Effect of working environment and procedural strategies on mechanical performance of bioresorbable vascular scaffolds
Open accessAbstract
Polymeric bioresorbable scaffolds (BRS), at their early stages of invention, were considered as a promising revolution in interventional cardiology. However, they failed dramatically compared to metal stents showing substantially higher incidence of device failure and clinical events, especially thrombosis. One problem is that use of paradigms inherited from metal stents ignores dependency of polymer material properties on working environment and manufacturing/deployment steps. Unlike metals, polymeric material characterization experiments cannot be considered identical under dry and submerged conditions at varying rates of operation. We demonstrated different material behaviors associated with variable testing environment and parameters. We, then, have employed extracted material models, which are verified by computational methods, to assess the performance of a full-scale BRS in different working condition and under varying procedural strategies. Our results confirm the accepted notion that slower rate of crimping and inflation can potentially reduce stress concentrations and thus reduce localized damages. However, we reveal that using a universal set of material properties derived from a benchtop experiment conducted regardless of working environment and procedural variability may lead to a significant error in estimation of stress-induced damages and overestimation of benefits procedural updates might offer. We conclude that, for polymeric devices, microstructural damages and localized loss of structural integrity should complement former macroscopic performance-assessment measures (fracture and recoil). Though, to precisely capture localized stress concentration and microstructural damages, context-related testing environment and clinically-relevant procedural scenarios should be devised in preliminary experiments of polymeric resorbable devices to enhance their efficacy and avoid unpredicted clinical events. STATEMENT OF SIGNIFICANCE: Bioresorbable scaffolds (BRS) with the hope to become the next cardiovascular interventional revolution failed in comparison to metal stents. When BRS were characterized using methods for metal stents, designers were misled to seek problem sources at erroneous timeframe and use inefficient indicators, and thus no signal of concern emerged. We demonstrated fundamental flaws associated with applying a universal set of material properties to study device performances in different phases of manufacturing/implantation, and these may be responsible for failure in predicting performance in first-generation BRS. We introduced new criterion for the assessment of structural integrity and device efficacy in next-generation BRS, and indeed all devices using polymeric materials which evolve with the environment they reside in.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Safety and Efficacy of Sirolimus- and Paclitaxel-Eluting Coronary Stents
Stone, G. W., Moses, J. W., Ellis, S. G., Schöfer, J., Dawkins, K. D., Morice, M., Colombo, A., Schampaert, É., Grube, E., Kirtane, A. J., et al. (2007) · New England Journal of Medicine · doi:10.1056/nejmoa067193 · Cited by 1,631 · see citing works
Outcomes associated with drug-eluting and bare-metal stents: a collaborative network meta-analysis
Stettler, C., Wandel, S., Allemann, S., Kastrati, A., Morice, M., Schömig, A., Pfisterer, M. E., Stone, G. W., Leon, M. B., de Lezo, J. S., et al. (2007) · The Lancet · doi:10.1016/s0140-6736(07)61444-5 · Cited by 1,429 · see citing works
Kolandaivelu, K., Swaminathan, R. V., Gibson, W. J., Kolachalama, V. B., Nguyen-Ehrenreich, K.-L., Giddings, V. L., Coleman, L. A., Wong, G. K., & Edelman, E. R. (2011) · Circulation · doi:10.1161/circulationaha.110.003210 · Cited by 801 · see citing works
