Chemical exchange saturation transfer (CEST): what is in a name and what isn't?
Open accessAbstract
Chemical exchange saturation transfer (CEST) imaging is a relatively new magnetic resonance imaging contrast approach in which exogenous or endogenous compounds containing either exchangeable protons or exchangeable molecules are selectively saturated and after transfer of this saturation, detected indirectly through the water signal with enhanced sensitivity. The focus of this review is on basic magnetic resonance principles underlying CEST and similarities to and differences with conventional magnetization transfer contrast. In CEST magnetic resonance imaging, transfer of magnetization is studied in mobile compounds instead of semisolids. Similar to magnetization transfer contrast, CEST has contributions of both chemical exchange and dipolar cross‐relaxation, but the latter can often be neglected if exchange is fast. Contrary to magnetization transfer contrast, CEST imaging requires sufficiently slow exchange on the magnetic resonance time scale to allow selective irradiation of the protons of interest. As a consequence, magnetic labeling is not limited to radio‐frequency saturation but can be expanded with slower frequency‐selective approaches such as inversion, gradient dephasing and frequency labeling. The basic theory, design criteria, and experimental issues for exchange transfer imaging are discussed. A new classification for CEST agents based on exchange type is proposed. The potential of this young field is discussed, especially with respect to in vivo application and translation to humans. Magn Reson Med, 2011. © 2011 Wiley‐Liss, Inc.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Differences in APT may also be influenced by pH and longitudinal relaxation time (T1), meaning that as a biomarker, CEST distinguishes normal controls from rTg4510 mice but does so in a non-specific manner.
“However the differences in APT may also be influenced by pH and the longitudinal relaxation time (T1), which may change due to cell death, for example (van Zijl and Yadav, 2011), meaning that as a biomarker, CEST distinguishes normal controls from rTg4510 mice but does so in a non-specific manner.”
From In vivo imaging of tau pathology using multi-parametric quantitative MRI (2015) · cites this paper as van Zijl and Yadav, 2011
The paper’s ‘In vivo CEST imaging and translation to humans’ section says pH reduction can reduce APT/CEST contrast, and ‘Signal and Parameter Quantification’ says water T1 must be known to quantify a CEST parameter. These passages support pH and T1 as influences, but the paper does not link a T1 change to cell death. The rTg4510 comparison is the citing authors’ conclusion, not a finding attributed to this source. (LLM, conf=0.84)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
No references could be resolved yet.
Research works printed but not linked: 142
- [1]Ward KM Aletras AH Balaban RS A new class of contrast agents for MRI based on proton chemical exchange dependent saturation transfer (CEST) J Magn Reson 2000 143 79 87 10698648
- [2]Zhang S Merritt M Woessner DE Lenkinski RE Sherry AD PARACEST agents: modulating MRI contrast via water proton exchange Acc Chem Res 2003 36783 36790 14567712
- [3]Woods M Woessner DE Sherry AD Paramagnetic lanthanide complexes as PARACEST agents for medical imaging Chem Soc Rev 2006 35 500 511 16729144 PMC2718840
- [4]Zhou J van Zijl P Chemical exchange saturation transfer imaging and spectroscopy Prog Nucl Magn Reson Spectrosc 2006 48 109 136 33198968
- [5]Sherry AD Woods M Chemical exchange saturation transfer contrast agents for magnetic resonance imaging Annu Rev Biomed Eng 2008 10 391 411 18647117 PMC2709739
- [6]Terreno E Delli Castelli D Cabella C Dastru W Sanino A Stancanello J Tei L Aime S Paramagnetic liposomes as innovative contrast agents for magnetic resonance (MR) molecular imaging applications Chem Biodivers 2008 5 1901 1912 18972531
