Super‐resolution T1estimation: Quantitative high resolution T1mapping from a set of low resolution T1‐weighted images with different slice orientations
Open accessAbstract
Purpose QuantitativeT1mapping is a magnetic resonance imaging technique that estimates the spin‐lattice relaxation time of tissues. Even thoughT1mapping has a broad range of potential applications, it is not routinely used in clinical practice as accurate and precise high resolutionT1mapping requires infeasibly long acquisition times. Method To improve the trade‐off between the acquisition time, signal‐to‐noise ratio and spatial resolution, we acquire a set of low resolutionT1‐weighted images and directly estimate a high resolutionT1map by means of super‐resolution reconstruction. Results Simulation and in vivo experiments show an increased spatial resolution of theT1map, while preserving a high signal‐to‐noise ratio and short scan time. Moreover, the proposed method outperforms conventional estimation in terms of root‐mean‐square error. Conclusion Super resolutionT1estimation enables resolution enhancement inT1mapping with the use of standard (inversion recovery)T1acquisition sequences. Magn Reson Med 77:1818–1830, 2017. © 2016 International Society for Magnetic Resonance in Medicine
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Super-resolution for MRI can be used to improve the trade-off between resolution, SNR, and acquisition time, generate 7T-like MR images on 3T MRI scanners, or obtain super-resolution T1 maps from a set of low resolution T1 weighted images.
“Super-resolution for MRI have been around for almost 10 years [192, 193] and can be used to improve the trade-off between resolution, SNR, and acquisition time [194], generate 7T-like MR images on 3T MRI scanners [195], or obtain super-resolution T 1 maps from a set of low resolution T 1 weighted images [196].”
From An overview of deep learning in medical imaging focusing on MRI (2019) · cites this paper as [196]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Levary, R. R. (1982) · European Journal of Operational Research · doi:10.1016/0377-2217(82)90093-5 · Cited by 4,170 · see citing works
The rician distribution of noisy mri data
Guðbjartsson, H., & Patz, S. (1995) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910340618 · Cited by 2,498 · see citing works
Multi-planar image formation using NMR spin echoes
Mansfield, P. (1977) · Journal of Physics C Solid State Physics · doi:10.1088/0022-3719/10/3/004 · Cited by 1,809 · see citing works
Handbook of MRI Pulse Sequences
Bernstein, M. A., King, K. F., & Zhou, X. J. (2004) · Elsevier eBooks · doi:10.1016/b978-0-12-092861-3.x5000-6 · Cited by 1,641 · see citing works
Coleman, T. F., & Li, Y. (1994) · Mathematical Programming · doi:10.1007/bf01582221 · Cited by 1,155 · see citing works
Time Saving in Measurement of NMR and EPR Relaxation Times
Look, D. C., & Locker, D. R. (1970) · Review of Scientific Instruments · doi:10.1063/1.1684482 · Cited by 826 · see citing works
Super-Resolution in Medical Imaging
Greenspan, H. (2008) · The Computer Journal · doi:10.1093/comjnl/bxm075 · Cited by 584 · see citing works
Musculoskeletal MRI at 3.0 T: Relaxation Times and Image Contrast
Gold, G. E., Han, E. T., Stainsby, J. A., et al. (2004) · American Journal of Roentgenology · doi:10.2214/ajr.183.2.1830343 · Cited by 518 · see citing works
Signal‐to‐noise measurements in magnitude images from NMR phased arrays
Constantinides, C., Atalar, E., & McVeigh, E. R. (1997) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910380524 · Cited by 512 · see citing works
Deoni, S., Peters, T. M., & Rutt, B. K. (2004) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20314 · Cited by 482 · see citing works
A review of 1H nuclear magnetic resonance relaxation in pathology: Are T1 and T2 diagnostic?
