Repeatability of 18F‐FDG PET radiomic features: A phantom study to explore sensitivity to image reconstruction settings, noise, and delineation method

Open access
Pfaehler, Elisabeth, Beukinga, Roelof J., de Jong, Johan R., Slart, Riemer H. J. A., Slump, Cornelis H., Dierckx, Rudi A. J. O., Boellaard, Ronald · 2019 · Medical Physics
Cited by 116

The DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.

Abstract

Background: 18 F‐fluoro‐2‐deoxy‐D‐Glucose positron emission tomography (18 F‐FDG PET) radiomics has the potential to guide the clinical decision making in cancer patients, but validation is required before radiomics can be implemented in the clinical setting. The aim of this study was to explore how feature space reduction and repeatability of 18 F‐FDG PET radiomic features are affected by various sources of variation such as underlying data (e.g., object size and uptake), image reconstruction methods and settings, noise, discretization method, and delineation method. Methods: The NEMA image quality phantom was scanned with various sphere‐to‐background ratios (SBR), simulating different activity uptakes, including spheres with low uptake, that is, SBR smaller than 1. Furthermore, images of a phantom containing 3D printed inserts reflecting realistic heterogeneity uptake patterns were acquired. Data were reconstructed using various matrix sizes, reconstruction algorithms, and scan durations (noise). For every specific reconstruction and noise level, ten statistically equal replicates were generated. The phantom inserts were delineated using CT and PET‐based segmentation methods. A total of 246 radiomic features was extracted from each image dataset. Images were discretized with a fixed number of 64 bins (FBN) and a fixed bin width (FBW) of 0.25 for the high and a FBW of 0.05 for the low uptake data. In terms of feature reduction, we determined the impact of these factors on the composition of feature clusters, which were defined on the basis of Spearman's correlation matrices. To assess feature repeatability, the intraclass correlation coefficient was calculated over the ten replicates. Results: In general, larger spheres with high uptake resulted in better repeatability compared to smaller low uptake spheres. In terms of repeatability, features extracted from heterogeneous phantom inserts were comparable to features extracted from bigger high uptake spheres. For example, for an EARL‐compliant reconstruction, larger and smaller high uptake spheres yielded good repeatability for 32% and 30% of the features, while the heterogeneous inserts resulted in 34% repeatable features. For the low uptake spheres, this was the case for 22% and 20% of the features for bigger and smaller spheres, respectively. Images reconstructed with point‐spread‐function (PSF) resulted in the highest repeatability when compared with OSEM or time‐of‐flight, for example, 53%, 30%, and 32% of repeatable features, respectively (for unsmoothed data, discretized with FBN, 300 s scan duration). Reducing image noise (increasing scan duration and smoothing) and using CT‐based segmentation for the low uptake spheres yielded improved repeatability. FBW discretization resulted in higher repeatability than FBN discretization, for example, 89% and 35% of the features, respectively (for the EARL‐compliant reconstruction and larger high uptake spheres). Conclusion: Feature space reduction and repeatability of 18 F‐FDG PET radiomic features depended on all studied factors. The high sensitivity of PET radiomic features to image quality suggests that a high level of image acquisition and preprocessing standardization is required to be used as clinical imaging biomarker.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1002/mp.13322

Cites

5 of 5 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 5 references

Cited by

See all 116

Metadata

Authors
Pfaehler, E., Beukinga, R. J., de Jong, J. R., Slart, R. H. J. A., Slump, C. H., Dierckx, R. A. J. O., & Boellaard, R.
Venue
Medical Physics
Publisher
Wiley
Published
2018-12-28
Volume
46
Issue
2
Pages
665-678
ISSN
  • 0094-2405 (ISSN-L)
  • 2473-4209
OpenAlex
W2902316253
Cited by
116
References
5
FWCI
6.62
Type
journal article
Language
en
Open access
hybrid

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Repeatability of 18F‐FDG PET radiomic features: A phantom study to explore sensitivity to image reconstruction settings, noise, and delineation method. https://app.cochat.ai/papers/10.1002/mp.13322
The paper
Pfaehler, E., Beukinga, R. J., de Jong, J. R., Slart, R. H. J. A., Slump, C. H., Dierckx, R. A. J. O., & Boellaard, R. (2019). Repeatability of 18F‐FDG PET radiomic features: A phantom study to explore sensitivity to image reconstruction settings, noise, and delineation method. Medical Physics. https://doi.org/10.1002/mp.13322

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.