Self-citationA Pooled Analysis of Data Comparing Sirolimus-Eluting Stents with Bare-Metal Stents
Spaulding, C. M., Daemen, J., Boersma, E. H., Cutlip, D. E., & Serruys, P. W. J. C. (2007) · New England Journal of Medicine · doi:10.1056/nejmoa066633 · Cited by 705 · see citing works
Serruys, P. W. J. C., Chevalier, B., Sotomi, Y., Cequier, A. R., Carrié, D., Piek, J. J., Boven, A. J. V., Dominici, M., Dudek, D., McClean, D., et al. (2016) · The Lancet · doi:10.1016/s0140-6736(16)32050-5 · Cited by 540 · see citing works
Stent fracture associated with drug‐eluting stents: Clinical characteristics and implications
Lee, M. S., Jurewitz, D. L., Aragon, J. R., et al. (2006) · Catheterization and Cardiovascular Interventions · doi:10.1002/ccd.20942 · Cited by 176 · see citing works
Lipinski, M. J., Escarcega, R. O., Baker, N. C., et al. (2016) · JACC: Cardiovascular Interventions · doi:10.1016/j.jcin.2015.09.024 · Cited by 168 · see citing works
Evaluation of poly(L-lactic acid) as a material for intravascular polymeric stents
Agrawal, C. M., Haas, K. F., Leopold, D. A., & Clark, H. G. (1992) · Biomaterials · doi:10.1016/0142-9612(92)90068-y · Cited by 166 · see citing works
A model for simultaneous crystallisation and biodegradation of biodegradable polymers
Han, X., & Pan, J. (2009) · Biomaterials · doi:10.1016/j.biomaterials.2008.10.001 · Cited by 143 · see citing works
Ortega‐Paz, L., Capodanno, D., Gori, T., et al. (2017) · EuroIntervention · doi:10.4244/eij-d-16-00974 · Cited by 136 · see citing works
Sotomi, Y., Suwannasom, P., Serruys, P. W. J. C., et al. (2017) · EuroIntervention · doi:10.4244/eij-d-16-00471 · Cited by 125 · see citing works
Wang, Q., Fang, G., Zhao, Y., et al. (2016) · Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · doi:10.1016/j.jmbbm.2016.08.033 · Cited by 116 · see citing works
Lee, S., Park, J., Shin, D., et al. (2007) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2007.03.073 · Cited by 114 · see citing works
Computational analysis of the radial mechanical performance of PLLA coronary artery stents
Pauck, R. G., & Reddy, B. D. (2014) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2014.09.014 · Cited by 108 · see citing works
Optimizing the design of a bioabsorbable metal stent using computer simulation methods
Grogan, J. A., Leen, S. B., & McHugh, P. E. (2013) · Biomaterials · doi:10.1016/j.biomaterials.2013.07.010 · Cited by 92 · see citing works
Strain-induced accelerated asymmetric spatial degradation of polymeric vascular scaffolds
Wang, P.-J., Ferralis, N., Conway, C., et al. (2018) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1716420115 · Cited by 77 · see citing works
Self-citationMicro-tension and micro-shear experiments to characterize stress-state dependent ductile fracture
Gorji, M. B., & Mohr, D. (2017) · Acta Materialia · doi:10.1016/j.actamat.2017.03.034 · Cited by 70 · see citing works
Self-citationTanaka, A., Latib, A., Kawamoto, H., et al. (2017) · EuroIntervention · doi:10.4244/eij-d-16-00247 · Cited by 65 · see citing works
Jiménez, J. M., Prasad, V., Yu, M. D., Kampmeyer, C. P., Kaakour, A.-H., Wang, P.-J., Maloney, S. F., Wright, N., Johnston, I., Jiang, Y., et al. (2014) · Journal of The Royal Society Interface · doi:10.1098/rsif.2013.1079 · Cited by 55 · see citing works
Self-citationComputational Bench Testing to Evaluate the Short-Term Mechanical Performance of a Polymeric Stent
Bobel, A. C., Petisco-Ferrero, S., Sarasua, J.-R., Wang, W., & McHugh, P. E. (2015) · Cardiovascular Engineering and Technology · doi:10.1007/s13239-015-0235-9 · Cited by 52 · see citing works
Bobel, A. C., Lohfeld, S., Shirazi, R. N., & McHugh, P. E. (2016) · Polymer Testing · doi:10.1016/j.polymertesting.2016.07.011 · Cited by 43 · see citing works
Heterogeneous random medium plasticity and fracture model of additively-manufactured Ti-6Al-4V