Bottomley, P. A., Hardy, C. J., Argersinger, R. E., et al. (1987) · Medical Physics · doi:10.1118/1.596111 · Cited by 472 · see citing works
(2003) · Wiley eBooks · doi:10.1002/0470869526 · Cited by 347 · see citing works
MRI inter-slice reconstruction using super-resolution
Greenspan, H. K., Oz, G., Kiryati, N., & Peled, S. (2002) · Magnetic Resonance Imaging · doi:10.1016/s0730-725x(02)00511-8 · Cited by 295 · see citing works
On the accuracy of T<sub>1</sub> mapping: Searching for common ground
Stikov, N., Boudreau, M., Levesque, I. R., Tardif, C. L., Barral, J. K., & Pike, G. B. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25135 · Cited by 285 · see citing works
Plenge, E., Poot, D. H. J., Bernsen, M. R., Kotek, G., Houston, J. G., Wielopolski, P. A., van der Weerd, L., Niessen, W. J., & Meijering, E. H. W. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24187 · Cited by 273 · see citing works
Self-citationSuper‐resolution in magnetic resonance imaging: A review
Reeth, É. V., Tham, I. W. K., Tan, C. H., & Poh, C. L. (2012) · Concepts in Magnetic Resonance Part A · doi:10.1002/cmr.a.21249 · Cited by 270 · see citing works
Wright, P., Mougin, O. E., Totman, J. J., Peters, A. M., Brookes, M. J., Coxon, R., Morris, P. E., Clemence, M., Francis, S., Bowtell, R. W., et al. (2008) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1007/s10334-008-0104-8 · Cited by 262 · see citing works
Quantitative Relaxometry of the Brain
Deoni, S. (2010) · Topics in Magnetic Resonance Imaging · doi:10.1097/rmr.0b013e31821e56d8 · Cited by 252 · see citing works
Deoni, S. (2007) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.21130 · Cited by 237 · see citing works
Preibisch, C., & Deichmann, R. (2008) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21776 · Cited by 233 · see citing works
Contrast manipulation and artifact assessment of 2D and 3D RARE sequences
Mulkern, R. V., Wong, S. T. S., Winalski, C. S., et al. (1990) · Magnetic Resonance Imaging · doi:10.1016/0730-725x(90)90132-l · Cited by 232 · see citing works
A comparison of one‐shot and recovery methods in T1 imaging
Crawley, A. P., & Henkelman, R. M. (1988) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910070104 · Cited by 199 · see citing works
Do white matter changes on MRI and CT differentiate vascular dementia from Alzheimer's disease?
Erkinjuntti, T., Ketonen, L., Sulkava, R. O., Sipponen, J. T., Vuorialho, M., & Iivanainen, M. (1987) · Journal of Neurology Neurosurgery & Psychiatry · doi:10.1136/jnnp.50.1.37 · Cited by 196 · see citing works
An Accurate Nuclear Magnetic Resonance Method for Measuring Spin-Lattice Relaxation Times
Hahn, E. L. (1949) · Physical Review · doi:10.1103/physrev.76.145 · Cited by 158 · see citing works
Relationships between diffusion tensor andq‐space MRI†
Basser, P. J. (2002) · Magnetic Resonance in Medicine · doi:10.1002/mrm.10052 · Cited by 149 · see citing works
Scherrer, B., Gholipour, A., & Warfield, S. K. (2012) · Medical Image Analysis · doi:10.1016/j.media.2012.05.003 · Cited by 125 · see citing works
Goodin, D. S. (2006) · Annals of Neurology · doi:10.1002/ana.20832 · Cited by 123 · see citing works
Rapid T1 mapping using multislice echo planar imaging
Clare, S., & Jezzard, P. (2001) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1085 · Cited by 118 · see citing works
REAL-TIME ECHO-PLANAR IMAGING BY NMR
Mansfield, P. (1984) · British Medical Bulletin · doi:10.1093/oxfordjournals.bmb.a071970 · Cited by 117 · see citing works
Analysis of noise in phase contrast MR imaging
Andersen, A. H., & Kirsch, J. E. (1996) · Medical Physics · doi:10.1118/1.597729 · Cited by 96 · see citing works
Kershaw, L. E., & Buckley, D. L. (2006) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21040 · Cited by 93 · see citing works
Practical Implementation and Optimization of One‐shot T1 imaging
Kay, I., & Henkelman, R. M. (1991) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910220249 · Cited by 89 · see citing works
Messroghli, D., Rudolph, A., Abdel‐Aty, H., et al. (2010) · BMC Medical Imaging · doi:10.1186/1471-2342-10-16 · Cited by 88 · see citing works
T 1 Mapping using variable flip angle SPGR data with flip angle correction
Liberman, G., Louzoun, Y., & Bashat, D. B. (2013) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24373 · Cited by 87 · see citing works
Hansen, M. F., Seshamani, S., Kroenke, C. D., Cheng, X. E., Chapman, T., Wilm, J., Rousseau, F., & Studholme, C. (2013) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2013.2284014 · Cited by 87 · see citing works
Effects of magnetization transfer on T 1 contrast in human brain white matter
van Gelderen, P., Jiang, X., & Duyn, J. H. (2015) · NeuroImage · doi:10.1016/j.neuroimage.2015.12.032 · Cited by 85 · see citing works
