CBCT‐based synthetic CT generation using deep‐attention cycleGAN for pancreatic adaptive radiotherapy

Open access
Liu, Yingzi, Lei, Yang, Wang, Tonghe, Fu, Yabo, Tang, Xiangyang, Curran, Walter J., Liu, Tian, Patel, Pretesh, Yang, Xiaofeng · 2020 · Medical Physics

Abstract

Purpose: Current clinical application of cone‐beam CT (CBCT) is limited to patient setup. Imaging artifacts and Hounsfield unit (HU) inaccuracy make the process of CBCT‐based adaptive planning presently impractical. In this study, we developed a deep‐learning‐based approach to improve CBCT image quality and HU accuracy for potential extended clinical use in CBCT‐guided pancreatic adaptive radiotherapy. Methods: Thirty patients previously treated with pancreas SBRT were included. The CBCT acquired prior to the first fraction of treatment was registered to the planning CT for training and generation of synthetic CT (sCT). A self‐attention cycle generative adversarial network (cycleGAN) was used to generate CBCT‐based sCT. For the cohort of 30 patients, the CT‐based contours and treatment plans were transferred to the first fraction CBCTs and sCTs for dosimetric comparison. Results: At the site of abdomen, mean absolute error (MAE) between CT and sCT was 56.89 ± 13.84 HU, comparing to 81.06 ± 15.86 HU between CT and the raw CBCT. No significant differences (P > 0.05) were observed in the PTV and OAR dose‐volume‐histogram (DVH) metrics between the CT‐ and sCT‐based plans, while significant differences (P < 0.05) were found between the CT‐ and the CBCT‐based plans. Conclusions: The image similarity and dosimetric agreement between the CT and sCT‐based plans validated the dose calculation accuracy carried by sCT. The CBCT‐based sCT approach can potentially increase treatment precision and thus minimize gastrointestinal toxicity.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1002/mp.14121

Cites

4 of 4 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 4 references

Cited by

See all 205

Metadata

Authors
Liu, Y., Lei, Y., Wang, T., Fu, Y., Tang, X., Curran, W. J., Liu, T., Patel, P., & Yang, X.
Venue
Medical Physics
Publisher
Wiley
Published
2020-03-28
Volume
47
Issue
6
Pages
2472-2483
ISSN
  • 0094-2405 (ISSN-L)
  • 2473-4209
OpenAlex
W3010126053
Cited by
205
References
4
FWCI
19.56
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: CBCT‐based synthetic CT generation using deep‐attention cycleGAN for pancreatic adaptive radiotherapy. https://app.cochat.ai/papers/10.1002/mp.14121
The paper
Liu, Y., Lei, Y., Wang, T., Fu, Y., Tang, X., Curran, W. J., Liu, T., Patel, P., & Yang, X. (2020). CBCT‐based synthetic CT generation using deep‐attention cycleGAN for pancreatic adaptive radiotherapy. Medical Physics. https://doi.org/10.1002/mp.14121

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.