- [7]Aime S Castelli DD Terreno E Chapter 10 - Lanthanide-loaded paramagnetic liposomes as switchable magnetically oriented nanovesicles Methods Enzymol 2009 464 193 210 19903556
- [8]Terreno E Stancanello J Longo D Castelli DD Milone L Sanders HM Kok MB Uggeri F Aime S Methods for an improved detection of the MRI-CEST effect Contrast Media Mol Imaging 2009 4 237 247 19839029
- [9]Hancu I Dixon WT Woods M Vinogradov E Sherry AD Lenkinski RE CEST and PARACEST MR contrast agents Acta Radiol 2010 51 910 923 20828299 PMC6594364
- [10]Terreno E Castelli DD Aime S Encoding the frequency dependence in MRI contrast media: the emerging class of CEST agents Contrast Media Mol Imaging 2010 5 78 98 20419761
- [11]Terreno E Castelli DD Viale A Aime S Challenges for molecular magnetic resonance imaging Chem Rev 2010 110 3019 3042 20415475
- [12]Wolff SD Balaban RS Magnetization transfer contrast (MTC) and tissue water proton relaxation in vivo Magn Reson Med 1989 10 135 144 2547135
- [13]Bryant RG The dynamics of water-protein interactions Annu Rev Biophys Biomol Struct 1996 25 29 53 8800463
- [14]Guivel-Scharen V Sinnwell T Wolff SD Balaban RS Detection of proton chemical exchange between metabolites and water in biological tissues J Magn Reson 1998 133 36 45 9654466
- [16]Li AX Hudson RH Barrett JW Jones CK Pasternak SH Bartha R Four-pool modeling of proton exchange processes in biological systems in the presence of MRI-paramagnetic chemical exchange saturation transfer (PARACEST) agents Magn Reson Med 2008 60 1197 1206 18958857
- [18]McMahon MT Gilad AA Zhou J Sun PZ Bulte JW van Zijl PC Quantifying exchange rates in chemical exchange saturation transfer agents using the saturation time and saturation power dependencies of the magnetization transfer effect on the magnetic resonance imaging signal (QUEST and QUESP): pH calibration for poly-L-lysine and a starburst dendrimer Magn Reson Med 2006 55 836 847 16506187 PMC2860536
- [19]Sun PZ Simplified and scalable numerical solution for describing multi-pool chemical exchange saturation transfer (CEST) MRI contrast J Magn Reson 2010 205 235 241 20570196 PMC2902598
- [20]Sun PZ Zhou J Huang J van Zijl P Simplified quantitative description of amide proton transfer (APT) imaging during acute ischemia Magn Reson Med 2007 57 405 410 17260362
- [22]Woessner DE Zhang S Merritt ME Sherry AD Numerical solution of the Bloch equations provides insights into the optimum design of PARACEST agents for MRI Magn Reson Med 2005 53 790 799 15799055
- [23]Zhou J Wilson DA Sun PZ Klaus JA Van Zijl PC Quantitative description of proton exchange processes between water and endogenous and exogenous agents for WEX, CEST, and APT experiments Magn Reson Med 2004 51 945 952 15122676
- [24]Snoussi K Bulte JW Gueron M van Zijl PC Sensitive CEST agents based on nucleic acid imino proton exchange: detection of poly(rU) and of a dendrimer-poly(rU) model for nucleic acid delivery and pharmacology Magn Reson Med 2003 49 998 1005 12768576
- [25]Zhou J Payen JF Wilson DA Traystman RJ van Zijl PC Using the amide proton signals of intracellular proteins and peptides to detect pH effects in MRI Nat Med 2003 9 1085 1090 12872167
- [26]van Zijl PC Jones CK Ren J Malloy CR Sherry AD MRI detection of glycogen in vivo by using chemical exchange saturation transfer imaging (glycoCEST) Proc Natl Acad Sci USA 2007 104 4359 4364 17360529 PMC1838607
- [27]Aime S Barge A Delli Castelli D Fedeli F Mortillaro A Nielsen FU Terreno E Paramagnetic lanthanide(III) complexes as pH-sensitive chemical exchange saturation transfer (CEST) contrast agents for MRI applications Magn Reson Med 2002 47 639 648 11948724
- [28]Aime S Delli Castelli D Fedeli F Terreno E A paramagnetic MRI-CEST agent responsive to lactate concentration J Am Chem Soc 2002 124 9364 9365 12167018
- [29]Zhang S Trokowski R Sherry AD A paramagnetic CEST agent for imaging glucose by MRI J Am Chem Soc 2003 125 15288 15289 14664562
