Quantification of intravoxel velocity standard deviation and turbulence intensity by generalizing phase‐contrast MRI
Open accessThe DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
Turbulent flow, characterized by velocity fluctuations, is a contributing factor to the pathogenesis of several cardiovascular diseases. A clinical noninvasive tool for assessing turbulence is lacking, however. It is well known that the occurrence of multiple spin velocities within a voxel during the influence of a magnetic gradient moment causes signal loss in phase‐contrast magnetic resonance imaging (PC‐MRI). In this paper a mathematical derivation of an expression for computing the standard deviation (SD) of the blood flow velocity distribution within a voxel is presented. The SD is obtained from the magnitude of PC‐MRI signals acquired with different first gradient moments. By exploiting the relation between the SD and turbulence intensity (TI), this method allows for quantitative studies of turbulence. For validation, the TI in an in vitro flow phantom was quantified, and the results compared favorably with previously published laser Doppler anemometry (LDA) results. This method has the potential to become an important tool for the noninvasive assessment of turbulence in the arterial tree. Magn Reson Med, 2006. © 2006 Wiley‐Liss, Inc.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Spin Diffusion Measurements: Spin Echoes in the Presence of a Time-Dependent Field Gradient
Stejskal, E. O., & Tanner, J. E. (1965) · The Journal of Chemical Physics · doi:10.1063/1.1695690 · Cited by 8,286
Hemodynamic shear stress and its role in atherosclerosis
Malek, A. M. (1999) · JAMA · doi:10.1001/jama.282.21.2035 · Cited by 3,650
McDonald's Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles
Vlachopoulos, C., O'Rourke, M., & Nichols, W. W. (2011) · doi:10.1201/b13568 · Cited by 1,613
Cited by
4D flow cardiovascular magnetic resonance consensus statement
Dyverfeldt, P., Bissell, M., Barker, A. J., Bolger, A. F., Carlhäll, C.-J., Ebbers, T., Francios, C. J., Frydrychowicz, A., Geiger, J., Giese, D., et al. (2015) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-015-0174-5 · Cited by 877
Self-citationMarkl, M., Kilner, P. J., & Ebbers, T. (2011) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-13-7 · Cited by 463
Self-citation4D Flow cardiovascular magnetic resonance consensus statement: 2023 update
Bissell, M. M., Raimondi, F., Ait Ali, L., Allen, B. D., Barker, A. J., Bolger, A., Burris, N., Carhäll, C.-J., Collins, J. D., Ebbers, T., et al. (2023) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-023-00942-z · Cited by 296
Self-citationStanković, Z. B., Allen, B. D., García, J., Jarvis, K., & Markl, M. (2014) · PubMed · doi:10.3978/j.issn.2223-3652.2014.01.02 · Cited by 281
Viscous energy loss in the presence of abnormal aortic flow
Barker, A. J., van Ooij, P., Bandi, K., Garcia, J., Albaghdadi, M., McCarthy, P., Bonow, R. O., Carr, J., Collins, J., Malaisrie, S. C., et al. (2014) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24962 · Cited by 181
Dyverfeldt, P., Hope, M. D., Tseng, E. E., & Saloner, D. (2013) · JACC: Cardiovascular Imaging · doi:10.1016/j.jcmg.2012.07.017 · Cited by 174
Self-citationClough, R. E., Waltham, M., Giese, D., Taylor, P. R., & Schaeffter, T. (2012) · Journal of Vascular Surgery · doi:10.1016/j.jvs.2011.11.005 · Cited by 159
Metadata
- Authors
- Dyverfeldt, P., Sigfridsson, A., Kvitting, J. E., & Ebbers, T.
- Venue
- Magnetic Resonance in Medicine
- Publisher
- Wiley
- Published
- 2006-09-06
- Volume
- 56
- Issue
- 4
- Pages
- 850-858
- ISSN
- 0740-3194 (ISSN-L)
- 1522-2594
- OpenAlex
- W1993237717
- Cited by
- 162
- References
- 3
- FWCI
- 3.02
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.