ashbrook, L., Auckley, D., Ayappa, I., Bakhtiar, H., Barrera, J. E., Bartley, B. L., et al. (2023) · International Forum of Allergy & Rhinology · doi:10.1002/alr.23079 · Cited by 350 · see citing works
Self-citationLucassen, E. A., Zhao, X., Rother, K. I., Mattingly, M. S., Courville, A. B., de Jonge, L., Csako, G., Cizza, G., & for the Sleep Extension Study Group (2013) · PLoS ONE · doi:10.1371/journal.pone.0056519 · Cited by 264 · see citing works
El Shayeb, M., Topfer, L.-A., Stafinski, T., Pawluk, L., & Menon, D. (2014) · Canadian Medical Association Journal · doi:10.1503/cmaj.130952 · Cited by 237 · see citing works
Masa, J. F., Corral, J., Pereira, R., Duran-Cantolla, J., Cabello, M., Hernandez-Blasco, L., Monasterio, C., Alonso, A., Chiner, E., Rubio, M., et al. (2011) · Thorax · doi:10.1136/thx.2010.152272 · Cited by 161 · see citing works
Clinical utility of the Epworth sleepiness scale
Walker, N. A., Sunderram, J., Zhang, P., Lu, S.-E., & Scharf, M. T. (2020) · Sleep and Breathing · doi:10.1007/s11325-020-02015-2 · Cited by 102 · see citing works
Guerrero, A., Embid, C., Isetta, V., Farre, R., Duran-Cantolla, J., Parra, O., Barbé, F., Montserrat, J. M., & Masa, J. F. (2014) · Sleep · doi:10.5665/sleep.3932 · Cited by 62 · see citing works
Chronic Rhinosinusitis Is an Independent Risk Factor for OSA in World Trade Center Responders
Sunderram, J., Weintraub, M., Black, K., Alimokhtari, S., Twumasi, A., Sanders, H., Udasin, I., Harrison, D., Chitkara, N., de la Hoz, R. E., et al. (2019) · Chest · doi:10.1016/j.chest.2018.10.015 · Cited by 38 · see citing works
Self-citationPredictors of sleep-disordered breathing in obese adults who are chronic short sleepers
Knutson, K. L., Zhao, X., Mattingly, M., Galli, G., & Cizza, G. (2012) · Sleep Medicine · doi:10.1016/j.sleep.2011.11.009 · Cited by 13 · see citing works
- Authors
- Ayappa, I., Norman, R. G., Seelall, V., & Rapoport, D. M.
- Affiliations
- New York University (US)
- Venue
- Journal of Clinical Sleep Medicine
- Published
- 2008-02-15
- Volume
- 4
- Issue
- 1
- Pages
- 26-37
- ISSN
- 1550-9389 (ISSN-L)
- 1550-9397
- PMC
- PMC2276822
- OpenAlex
- W2139611461
- Cited by
- 125
- References
- 0
- Citations by year
- 2026: 1 · 2025: 3 · 2024: 1 · 2023: 3 · 2022: 6 · 2021: 6 · 2020: 11 · 2019: 4 · 2018: 5 · 2017: 8 · 2016: 10 · 2015: 8 · 2014: 8 · 2013: 12 · 2012: 8
- FWCI
- 6.18
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Journal of Clinical Sleep Medicine (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Center for Research Resources (M01RR000096)
- Topics
- Health Sciences › Medicine › Physiology › Obstructive Sleep Apnea Research
- Social Sciences › Psychology › Experimental and Cognitive Psychology › Sleep and related disorders
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- sleep-disordered breathing, obstructive sleep apnea-hypopnea syndrome, home sleep apnea testing, device validation, apnea-hypopnea index, respiratory disturbance index, diagnostic accuracy, polysomnography comparison, sensitivity and specificity, likelihood ratios, sleep apnea screening
- MeSH terms
- Adult, Algorithms, Diagnosis, Computer-Assisted, Female, Home Care Services, Humans, Male, Middle Aged, Prospective Studies, Self Care, Sensitivity and Specificity, Signal Processing, Computer-Assisted, Software, Technology Assessment, Biomedical, Polysomnography, Sleep Apnea, Obstructive
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review