- [30]Zhang S Winter P Wu K Sherry AD A novel europium(III)-based MRI contrast agent J Am Chem Soc 2001 123 1517 1518 11456734
- [31]Goffeney N Bulte JW Duyn J Bryant LH Jr van Zijl PC Sensitive NMR detection of cationic-polymer-based gene delivery systems using saturation transfer via proton exchange J Am Chem Soc 2001 123 8628 8629 11525684
- [32]Ward KM Balaban RS Determination of pH using water protons and chemical exchange dependent saturation transfer (CEST) Magn Reson Med 2000 44 799 802 11064415
- [33]Zhou J Lal B Wilson DA Laterra J van Zijl PC Amide proton transfer (APT) contrast for imaging of brain tumors Magn Reson Med 2003 50 1120 1126 14648559
- [34]Ling W Regatte RR Navon G Jerschow A Assessment of glycosaminoglycan concentration in vivo by chemical exchange-dependent saturation transfer (gagCEST) Proc Natl Acad Sci USA 2008 105 2266 2270 18268341 PMC2268124
- [35]Aime S Delli Castelli D Terreno E Highly sensitive MRI chemical exchange saturation transfer agents using liposomes Angew Chem Int Ed 2005 44 5513 5515 16052647
- [36]Friedman JI McMahon MT Stivers JT van Zijl PC Indirect detection of labile solute proton spectra via the water signal using frequency-labeled exchange (FLEX) transfer J Am Chem Soc 2010 132 1813 1815 20095603 PMC2822835
- [37]Zhao JM Har-el YE McMahon MT Zhou J Sherry AD Sgouros G Bulte JW van Zijl PC Size-induced enhancement of chemical exchange saturation transfer (CEST) contrast in liposomes J Am Chem Soc 2008 130 5178 5184 18361490 PMC2759111
- [38]Schroder L Lowery TJ Hilty C Wemmer DE Pines A Molecular imaging using a targeted magnetic resonance hyperpolarized biosensor Science 2006 314 446 449 17053143
- [39]Ratnakar SJ Woods M Lubag AJ Kovacs Z Sherry AD Modulation of water exchange in europium(III) DOTA-tetraamide complexes via electronic substituent effects J Am Chem Soc 2008 130 6 7 18067296 PMC2716115
- [40]Cavanagh J Fairbrother WJ Palmer AF III Skelton NJ Protein NMR spectroscopy: principles and practice 1996 San Diego Academic Press
- [41]Ernst RR Bodenhausen G Wokaun A Principles of nuclear magnetic resonance in one and two dimensions 1988 Oxford Oxford University Press
- [42]Wüthrich K NMR of proteins and nucleic acids 1986 New York Wiley
- [43]Ling W Eliav U Navon G Jerschow A Chemical exchange saturation transfer by intermolecular double-quantum coherence J Magn Reson 2008 194 29 32 18571444 PMC2562429
- [44]Hwang TL van Zijl PC Mori S Accurate quantitation of water-amide proton exchange rates using the phase-modulated CLEAN chemical EXchange (CLEANEX-PM) approach with a Fast-HSQC (FHSQC) detection scheme J Biomol NMR 1998 11 221 226 9679296
- [45]Mori S Abeygunawardana C van Zijl PC Berg JM Water exchange filter with improved sensitivity (WEX II) to study solvent-exchangeable protons. Application to the consensus zinc finger peptide CP-1 J Magn Reson B 1996 110 96 101 8556240
- [47]Mori S Eleff SM Pilatus U Mori N van Zijl PC Proton NMR spectroscopy of solvent-saturable resonances: a new approach to study pH effects in situ Magn Reson Med 1998 40 36 42 9660550
- [48]van Zijl PC Zhou J Mori N Payen JF Wilson D Mori S Mechanism of magnetization transfer during on-resonance water saturation. A new approach to detect mobile proteins, peptides, and lipids Magn Reson Med 2003 49 440 449 12594746
- [49]Otting G Liepinsh E Wuthrich K Protein hydration in aqueous solution Science 1991 254 974 980 1948083
- [50]Hua J Jones CK Blakeley J Smith SA van Zijl PC Zhou J Quantitative description of the asymmetry in magnetization transfer effects around the water resonance in the human brain Magn Reson Med 2007 58 786 793 17899597 PMC3707117
- [51]Pekar J Jezzard P Roberts DA Leigh JS Jr Frank JA McLaughlin AC Perfusion imaging with compensation for asymmetric magnetization transfer effects Magn Reson Med 1996 35 70 79 8771024
- [52]Swanson S Pang Y MT is symmetric but shifted with respect to water 2003 Proceedings of the 11th Annual Meeting of ISMRM Ontario, Canada 660
