Surveillance of Medical Device–Related Hazards and Adverse Events in Hospitalized Patients
Abstract
CONTEXT: Although adverse drug events have been extensively evaluated by computer-based surveillance, medical device errors have no comparable surveillance techniques. OBJECTIVES: To determine whether computer-based surveillance can reliably identify medical device-related hazards (no known harm to patient) and adverse medical device events (AMDEs; patient experienced harm) and to compare alternative methods of detection of device-related problems. DESIGN, SETTING, AND PARTICIPANTS: This descriptive study was conducted from January through September 2000 at a 520-bed tertiary teaching institution in the United States with experience in using computer tools to detect and prevent adverse drug events. All 20 441 regular and short-stay patients (excluding obstetric and newborn patients) were included. MAIN OUTCOME MEASURES: Medical device events as detected by computer-based flags, telemetry problem checklists, International Classification of Diseases, Ninth Revision (ICD-9) discharge code (which could include AMDEs present at admission), clinical engineering work logs, and patient survey results were compared with each other and with routine voluntary incident reports to determine frequencies, proportions, positive predictive values, and incidence rates by each technique. RESULTS: Of the 7059 flags triggered, 552 (7.8%) indicate a device-related hazard or AMDE. The estimated 9-month incidence rates (number per 1000 admissions [95% confidence intervals]) for AMDEs were 1.6 (0.9-2.5) for incident reports, 27.7 (24.9-30.7) for computer flags, and 64.6 (60.4-69.1) for ICD-9 discharge codes. Few of these events were detected by more than 1 surveillance method, giving an overall incidence of AMDE detected by at least 1 of these methods of 83.7 per 1000 (95% confidence interval, 78.8-88.6) admissions. The positive predictive value of computer flags for detecting device-related hazards and AMDEs ranged from 0% to 38%. CONCLUSIONS: More intensive surveillance methods yielded higher rates of medical device problems than found with traditional voluntary reporting, with little overlap between methods. Several detection methods had low efficiency in detecting AMDEs. The high rate of AMDEs suggests that AMDEs are an important patient safety issue, but additional research is necessary to identify optimal AMDE detection strategies.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Incident reports provided by healthcare professionals have been shown to be useful in examining the risks and harm caused by healthcare.
“Incident reports provided by healthcare professionals have been shown to be useful in examining the risks and harm caused by healthcare (eg, falls, medication errors, therapeutic devices, and equipment).18 19”
From An analysis of computer-related patient safety incidents to inform the development of a classification (2010) · cites this paper as [19]
Cites
CDC definitions for nosocomial infections, 1988
Garner, J. S., Jarvis, W. R., Emori, T. G., Horan, T. C., & Hughes, J. M. (1988) · American Journal of Infection Control · doi:10.1016/0196-6553(88)90053-3 · Cited by 5,665 · see citing works
CDC definitions for nosocomial infections
Yangco, B. G. (1989) · American Journal of Infection Control · doi:10.1016/s0196-6553(89)80013-6 · Cited by 3,779 · see citing works
The Nature of Adverse Events in Hospitalized Patients
Leape, L. L., Brennan, T. A., Laird, N. M., Lawthers, A. G., Localio, A. R., Barnes, B. A., Hebert, L. E., Newhouse, J. P., Weiler, P. C., & Hiatt, H. H. (1991) · New England Journal of Medicine · doi:10.1056/nejm199102073240605 · Cited by 3,766 · see citing works
Incidence of Adverse Drug Events and Potential Adverse Drug Events: Implications for Prevention
Bates, D. W. (1995) · JAMA · doi:10.1001/jama.1995.03530010043033 · Cited by 2,127 · see citing works
Incidence and Types of Adverse Events and Negligent Care in Utah and Colorado
Thomas, E. J., Studdert, D. M., Burstin, H. R., Orav, E. J., Zeena, T., WILLIAMS, E. J., Howard, K., Weiler, P. C., & Brennan, T. A. (2000) · Medical Care · doi:10.1097/00005650-200003000-00003 · Cited by 1,673 · see citing works
Computerized Surveillance of Adverse Drug Events in Hospital Patients
Classen, D. C. (1991) · JAMA · doi:10.1001/jama.1991.03470200059035 · Cited by 643 · see citing works
The Incident Reporting System Does Not Detect Adverse Drug Events: A Problem for Quality Improvement