Gorji, M. B., Tancogne‐Dejean, T., & Mohr, D. (2018) · Acta Materialia · doi:10.1016/j.actamat.2018.02.025 · Cited by 40 · see citing works
Self-citationNaseem, R., Zhao, L., Liu, Y., & Silberschmidt, V. V. (2017) · Mechanics of Advanced Materials and Modern Processes · doi:10.1186/s40759-017-0028-y · Cited by 36 · see citing works
Magnetically enhanced cell delivery for accelerating recovery of the endothelium in injured arteries
Adamo, R. F., Fishbein, I., Zhang, K., Wen, J., Levy, R. J., Alferiev, I. S., & Chorny, M. (2015) · Journal of Controlled Release · doi:10.1016/j.jconrel.2015.12.025 · Cited by 30 · see citing works
Effects of fatigue on the chemical and mechanical degradation of model stent sub-units
Dreher, M. L., Nagaraja, S., & Batchelor, B. L. (2015) · Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · doi:10.1016/j.jmbbm.2015.12.020 · Cited by 28 · see citing works
Collet, C. A., Asano, T., Sotomi, Y., et al. (2016) · Minerva Cardiology and Angiology · doi:10.23736/s0026-4725.16.04238-9 · Cited by 27 · see citing works
Longitudinal compression behaviour of coronary stents: a bench-top comparative study
Barragan, P. T., Garitey, V., Mouneimne, K., et al. (2014) · EuroIntervention · doi:10.4244/eijv9i12a243 · Cited by 20 · see citing works
The Systems Biocompatibility of Coronary Stenting
Kolandaivelu, K., & Nezami, F. R. (2016) · Interventional Cardiology Clinics · doi:10.1016/j.iccl.2016.02.001 · Cited by 13 · see citing works
Self-citationNeedles in Our Technology Haystacks
Edelman, E. R., & Wang, P.-J. (2017) · Circulation Cardiovascular Interventions · doi:10.1161/circinterventions.117.006059 · Cited by 3 · see citing works
Self-citationEdelman, E. R., & Wang, P.-J. (2017) · PMC · Cited by 1 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Oosterbeek, R. N., Kwon, K.-A., Duffy, P., McMahon, S., Zhang, X. C., Best, S. M., & Cameron, R. E. (2019) · Polymer Degradation and Stability · doi:10.1016/j.polymdegradstab.2019.109015 · Cited by 30 · see citing works
“3), consistent with recent works illustrating the plasticising effect of water on the bulk mechanical properties of PLLA [44, 45].”
Background
- Authors
- Wang, P.-J., Nezami, F. R., Gorji, M. B., Berti, F., Petrini, L., Wierzbicki, T., Migliavacca, F., & Edelman, E. R.
- Affiliations
- Boston University (US)
- Massachusetts Institute of Technology (US)
- Politecnico di Milano (IT)
- Brigham and Women's Hospital (US)
- Corresponding author
- Wang, Pei-Jiang
- ORCID
- Venue
- Acta Biomaterialia
- Publisher
- Elsevier BV
- Published
- 2018-12
- Volume
- 82
- Pages
- 34-43
- ISSN
- 1742-7061 (ISSN-L)
- PMC
- PMC6295666
- OpenAlex
- W2897287626
- Cited by
- 31
- References
- 30
- Citations by year
- 2025: 4 · 2024: 4 · 2023: 3 · 2022: 8 · 2021: 3 · 2020: 4 · 2019: 4 · 2018: 1
- FWCI
- 3.12
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- Locations
- Acta Biomaterialia (acceptedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- Virtual Community of Pathological Anatomy (University of Castilla La Mancha) (submittedVersion, other-oa)
- DSpace@MIT (Massachusetts Institute of Technology) (submittedVersion, cc-by-nc-nd)
- PMC
- Funders
- Boston Scientific Corporation
- National Institutes of Health
- Topics
- Health Sciences › Medicine › Surgery › Coronary Interventions and Diagnostics
- Physical Sciences › Engineering › Biomedical Engineering › Elasticity and Material Modeling
- Physical Sciences › Materials Science › Biomaterials › biodegradable polymer synthesis and properties
- Keywords
- bioresorbable vascular scaffolds, polymer properties, work environment, stress concentration, microstructural damage, structural integrity, thrombosis, device failure, material models, coronary stents
- MeSH terms
- Blood Vessel Prosthesis, Humans, Prosthesis Design, Stents, Absorbable Implants, Tissue Scaffolds
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review