- [53]Englander SW Kallenbach NR Hydrogen exchange and structural dynamics of proteins and nucleic acids Q Rev Biophys 1984 16 521 655 6204354
- [54]Gillies RJ Raghunand N Garcia-Martin ML Gatenby RA pH imaging. A review of pH measurement methods and applications in cancers IEEE Eng Med Biol Mag 2004 23 57 64 15565800
- [55]Griffiths JR Are cancer cells acidic? Br J Cancer 1991 64 425 427 1911181 PMC1977628
- [56]Balaban RS Ceckler TL Magnetization transfer contrast in magnetic resonance imaging Magn Reson Q 1992 8 116 137 1622774
- [57]Henkelman RM Huang X Xiang QS Stanisz GJ Swanson SD Bronskill MJ Quantitative interpretation of magnetization transfer Magn Reson Med 1993 29 759 766 8350718
- [58]Henkelman RM Stanisz GJ Graham SJ Magnetization transfer in MRI: a review NMR Biomed 2001 14 57 64 11320533
- [59]Leibfritz D Dreher W Magnetization transfer MRS NMR Biomed 2001 14 65 76 11320534
- [60]McGowan JC 3rd Schnall MD Leigh JS Magnetization transfer imaging with pulsed off-resonance saturation: variation in contrast with saturation duty cycle J Magn Reson Imaging 1994 4 79 82 8148561
- [61]Pike GB Pulsed magnetization transfer contrast in gradient echo imaging: a two-pool analytic description of signal response Magn Reson Med 1996 36 95 103 8795027
- [62]Pike GB Glover GH Hu BS Enzmann DR Pulsed magnetization transfer spin-echo MR imaging J Magn Reson Imaging 1993 3 531 539 8324313
- [63]Graham SJ Henkelman RM Understanding pulsed magnetization transfer J Magn Reson Imaging 1997 7 903 912 9307918
- [64]Hu BS Conolly SM Wright GA Nishimura DG Macovski A Pulsed saturation transfer contrast Magn Reson Med 1992 26 231 240 1325023
- [65]Yeung HN Aisen AM Magnetization transfer contrast with periodic pulsed saturation Radiology 1992 183 209 214 1549673
- [66]Gochberg DF Gore JC Quantitative imaging of magnetization transfer using an inversion recovery sequence Magn Reson Med 2003 49 501 505 12594753
- [67]Sun PZ Benner T Kumar A Sorensen AG Investigation of optimizing and translating pH-sensitive pulsed-chemical exchange saturation transfer (CEST) imaging to a 3T clinical scanner Magn Reson Med 2008 60 834 841 18816867 PMC2678923
- [68]Sun PZ Zhou J Sun W Huang J van Zijl PC Detection of the ischemic penumbra using pH-weighted MRI J Cereb Blood Flow Metab 2007 27 1129 1136 17133226
- [69]Vinogradov E He H Lubag A Balschi JA Sherry AD Lenkinski RE MRI detection of paramagnetic chemical exchange effects in mice kidneys in vivo Magn Reson Med 2007 58 650 655 17899603
- [70]Vinogradov E Zhang S Lubag A Balschi JA Sherry AD Lenkinski RE On-resonance low B1 pulses for imaging of the effects of PARACEST agents J Magn Reson 2005 176 54 63 15979362
- [71]Eliav U Navon G Enhancement of magnetization transfer effects by inter-molecular multiple quantum filtered NMR J Magn Reson 2008 190 149 153 17936657
- [72]Warren WS Richter W Andreotti AH Farmer BT 2nd Generation of impossible cross-peaks between bulk water and biomolecules in solution NMR Science 1993 262 2005 2009 8266096
- [73]Zhang S Zhu X Chen Z Cai C Lin T Zhong J Improvement in the contrast of CEST MRI via intermolecular double quantum coherences Phys Med Biol 2008 53 N287 N296 18574314
- [74]Ropele S Stollberger R Hartung HP Fazekas F Estimation of magnetization transfer rates from PACE experiments with pulsed RF saturation J Magn Reson Imaging 2000 12 749 756 11050646
- [75]Sled JG Pike GB Quantitative imaging of magnetization transfer exchange and relaxation properties in vivo using MRI Magn Reson Med 2001 46 923 931 11675644
- [76]Yarnykh VL Pulsed Z-spectroscopic imaging of cross-relaxation parameters in tissues for human MRI: theory and clinical applications Magn Reson Med 2002 47 929 939 11979572
- [77]Atreya HS Szyperski T Rapid NMR data collection Methods in enzymology 2005 394 78 108 15808218
- [78]Frydman L Scherf T Lupulescu A The acquisition of multidimensional NMR spectra within a single scan Proc Natl Acad Sci USA 2002 99 15858 15862 12461169 PMC138528