Badve, C. A., Yu, A. C., Rogers, M., Ma, D., Liu, Y., SCHLUCHTER, M. D., Sunshine, J. L., Griswold, M. A., & Gulani, V. (2015) · Tomography · doi:10.18383/j.tom.2015.00166 · Cited by 82 · see citing works
A Direct Method of Measuring Nuclear Spin-Lattice Relaxation Times
Drain, L. E. (1949) · Proceedings of the Physical Society Section A · doi:10.1088/0370-1298/62/5/306 · Cited by 79 · see citing works
A single-scan fourier transform method for measuring spin-lattice relaxation times
Kaptein, R., Dijkstra, K., & Tarr, C. E. (1976) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(76)90039-1 · Cited by 78 · see citing works
Zhu, D. C. D., & Penn, R. D. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20602 · Cited by 77 · see citing works
Data distributions in magnetic resonance images: A review
den Dekker, A. J., & Sijbers, J. (2014) · Physica Medica · doi:10.1016/j.ejmp.2014.05.002 · Cited by 76 · see citing works
Self-citationRARE/turbo spin echo imaging with simultaneous multislice Wave-CAIPI
Gagoski, B. A., Bilgiç, B., Eichner, C., Bhat, H., Grant, P. E., Wald, L. L., & Setsompop, K. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25615 · Cited by 73 · see citing works
Cheng, H. M. (2007) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.20898 · Cited by 73 · see citing works
Neely, R. K., Vandenberg, W., & Dedecker, P. (2016) · CRC Press eBooks · doi:10.1201/b18007-25 · Cited by 73 · see citing works
Magnetic field dependence of the distribution of NMR relaxation times in the living human brain
Oros‐Peusquens, A., Laurila, M., & Shah, N. J. (2008) · Magnetic Resonance Materials in Physics Biology and Medicine · doi:10.1007/s10334-008-0107-5 · Cited by 68 · see citing works
Naruse, S., Horikawa, Y., Tanaka, C., et al. (1986) · Magnetic Resonance Imaging · doi:10.1016/0730-725x(86)91039-8 · Cited by 65 · see citing works
Steenkiste, G. V., Jeurissen, B., Veraart, J., den Dekker, A. J., Parizel, P. M., Poot, D. H. J., & Sijbers, J. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25597 · Cited by 58 · see citing works
Self-citationSuper‐resolution for multislice diffusion tensor imaging
Poot, D. H. J., Jeurissen, B., Bastiaensen, Y., Veraart, J., Hecke, W. V., Parizel, P. M., & Sijbers, J. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24233 · Cited by 58 · see citing works
Self-citationPeeters, F., Annet, L., Hermoye, L., et al. (2004) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20032 · Cited by 54 · see citing works
General and Efficient Super-Resolution Method for Multi-slice MRI
Poot, D. H. J., Meir, V. V., & Sijbers, J. (2010) · Lecture notes in computer science · doi:10.1007/978-3-642-15705-9_75 · Cited by 51 · see citing works
Self-citationModel-based registration to correct for motion between acquisitions in diffusion MR imaging
Bai, Y., & Alexander, D. C. (2008) · IEEE International Symposium on Biomedical Imaging: From Nano to Macro (ISBI) · doi:10.1109/isbi.2008.4541154 · Cited by 43 · see citing works
Magnetic resonance imaging in epilepsy: a controlled study
Conlon, P., Trimble, M., Rogers, D., et al. (1988) · Epilepsy Research · doi:10.1016/0920-1211(88)90008-3 · Cited by 40 · see citing works
Estimating T1 from multichannel variable flip angle SPGR sequences
Trzasko, J. D., Mostardi, P. M., Riederer, S. J., et al. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24401 · Cited by 27 · see citing works
Kurki, T. J. I., & Komu, M. E. S. (1995) · Magnetic Resonance Imaging · doi:10.1016/0730-725x(94)00126-n · Cited by 23 · see citing works
Quantitative MR in multi‐center clinical trials
Ashton, E. A. (2010) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.22022 · Cited by 22 · see citing works
Relaxation times in relation to grade of malignancy and tissue necrosis in astrocytic gliomas
Englund, E. M., Brun, A. E., Györffy-Wagner, Z., et al. (1986) · Magnetic Resonance Imaging · doi:10.1016/0730-725x(86)90051-2 · Cited by 21 · see citing works
TOWERS: T‐One with Enhanced Robustness and Speed
Eldeniz, C., Finsterbusch, J., Lin, W., et al. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25864 · Cited by 12 · see citing works
Ramos‐Llordén, G., den Dekker, A. J., Steenkiste, G. V., et al. (2015) · Proceedings - International Conference on Image Processing · doi:10.1109/icip.2015.7351386 · Cited by 5 · see citing works
Wansapura (1999) · doi:10.1002/(sici)1522-2586(199904)9:4
Flagged first, then by citations
Research works printed but not linked: 9
- Truyen L, van Waesberghe JHTM, van Walderveen MAA, van Oosten BW, Polman CH, Hommes OR, Ade `r HJA, Barkhof F. Accumulation of hypointense lesions ("black holes") on T1 spin-echo MRI correlates with disease progression in multiple sclerosis. Neurology 1996;47:1469-1476.
