Continuous Glucose Monitoring for Diabetes: Potential Pitfalls for the General Physician

Open access
Herrod, Samuel S, Liversedge, Grace, Vaidya, Bijay, Walker, Neil·2022·Clinical Medicine

Abstract

A 31-year-old man presented systemically unwell with diabetic ketoacidosis (DKA). He was using an intermittently scanned continuous glucose monitoring (CGM) device that had been recording low or normal glucose readings for the 48 hours prior to admission. The sensor site had become infected, and we believe this soft tissue infection caused his CGM device to record falsely low glucose readings leading the patient to erroneously lower his insulin doses and take extra carbohydrates, precipitating DKA. CGM devices measure glucose levels in the interstitial fluid. When interstitial glucose readings do not match symptoms or expectations, a capillary blood glucose reading should be taken to correlate and impact treatment decisions. There will be an increase in patients presenting to hospital with CGM devices as the National Institute for Health and Care Excellence guidelines have recently been updated. We use this interesting clinical case to provide context for key learning points about CGM devices for the general physician.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.7861/cm-2022-0199.r1

Cites

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • CGMs measure interstitial fluid glucose that has diffused from capillaries into subcutaneous tissue and may indicate a falsely low blood glucose due to device compression, device calibration errors, and skin/soft tissue infections.

    “These devices measure interstitial fluid glucose that has diffused from the capillaries and into the subcutaneous tissue and may indicate a falsely low BG due to device compression, device calibration errors, and skin/soft tissue infections [4,6,7].”

    From Non-Diabetic Hypoglycemia: Evaluation and Management in Adults (2025) · cites this paper as [7]

    The section “What are continuous glucose monitoring devices?” says CGMs measure interstitial-fluid glucose and reports a case in which a soft-tissue infection was believed to cause falsely low readings (page unavailable). Our copy does not state that the glucose diffused from capillaries into subcutaneous tissue or identify device compression or calibration errors as causes of falsely low readings. (LLM, conf=0.87)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

No references could be resolved yet.

Research works printed but not linked: 3

  1. Lind M, Hirsch PWI, Heise T et al. Continuous glucose monitoring vs conventional therapy for glycemic control in adults with type 1 diabetes treated with multiple daily insulin injections the GOLD randomized clinical trial. JAMA 2017;317:379-87.
  2. Tokarski F, Hobson C, Lartigue M, Lemaignen A, Lioger B. Streptococcus pyogenes necrotizing soft tissue infection following glucose monitoring with FreeStyle Libre device. QJM 2021;114: 58-9.
  3. McClatchey PM, McClain ES, Williams IM et al. Fibrotic encapsula- tion is the dominant source of continuous glucose monitor delays. Diabetes 2019;68:1892-901.

Cited by

Sorted by citations

Authors
Herrod, S. S., Liversedge, G., Vaidya, B., & Walker, N.
Venue
Clinical Medicine
Publisher
Elsevier BV
Published
2022-09
Volume
22
Issue
5
Pages
482-484
ISSN
1470-2118 (ISSN-L)
OpenAlex
W4297914158
Cited by
5
References
0
Citations by year
2026: 1 · 2025: 2 · 2024: 2
FWCI
0.74
Type
journal article
Language
en
Open access
Locations
  • Clinical Medicine (publishedVersion, cc-by-nc-nd)
  • PubMed Central (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
Keywords
continuous glucose monitoring, diabetic ketoacidosis, soft tissue infection, flash glucose monitoring, interstitial glucose, capillary blood glucose, insulin dose adjustment, diabetes management
MeSH terms
Adult, Blood Glucose, Diabetes Mellitus, Type 1, Glucose, Glycated Hemoglobin, Humans, Male, Physicians, Blood Glucose Self-Monitoring, Diabetic Ketoacidosis

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review