Cullen, D. J., Bates, D. W., Small, S., Cooper, J. B., Nemeskal, A. R., & Leape, L. L. (1995) · The Joint Commission Journal on Quality Improvement · doi:10.1016/s1070-3241(16)30180-8 · Cited by 601 · see citing works
Jha, A. K., Kuperman, G. J., Teich, J. M., Leape, L. L., Shea, B. S., Rittenberg, E., Burdick, E., Seger, D. L., Vliet, M. V., & Bates, D. W. (1998) · Journal of the American Medical Informatics Association · doi:10.1136/jamia.1998.0050305 · Cited by 514 · see citing works
Drug complications in outpatients
Gandhi, T. K., Burstin, H. R., Cook, E. F., Puopolo, A. L., Haas, J. S., Brennan, T. A., & Bates, D. W. (2000) · Journal of General Internal Medicine · doi:10.1046/j.1525-1497.2000.04199.x · Cited by 430 · see citing works
Poor prognosis for existing monitors in the intensive care unit
Tsien, C. L., & Fackler, J. C. (1997) · Critical Care Medicine · doi:10.1097/00003246-199704000-00010 · Cited by 309 · see citing works
Adverse Events With Transvenous Implantable Cardioverter-Defibrillators
Rosenqvist, M., Beyer, T., Block, M. S., den Dulk, K., Minten, J., & Lindemans, F. W. (1998) · Circulation · doi:10.1161/01.cir.98.7.663 · Cited by 231 · see citing works
Human factors engineering and patient safety
Gosbee, J. W. (2002) · BMJ Quality & Safety · doi:10.1136/qhc.11.4.352 · Cited by 196 · see citing works
Alarms in the intensive care unit: how can the number of false alarms be reduced?
Chambrin, M.-C. (2001) · Critical Care · doi:10.1186/cc1021 · Cited by 176 · see citing works
Identification of Adverse Events Occurring during Hospitalization
Brennan, T. A., Localio, A. R., Leape, L. L., Laird, N. M., Peterson, L., Hiatt, H. H., & Barnes, B. A. (1990) · Annals of Internal Medicine · doi:10.7326/0003-4819-112-3-221 · Cited by 168 · see citing works
Lin, L., Vicente, K. J., & Doyle, D. J. (2001) · Journal of Biomedical Informatics · doi:10.1006/jbin.2001.1028 · Cited by 142 · see citing works
Development of a computerized adverse drug event monitor.
Evans, R. S., Pestotnik, S. L., Classen, D. C., Bass, S. B., Menlove, R. L., Gardner, R. M., & Burke, J. P. (1991) · PubMed Central · Cited by 81 · see citing works
Evaluation of Screening Criteria for Adverse Events in Medical Patients
Bates, D. W., Neil, A. C. O., Petersen, L. A., Lee, T. H.-Y., & Brennan, T. A. (1995) · Medical Care · doi:10.1097/00005650-199505000-00002 · Cited by 80 · see citing works
Burns from Warming Devices in Anesthesia
Cheney, F. W., Posner, K. L., Caplan, R. A., et al. (1994) · Anesthesiology · doi:10.1097/00000542-199404000-00012 · Cited by 77 · see citing works
Learning from adverse incidents involving medical devices
Amoore, J., & Ingram, P. (2002) · BMJ · doi:10.1136/bmj.325.7358.272 · Cited by 75 · see citing works
Prevention of adverse drug events through computerized surveillance.
Evans, R. S., Pestotnik, S. L., Classen, D. C., Bass, S. B., & Burke, J. P. (1992) · PubMed Central · Cited by 74 · see citing works
Adverse Incident Reporting in Intensive Care
Hart, G. K., Baldwin, I. T., Gutteridge, G. A., & Ford, J. (1994) · Anaesthesia and Intensive Care · doi:10.1177/0310057x9402200509 · Cited by 60 · see citing works
Airway Injury During Anesthesia: A Closed Claims Analysis
Domino, K. B., Posner, K. L., Caplan, R. A., et al. (2000) · Survey of Anesthesiology · doi:10.1097/00132586-200010000-00062 · Cited by 57 · see citing works
Human factors engineering design demonstrations can enlighten your RCA team
Gosbee, J. W. (2003) · BMJ Quality & Safety · doi:10.1136/qhc.12.2.119 · Cited by 48 · see citing works
Human error and patient-controlled analgesia pumps.
Brown, S. L., Bogner, M. S., Parmentier, C. M., et al. (1998) · PubMed · Cited by 34 · see citing works
Infection surveillance in home care: Device-related incidence rates
Rosenheimer, L., Embry, F. C., Sanford, J., et al. (1998) · American Journal of Infection Control · doi:10.1016/s0196-6553(98)80018-7 · Cited by 31 · see citing works
Perspectives on Development of IEEE 1073: The Medical Information Bus (MIB) Standard
Kennelly, R. J., & Gardner, R. M. (1997) · International journal of clinical monitoring and computing · doi:10.1007/bf03356588 · Cited by 26 · see citing works
Improving Patient Care by Reporting Problems with Medical Devices
White, G. G., Weick-Brady, M. D., & Gross, T. P. (1998) · Journal of Toxicology Clinical Toxicology · doi:10.3109/15563659809028066 · Cited by 26 · see citing works
Infusion pump adverse events: experience from medical device reports.