- [79]Mobli M Stern AS Hoch JC Spectral reconstruction methods in fast NMR: reduced dimensionality, random sampling and maximum entropy J Magn Reson 2006 182 96 105 16815055
- [80]Zhou J Blakeley JO Hua J Kim M Laterra J Pomper MG van Zijl PC Practical data acquisition method for human brain tumor amide proton transfer (APT) imaging Magn Reson Med 2008 60 842 849 18816868 PMC2579754
- [81]Kim M Gillen J Landman BA Zhou J van Zijl PC Water saturation shift referencing (WASSR) for chemical exchange saturation transfer (CEST) experiments Magn Reson Med 2009 61 1441 1450 19358232 PMC2860191
- [82]Mulkern RV Williams ML The general solution to the Bloch equation with constant rf and relaxation terms: application to saturation and slice selection Med Phys 1993 20 5 13 8455512
- [83]Metzger GJ Snyder C Akgun C Vaughan T Ugurbil K Van de Moortele PF Local B1+ shimming for prostate imaging with transceiver arrays at 7T based on subject-dependent transmit phase measurements Magn Reson Med 2008 59 396 409 18228604 PMC4406352
- [84]Setsompop K Alagappan V Zelinski AC Potthast A Fontius U Hebrank F Schmitt F Wald LL Adalsteinsson E High-flip-angle slice-selective parallel RF transmission with 8 channels at 7 T J Magn Reson 2008 195 76 84 18799336 PMC2610679
- [85]Zhu H Jones CK van Zijl PC Barker PB Zhou J Fast 3D chemical exchange saturation transfer (CEST) imaging of the human brain Magn Reson Med 2010 64 638 644 20632402 PMC2932772
- [86]Dixon WT Hancu I Ratnakar SJ Sherry AD Lenkinski RE Alsop DC A multislice gradient echo pulse sequence for CEST imaging Magn Reson Med 2010 63 253 256 19918889 PMC2925508
- [87]Mougin OE Coxon RC Pitiot A Gowland PA Magnetization transfer phenomenon in the human brain at 7 T NeuroImage 2010 49 272 281 19683581
- [88]Sun PZ Murata Y Lu J Wang X Lo EH Sorensen AG Relaxation-compensated fast multislice amide proton transfer (APT) imaging of acute ischemic stroke Magn Reson Med 2008 59 1175 1182 18429031
- [89]Sun PZ Zhou J Sun W Huang J van Zijl PC Suppression of lipid artifacts in amide proton transfer imaging Magn Reson Med 2005 54 222 225 15968669
- [90]Dixon WT Ren J Lubag AJ Ratnakar J Vinogradov E Hancu I Lenkinski RE Sherry AD A concentration-independent method to measure exchange rates in PARACEST agents Magn Reson Med 2010 63 625 632 20187174 PMC2925112
- [91]Aime S Delli Castelli D Terreno E Novel pH-reporter MRI contrast agents Angew Chem Int Ed 2002 41 4334 4336 12434381
- [92]Terreno E Delli Castelli D Cravotto G Milone L Aime S Ln(III)-DOTAMGly complexes: a versatile series to assess the determinants of the efficacy of paramagnetic chemical exchange saturation transfer agents for magnetic resonance imaging applications Invest Radiol 2004 39 235 243 15021328
- [93]Delli Castelli D Dastru W Terreno E Cittadino E Mainini F Torres E Spadaro M Aime S In vivo MRI multicontrast kinetic analysis of the uptake and intracellular trafficking of paramagnetically labeled liposomes J Control Release 2010 144 271 279 20230865
- [94]Burdinski D Pikkemaat JA Lub J de Peinder P Nieto Garrido L Weyhermuller T Lanthanide complexes of triethylenetetramine tetra-, penta-, and hexaacetamide ligands as paramagnetic chemical exchange-dependent saturation transfer contrast agents for magnetic resonance imaging: nona- versus decadentate coordination Inorg Chem 2009 48 6692 6712 19507818
- [95]Huang CH Hammell J Ratnakar SJ Sherry AD Morrow JR Activation of a PARACEST agent for MRI through selective outersphere interactions with phosphate diesters Inorg Chem 2010 49 5963 5970 20509631 PMC2893239
- [96]Huang CH Morrow JR A PARACEST agent responsive to inner- and outer-sphere phosphate ester interactions for MRI applications J Am Chem Soc 2009 131 4206 4207 19317496 PMC2744982
- [97]Huang CH Morrow JR Cerium(III), europium(III), and ytterbium(III) complexes with alcohol donor groups as chemical exchange saturation transfer agents for MRI Inorg Chem 2009 48 7237 7243 19722692 PMC2739124
