The DOI matches its Crossref record.
Stent fracture associated with drug‐eluting stents: Clinical characteristics and implications
Abstract
Objective: To evaluate the clinical characteristics and implications of stent fracture in drug‐eluting stents. Background: Approximately 2.5 million drug‐eluting stents are implanted every year worldwide. In 10 randomized controlled trials involving 2,602 patients, no incidence of stent fracture was recognized or reported. Methods: From April 2003 to December 2005, 2,728 patients underwent drug‐eluting stenting. The angiograms of all 530 patients who underwent repeat angiography were analyzed to identify the presence of stent fracture. We then documented the incidence of adverse events associated with drug‐eluting stent fracture and systematically analyzed the clinical, procedural, and structural factors, which might predispose to stent fracture. Results: Stent fracture was identified in 10 patients. None of these fractures were detectable at the time of stent placement. The median time interval from stent implantation to detection of fracture at repeat angiography was 226 days (range, 7–620 days). Adverse clinical outcomes associated with stent fracture occurred in 7 patients (6 patients had binary restenosis and 1 patient had stent thrombosis), all necessitating repeat intervention. Analysis of potential predisposing clinical, procedural, and structural factors revealed that 4 patients had excessive tortuosity in the proximal segment, and overlapping stents were used in 5 cases. All fractures occurred in sirolimus‐eluting stents. Conclusions: Stent fracture may represent a new potential mechanism of restenosis and stent thrombosis in drug‐eluting stents. Predisposing clinical and procedural factors may be vessel tortuosity and use of overlapping stents. The most important predisposing factor, however, may be stent structure, since all fractures occurred in sirolimus‐eluting stents. © 2006 Wiley‐Liss, Inc.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
In-Stent Restenosis in the Drug-Eluting Stent Era
Dangas, G. D., Claessen, B. E., Caixeta, A., Sanidas, E. A., Mintz, G. S., & Mehran, R. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2010.07.028 · Cited by 803
Garg, S., & Serruys, P. W. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2010.06.007 · Cited by 501
Wang, P.-J., Nezami, F. R., Gorji, M. B., Berti, F., Petrini, L., Wierzbicki, T., Migliavacca, F., & Edelman, E. R. (2018) · Acta Biomaterialia · doi:10.1016/j.actbio.2018.10.020 · Cited by 31
Metadata
- Authors
- Lee, M. S., Jurewitz, D., Aragon, J., Forrester, J., Makkar, R. R., & Kar, S.
- Venue
- Catheterization and Cardiovascular Interventions
- Publisher
- Wiley
- Published
- 2006-12-28
- Volume
- 69
- Issue
- 3
- Pages
- 387-394
- ISSN
- 1522-1946 (ISSN-L)
- 1522-726X
- OpenAlex
- W1980940890
- Cited by
- 176
- References
- 0
- FWCI
- 15.58
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.