Effect of Mechanically Expanded vs Self-Expanding Transcatheter Aortic Valve Replacement on Mortality and Major Adverse Clinical Events in High-Risk Patients With Aortic Stenosis: The REPRISE III Randomized Clinical Trial

Open access
Feldman, Ted E., Reardon, Michael J., Rajagopal, Vivek, Makkar, Raj R., Bajwa, Tanvir K., Kleiman, Neal S., Linke, Axel, Kereiakes, Dean J., Waksman, Ron, Thourani, Vinod H., Stoler, Robert C., Mishkel, Gregory J., Rizik, David G., Iyer, Vijay S., Gleason, Thomas G., Tchétché, Didier, Rovin, Joshua D., Buchbinder, Maurice, Meredith, Ian T., Götberg, Matthias, Bjursten, Henrik, Meduri, Christopher, Salinger, Michael H., Allocco, Dominic J., Dawkins, Keith D. · 2018 · JAMA
Cited by 168

The DOI matches its Crossref record; 45 of the 45 references it cites were found in a registry or an index.

Abstract

Importance: Transcatheter aortic valve replacement (TAVR) is established for selected patients with severe aortic stenosis. However, limitations such as suboptimal deployment, conduction disturbances, and paravalvular leak occur. Objective: To evaluate if a mechanically expanded valve (MEV) is noninferior to an approved self-expanding valve (SEV) in high-risk patients with aortic stenosis undergoing TAVR. Design, Setting, and Participants: The REPRISE III trial was conducted in 912 patients with high or extreme risk and severe, symptomatic aortic stenosis at 55 centers in North America, Europe, and Australia between September 22, 2014, and December 24, 2015, with final follow-up on March 8, 2017. Interventions: Participants were randomized in a 2:1 ratio to receive either an MEV (n = 607) or an SEV (n = 305). Main Outcomes and Measures: The primary safety end point was the 30-day composite of all-cause mortality, stroke, life-threatening or major bleeding, stage 2/3 acute kidney injury, and major vascular complications tested for noninferiority (margin, 10.5%). The primary effectiveness end point was the 1-year composite of all-cause mortality, disabling stroke, and moderate or greater paravalvular leak tested for noninferiority (margin, 9.5%). If noninferiority criteria were met, the secondary end point of 1-year moderate or greater paravalvular leak was tested for superiority in the full analysis data set. Results: Among 912 randomized patients (mean age, 82.8 [SD, 7.3] years; 463 [51%] women; predicted risk of mortality, 6.8%), 874 (96%) were evaluable at 1 year. The primary safety composite end point at 30 days occurred in 20.3% of MEV patients and 17.2% of SEV patients (difference, 3.1%; Farrington-Manning 97.5% CI, -∞ to 8.3%; P = .003 for noninferiority). At 1 year, the primary effectiveness composite end point occurred in 15.4% with the MEV and 25.5% with the SEV (difference, -10.1%; Farrington-Manning 97.5% CI, -∞ to -4.4%; P<.001 for noninferiority). The 1-year rates of moderate or severe paravalvular leak were 0.9% for the MEV and 6.8% for the SEV (difference, -6.1%; 95% CI, -9.6% to -2.6%; P < .001). The superiority analysis for primary effectiveness was statistically significant (difference, -10.2%; 95% CI, -16.3% to -4.0%; P < .001). The MEV had higher rates of new pacemaker implants (35.5% vs 19.6%; P < .001) and valve thrombosis (1.5% vs 0%) but lower rates of repeat procedures (0.2% vs 2.0%), valve-in-valve deployments (0% vs 3.7%), and valve malpositioning (0% vs 2.7%). Conclusions and Relevance: Among high-risk patients with aortic stenosis, use of the MEV compared with the SEV did not result in inferior outcomes for the primary safety end point or the primary effectiveness end point. These findings suggest that the MEV may be a useful addition for TAVR in high-risk patients. Trial Registration: ClinicalTrials.gov Identifier: NCT02202434.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1001/jama.2017.19132

Cites

45 of 45 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 45 references

Cited by

See all 168

Metadata

Authors
Feldman, T. E., Reardon, M. J., Rajagopal, V., Makkar, R. R., Bajwa, T. K., Kleiman, N. S., Linke, A., Kereiakes, D. J., Waksman, R., Thourani, V. H., Stoler, R. C., Mishkel, G. J., Rizik, D. G., Iyer, V. S., Gleason, T. G., Tchétché, D., Rovin, J. D., Buchbinder, M., Meredith, I. T., Götberg, M., Bjursten, H., Meduri, C., Salinger, M. H., Allocco, D. J., & Dawkins, K. D.
Venue
JAMA
Published
2018-01-02
Volume
319
Issue
1
Pages
27
ISSN
  • 0002-9955 (ISSN-L)
  • 0098-7484
OpenAlex
W2781876810
Cited by
168
References
45
FWCI
18.75
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Effect of Mechanically Expanded vs Self-Expanding Transcatheter Aortic Valve Replacement on Mortality and Major Adverse Clinical Events in High-Risk Patients With Aortic Stenosis: The REPRISE III Randomized Clinical Trial. https://app.cochat.ai/papers/10.1001/jama.2017.19132
The paper
Feldman, T. E., Reardon, M. J., Rajagopal, V., Makkar, R. R., Bajwa, T. K., Kleiman, N. S., Linke, A., Kereiakes, D. J., Waksman, R., Thourani, V. H., Stoler, R. C., Mishkel, G. J., Rizik, D. G., Iyer, V. S., Gleason, T. G., Tchétché, D., Rovin, J. D., Buchbinder, M., Meredith, I. T., . . . Dawkins, K. D. (2018). Effect of Mechanically Expanded vs Self-Expanding Transcatheter Aortic Valve Replacement on Mortality and Major Adverse Clinical Events in High-Risk Patients With Aortic Stenosis: The REPRISE III Randomized Clinical Trial. JAMA. https://doi.org/10.1001/jama.2017.19132

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.