- [98]Perez-Mayoral E Negri V Soler-Padros J Cerdan S Ballesteros P Chemistry of paramagnetic and diamagnetic contrast agents for magnetic resonance imaging and spectroscopy pH responsive contrast agents Eur J Radiol 2008 67 453 458 18455343
- [99]Pikkemaat JA Wegh RT Lamerichs R van de Molengraaf RA Langereis S Burdinski D Raymond AY Janssen HM de Waal BF Willard NP Meijer EW Grull H Dendritic PARACEST contrast agents for magnetic resonance imaging Contrast Media Mol Imaging 2007 2 229 239 17937448
- [100]Sun PZ Sorensen AG Imaging pH using the chemical exchange saturation transfer (CEST) MRI: Correction of concomitant RF irradiation effects to quantify CEST MRI for chemical exchange rate and pH Magn Reson Med 2008 60 390 397 18666128
- [101]Longo DL Dastru W Digilio G Keupp J Langereis S Lanzardo S Prestigio S Steinbach O Terreno E Uggeri F Aime S Iopamidol as a responsive MRI-chemical exchange saturation transfer contrast agent for pH mapping of kidneys: in vivo studies in mice at 7T Magn Reson Med 2010 In press 20949634
- [102]Aime S Barge A Cabella C Crich SG Gianolio E Targeting cells with MR imaging probes based on paramagnetic Gd(III) chelates Curr Pharm Biotechnol 2004 5 509 518 15579040
- [103]Liu G Li Y Pagel MD Design and characterization of a new irreversible responsive PARACEST MRI contrast agent that detects nitric oxide Magn Reson Med 2007 58 1249 1256 18046705
- [104]Trokowski R Zhang S Sherry AD Cyclen-based phenylboronate ligands and their Eu3+ complexes for sensing glucose by MRI Bioconjug Chem 2004 15 1431 1440 15546212
- [105]Jones CK Schlosser MJ van Zijl PC Pomper MG Golay X Zhou J Amide proton transfer imaging of human brain tumors at 3T Magn Reson Med 2006 56 585 592 16892186
- [106]Wen Z Hu S Huang F Wang X Guo L Quan X Wang S Zhou J MR imaging of high-grade brain tumors using endogenous protein and peptide-based contrast NeuroImage 2010 51 616 622 20188197 PMC2856810
- [107]Angelovski G Chauvin T Pohmann R Logothetis NK Toth E Calcium-responsive paramagnetic CEST agents Bioorg Med Chem 2010 20691598
- [108]Trokowski R Ren J Kalman FK Sherry AD Selective sensing of zinc ions with a PARACEST contrast agent Angew Chem Int Ed 2005 44 6920 6923 16206314
- [109]Aime S Delli Castelli D Lawson D Terreno E Gd-loaded liposomes as T1, susceptibility, and CEST agents, all in one J Am Chem Soc 2007 129 2430 2431 17288421
- [110]Terreno E Castelli DD Milone L Rollet S Stancanello J Violante E Aime S First ex-vivo MRI co-localization of two LIPOCEST agents Contrast Media Mol Imaging 2008 3 38 43 18335476
- [111]Terreno E Delli Castelli D Violante E Sanders HM Sommerdijk NA Aime S Osmotically shrunken LIPOCEST agents: an innovative class of magnetic resonance imaging contrast media based on chemical exchange saturation transfer Chemistry 2009 15 1440 1448 19115311
- [112]Vasalatiy O Gerard RD Zhao P Sun X Sherry AD Labeling of adenovirus particles with PARACEST agents Bioconjug Chem 2008 19 598 606 18254605 PMC2696891
- [113]Aime S Delli Castelli D Terreno E Supramolecular adducts between poly-L-arginine and [TmIIIdotp]: a route to sensitivity-enhanced magnetic resonance imaging-chemical exchange saturation transfer agents Angew Chem Int Ed 2003 42 4527 4529 14520757
- [114]Ali MM Liu G Shah T Flask CA Pagel MD Using two chemical exchange saturation transfer magnetic resonance imaging contrast agents for molecular imaging studies Acc Chem Res 2009 42 915 924 19514717 PMC6010180
- [115]Ali MM Yoo B Pagel MD Tracking the relative in vivo pharmacokinetics of nanoparticles with PARACEST MRI Mol Pharmaceutics 2009 6 1409 1416 PMC4216567 19298054
- [116]Ali MM Woods M Suh EH Kovacs Z Tircso G Zhao P Kodibagkar VD Sherry AD Albumin-binding PARACEST agents J Biol Inorg Chem 2007 12 855 865 17534672 PMC2759689
- [117]Nwe K Andolina CM Huang CH Morrow JR PARACEST properties of a dinuclear neodymium(III) complex bound to DNA or carbonate Bioconjug Chem 2009 20 1375 1382 19555071 PMC2717892