Brown, S. L., Morrison, A. E., Parmentier, C. M., et al. (1997) · PubMed · Cited by 19 · see citing works
Dental device–associated problems
FULLER, J., & PARMENTIER, C. (2001) · The Journal of the American Dental Association · doi:10.14219/jada.archive.2001.0087 · Cited by 17 · see citing works
Improving acute care through use of medical device data
Kennelly, R. J. (1998) · International Journal of Medical Informatics · doi:10.1016/s1386-5056(97)00120-2 · Cited by 16 · see citing works
Health Effects Associated With Medical Glove Use
Dillard, S. F., Hefflin, B. J., Kaczmarek, R. G., et al. (2002) · AORN Journal · doi:10.1016/s0001-2092(06)61098-3 · Cited by 13 · see citing works
An association of silicone-gel breast implant rupture and fibromyalgia
Brown, S. L., Duggirala, H. J., & Pennello, G. A. (2002) · Current Rheumatology Reports · doi:10.1007/s11926-002-0037-6 · Cited by 13 · see citing works
Austin, A. (2000) · British Journal of Plastic Surgery · doi:10.1054/bjps.2000.3353 · Cited by 12 · see citing works
Ward, R. A. (2002) · Artificial Organs · doi:10.1046/j.1525-1594.2002.00913.x · Cited by 8 · see citing works
The human factors implications of peritoneal dialysis: Cycler overfill incident reports
Reid, M. H., & Sawyer, D. (1999) · International Journal of Trauma Nursing · doi:10.1016/s1075-4210(99)90034-2 · Cited by 3 · see citing works
Medical device adverse events and the temporary invasive cardiac pacemaker
Dwyer, D. (2001) · International Journal of Trauma Nursing · doi:10.1067/mtn.2001.115213 · Cited by 3 · see citing works
Medical device surveillance world-wide.
Badea, C. T., Kolitsi, Z., Pallikarakis, N. E., et al. (1996) · PubMed · Cited by 1 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Magrabi, F., Ong, M.-S., Runciman, W., & Coiera, E. (2010) · Journal of the American Medical Informatics Association · doi:10.1136/jamia.2009.002444 · Cited by 162 · see citing works
Technology, governance and patient safety: Systems issues in technology and patient safety
Balka, E., Doyle-Waters, M., Lecznarowicz, D., & FitzGerald, J. M. (2007) · International Journal of Medical Informatics · doi:10.1016/j.ijmedinf.2006.05.046 · Cited by 53 · see citing works
- Authors
- Samore, M. H.
- Affiliations
- University of Utah (US)
- Corresponding author
- Samore, Matthew H.
- ORCID
- Samore, Matthew H.
- Venue
- JAMA
- Publisher
- American Medical Association (AMA)
- Published
- 2004-01-21
- Volume
- 291
- Issue
- 3
- Pages
- 325
- ISSN
- 0002-9955 (ISSN-L)
- 0098-7484
- OpenAlex
- W2125794271
- Cited by
- 142
- References
- 37
- Citations by year
- 2026: 2 · 2025: 5 · 2024: 9 · 2023: 2 · 2022: 2 · 2021: 8 · 2020: 5 · 2019: 2 · 2018: 4 · 2017: 6 · 2016: 1 · 2015: 5 · 2014: 10 · 2013: 6 · 2012: 7
- FWCI
- 10.74
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- JAMA (publishedVersion)
- PubMed (publishedVersion)
- Funders
- U.S. Public Health Service (223-99-6101)
- Topics
- Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
- Health Sciences › Health Professions › Emergency Medical Services › Patient Safety and Medication Errors
- Life Sciences › Pharmacology, Toxicology and Pharmaceutics › Toxicology › Pharmacovigilance and Adverse Drug Reactions
- Keywords
- medical device errors, hospital patient safety, incident reporting, positive predictive value, hospitalized patients
- MeSH terms
- Adult, Aged, Computing Methodologies, Equipment and Supplies, Equipment Failure, Female, Hospital Records, Hospitalization, Hospitals, Teaching, Humans, Male, Middle Aged, Pilot Projects, Product Surveillance, Postmarketing, United States, Medical Records Systems, Computerized, Medical Errors, International Classification of Diseases
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review