Are Aneroid Sphygmomanometers Accurate in Hospital and Clinic Settings?

Canzanello, Vincent J., Jensen, Patricia L., Schwartz, Gary L. · 2001 · Archives of Internal Medicine
Cited by 133

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

BACKGROUND: The aneroid sphygmomanometer is commonly used for the indirect measurement of blood pressure despite significant concerns about its accuracy. Although the mercury sphygmomanometer is highly accurate, there are concerns about the environmental toxicity of mercury. In response to various external pressures to become essentially mercury free, the Mayo Clinic, Rochester, Minn, has replaced many mercury sphygmomanometers with aneroid devices. Since 1993, a maintenance protocol has been in place to ensure proper function and accuracy of these devices. METHODS: We assessed the accuracy of 283 aneroid devices using as the reference standard a digital pressure and vacuum meter that was calibrated using a mercury sphygmomanometer. RESULTS: The mean +/- SD values from the aneroid device in millimeters of mercury at each reference point (at 20-mm Hg intervals from 60 to 240 mm Hg defined by the reference device) were 59.9 +/- 1.9 at 60; 79.9 +/- 1.9 at 80; 100.0 +/- 1.8 at 100; 120.3 +/- 1.8 at 120; 140.7 +/- 1.4 at 140; 160.7 +/- 1.7 at 160; 180.9 +/- 1.3 at 180; 200.7 +/- 5.0 at 200; 221.0 +/- 1.3 at 220; and 240.8 +/- 1.6 at 240 (r = 0.99; P<.001). The values from the aneroid device underestimated those of the reference device by a mean of 0.5 mm Hg (95% confidence interval, 0.3-0.7). Virtually 100% of the values from the aneroid device were within the 4-mm Hg range recommended by the Association for the Advancement of Medical Instrumentation. CONCLUSION: Aneroid sphygmomanometers provide accurate pressure measurements when a proper maintenance protocol is followed.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1001/archinte.161.5.729

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 133

Metadata

Authors
Canzanello, V. J., Jensen, P. L., & Schwartz, G. L.
Venue
Archives of Internal Medicine
Published
2001-03-12
Volume
161
Issue
5
Pages
729
ISSN
0003-9926 (ISSN-L)
OpenAlex
W2044656600
Cited by
133
References
1
FWCI
6.80
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Are Aneroid Sphygmomanometers Accurate in Hospital and Clinic Settings? https://app.cochat.ai/papers/10.1001/archinte.161.5.729
The paper
Canzanello, V. J., Jensen, P. L., & Schwartz, G. L. (2001). Are Aneroid Sphygmomanometers Accurate in Hospital and Clinic Settings? Archives of Internal Medicine. https://doi.org/10.1001/archinte.161.5.729

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.