- [118]Langereis S Keupp J van Velthoven JL de Roos IH Burdinski D Pikkemaat JA Grull H A temperature-sensitive liposomal 1H CEST and 19F contrast agent for MR image-guided drug delivery J Am Chem Soc 2009 131 1380 1381 19173663
- [119]Li AX Wojciechowski F Suchy M Jones CK Hudson RH Menon RS Bartha R A sensitive PARACEST contrast agent for temperature MRI: Eu3+-DOTAM-glycine (Gly)-phenylalanine (Phe) Magn Reson Med 2008 59 374 381 18228602
- [120]Zhang S Malloy CR Sherry AD MRI thermometry based on PARACEST agents J Am Chem Soc 2005 127 17572 17573 16351064 PMC2570850
- [121]Chauvin T Durand P Bernier M Meudal H Doan BT Noury F Badet B Beloeil JC Toth E Detection of enzymatic activity by PARACEST MRI: a general approach to target a large variety of enzymes Angew Chem Int Ed 2008 47 4370 4372 18454438
- [122]Liu G Bernard SM Tse T Walczak P McMahon MT Bulte JW van Zijl PCM Gilad AA Direct detection of cytosine deaminase enzymatic activity using CEST MRI 2010 Proceedings of the 18th Annual Meeting of ISMRM Stockholm, Sweden 36
- [123]Yoo B Pagel MD A PARACEST MRI contrast agent to detect enzyme activity J Am Chem Soc 2006 128 14032 14033 17061878
- [124]Yoo B Pagel MD An overview of responsive MRI contrast agents for molecular imaging Front Biosci 2008 13 1733 1752 17981664
- [125]Yoo B Raam MS Rosenblum RM Pagel MD Enzyme-responsive PARACEST MRI contrast agents: a new biomedical imaging approach for studies of the proteasome Contrast Media Mol Imaging 2007 2 189 198 17712869
- [126]Gilad AA McMahon MT Walczak P Winnard PT Jr Raman V van Laarhoven HW Skoglund CM Bulte JW van Zijl PC Artificial reporter gene providing MRI contrast based on proton exchange Nat Biotechnol 2007 25 217 219 17259977
- [127]Liu G Moake MM Gilad AA Jamil M Har-el Y-E Long C Walczak P Zhang J Cardona A DeLiso MA Sgouros G Bulte JW van Zijl PC McMahon MT Multi-color in vivo MR imaging of lymph nodes usng DIACEST liposomes 2010 Proceedings of the 18th Annual Meeting of ISMRM Stockholm, Sweden 454
- [128]McMahon MT Gilad AA DeLiso MA Berman SM Bulte JW van Zijl PC New "multicolor" polypeptide diamagnetic chemical exchange saturation transfer (DIACEST) contrast agents for MRI Magn Reson Med 2008 60 803 812 18816830 PMC2614370
- [129]Aime S Castelli DD Crich SG Gianolio E Terreno E Pushing the sensitivity envelope of lanthanide-based magnetic resonance imaging (MRI) contrast agents for molecular imaging applications Acc Chem Res 2009 42 822 831 19534516
- [130]Dagher AP Aletras A Choyke P Balaban RS Imaging of urea using chemical exchange-dependent saturation transfer at 1.5T J Magn Reson Imaging 2000 12 745 748 11050645
- [131]Zhou J Tryggestad E Wen Z Lal B Zhou T Grossman R Wang S Yan K Fu D-X Ford E Tyler B Blakeley J Laterra J van Zijl PC Differentiation between glioma and radiation necrosis using molecular magnetic resonance imaging of endogenous proteins and peptides Nature Med 2010 In press PMC3058561 21170048
- [132]Dula AN Dortch RD Landman BA Pawate S Lavin PJ Welch EB Gore JC Smith SA Application of CEST imaging to study amide proton transfer (APT) in healthy controls and multiple sclerosis pathology at 7 Tesla 2010 Proceedings of the 18th Annual Meeting of ISMRM Stockholm, Sweden 4311
- [133]Cai K Haris M Singh A Gaddam S Daye D Grasley K II GZ Hariharan H Reddy R Assessment of chemical exchange saturation transfer effects in liver tissue at 7T 2010 Proceedings of the 18th Annual Meeting of ISMRM Stockholm, Sweden 2588
- [134]Al-Radaideh A Mougin O Lim S-Y Tench C Constantinescu C Gowland P Magnetization transfer (MT) and endogenous chemical exchange saturation transfer (CEST) effects in patients with clinically isolated syndrome 2010 Proceedings of the 18th Annual Meeting of ISMRM Stockholm, Sweden 4310
- [135]Gilad AA Winnard PT Jr van Zijl PC Bulte JW Developing MR reporter genes: promises and pitfalls NMR Biomed 2007 20 275 290 17451181
- [136]Chan KW Liu G Arifin DR Bulte JW McMahon MT Remote MR sensing of pH and cell viability using LipoCEST-filled microcapsules 2010 Proceedings of the 18th Annual Meeting of ISMRM Stockholm, Sweden 1889
- [137]Arifin DR Chan KW Liu G Cardona A Jamil M Bulte JW McMahon MT Remote MRI sensing of pH and cell viability using immunoprotective microcapsules crosslinked with polycationic DIACEST peptides 2010 Proceedings of the 18th Annual Meeting of ISMRM Stockholm, Sweden 42
- [138]Pasha A Tircso G Benyo ET Brucher E Sherry AD Synthesis and characterization of DOTA-(amide)(4) derivatives: equilibrium and kinetic behavior of their lanthanide(III) complexes Eur J Inorg Chem 2007 4340 4349 19802361 PMC2756095
- [139]Sherry AD Caravan P Lenkinski RE Primer on gadolinium chemistry J Magn Reson Imaging 2009 30 1240 1248 19938036 PMC2853020
- [140]Li AX Suchy M Li C Poppe C Gati J Meakin S Hudson RH Menon RS Bartha M In vivo detection of a PARACEST agent in mouse brain tumors 2010 Proceedings of the 18th Annual Meeting of ISMRM Stockholm, Sweden 3752
- [141]Jones CK Li AX Suchy M Hudson RH Menon RS Bartha R In vivo detection of PARACEST agents with relaxation correction Magn Reson Med 2010 63 1184 1192 20432289 PMC3427884
- [142]Soesbe TC Rojas-Quijano FA Merritt ME Sherry AD Eu3+-based PARACEST agents with intermediate water exchange rates also act as T2 exchange (T2exch) contrast agents 2010 Proceedings of the 18th Annual Meeting of ISMRM Stockholm, Sweden 39
- [143]Ren J Trokowski R Zhang S Malloy CR Sherry AD Imaging the tissue distribution of glucose in livers using a PARACEST sensor Magn Reson Med 2008 60 1047 1055 18958853 PMC2701623
- [144]Wu Y Zhou Y Ouari O Woods M Zhao P Soesbe TC Kiefer GE Sherry AD Polymeric PARACEST agents for enhancing MRI contrast sensitivity J Am Chem Soc 2008 130 13854 13855 18817395 PMC2723842
- [145]Liepinsh E Otting G Proton exchange rates from amino acid side chains--implications for image contrast Magn Reson Med 1996 35 30 42 8771020
- [146]Viswanathan S Kovacs Z Green KN Ratnakar SJ Sherry AD Alternatives to gadolinium-based metal chelates for magnetic resonance imaging Chemical reviews 2010 110 2960 3018 20397688 PMC2874212
Cited by
Sorted by citations
In vivo imaging of tau pathology using multi-parametric quantitative MRI
Wells, J. A., O'Callaghan, J. M., Holmes, H. E., Powell, N. M., Johnson, R. A., Siow, B., Torrealdea, F., Ismail, O., Walker-Samuel, S., Golay, X., et al. (2015) · NeuroImage · doi:10.1016/j.neuroimage.2015.02.023 · Cited by 93 · see citing works
Bartelle, B. B., Barandov, A., & Jasanoff, A. (2016) · The Journal of Neuroscience · doi:10.1523/jneurosci.4050-15.2016 · Cited by 45 · see citing works
- Authors
- van Zijl, P. C. M., & Yadav, N. N.
- Affiliations
- Kennedy Krieger Institute (US)
- ORCID
- Yadav, Nirbhay N.
- Venue
- Magnetic Resonance in Medicine
- Publisher
- Wiley
- Published
- 2011-02-17
- Volume
- 65
- Issue
- 4
- Pages
- 927-948
- ISSN
- 0740-3194 (ISSN-L)
- 1522-2594
- PMC
- PMC3148076
- OpenAlex
- W1995721991
- Cited by
- 1,155
- References
- 0
- Citations by year
- 2026: 77 · 2025: 97 · 2024: 80 · 2023: 100 · 2022: 95 · 2021: 84 · 2020: 82 · 2019: 95 · 2018: 74 · 2017: 86 · 2016: 69 · 2015: 59 · 2014: 51 · 2013: 69 · 2012: 24
- FWCI
- 38.19
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Magnetic Resonance in Medicine (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- NIH-NCRR (P41-RR15241)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Physical Sciences › Chemistry › Spectroscopy › Advanced NMR Techniques and Applications
- Physical Sciences › Materials Science › Materials Chemistry › Lanthanide and Transition Metal Complexes
- Keywords
- chemical exchange saturation transfer, CEST imaging, magnetization transfer contrast, chemical exchange, slow exchange, inversion, MRI contrast agents, in vivo applications
- MeSH terms
- Algorithms, Animals, Computer Simulation, Contrast Media, Humans, Magnetic Resonance Imaging, Models, Biological, Magnetic Resonance Spectroscopy, Protons
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review