- Stikov N, Boudreau M, Levesque IR, Tardif CL, Barral JK, Pike GB. On the accuracy of T1 mapping: searching for common ground. Magn Reson Med 2015;73:514-522.
- Liu Y, Buck JR, Zheng S, Ikonomidou VN. Optimizing and comparing the efficiencies of relaxometry sequences in quantitative T1 and T2 imaging. In Proc Intl Soc Mag Reson Med, Toronto, Ontario, Canada; Vol. 23, 2015. p. 1689.
- Mintzopoulos D, Inati S. The effects of magnetic field inhomogeneity on FLASH-based T1 measurements. In Proc Intl Soc Mag Reson Med, Seattle, Washington, USA; Vol. 14, 2006. p. 923.
- Fogtmann M, Seshamani S, Kim K, Chapman T, Studholme C. A uni- fied approach for motion estimation and super resolution reconstruc- tion from structural magnetic resonance imaging on moving subjects. In MICCAI Workshop on Perinatal and Paediatric Imaging, Nice, France; 2012. pp. 9-26.
- Van Steenkiste G, Poot DHJ, Jeurissen B, den Dekker AJ, Sijbers J. Super-resolution T1 mapping: a simulation study. In Proc Intl Soc Mag Reson Med, 2015. p. 1679.
- Scheffler K. Superresolution in MRI? Magn Reson Med 2002;47:392- 397.
- Cocosco CA, Kollokian V, Kwan RKS, Evans AC. BrainWeb: online interface to a 3D MRI simulated brain database. Neuroimage 1997;5: S425.
- Gold GE, Suh B, Sawyer-Glover A, Beaulieu C. Musculoskeletal MRI at 3.0 T: initial clinical experience. Am J Roentgenol 2004;183:343-351.
Cited by
Sorted by citations
An overview of deep learning in medical imaging focusing on MRI
Lundervold, A. S., & Lundervold, A. (2019) · Zeitschrift für Medizinische Physik · doi:10.1016/j.zemedi.2018.11.002 · Cited by 2,058 · see citing works
- Authors
- Van Steenkiste, G., Poot, D. H. J., Jeurissen, B., den Dekker, A. J., Vanhevel, F., Parizel, P. M., & Sijbers, J.
- Affiliations
- iMinds‐Vision Lab Department of Physics, University of Antwerp Antwerp Belgium
- Erasmus MC (NL)
- University of Antwerp (BE)
- Venue
- Magnetic Resonance in Medicine
- Publisher
- Wiley
- Published
- 2016-07
- Volume
- 77
- Issue
- 5
- Pages
- 1818-1830
- ISSN
- 0740-3194 (ISSN-L)
- 1522-2594
- OpenAlex
- W2462044004
- Cited by
- 20
- References
- 59
- Citations by year
- 2026: 1 · 2024: 3 · 2022: 6 · 2021: 1 · 2020: 2 · 2019: 3 · 2018: 3 · 2016: 1
- FWCI
- 2.02
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Magnetic Resonance in Medicine (publishedVersion)
- PubMed (publishedVersion)
- EUR Research Repository (Erasmus University Rotterdam) (publishedVersion)
- EUR Research Repository (Erasmus University Rotterdam) (submittedVersion)
- Funders
- Fonds Wetenschappelijk Onderzoek
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Advanced Image Processing Techniques
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Keywords
- quantitative T1 mapping, super-resolution reconstruction, T1 relaxation time, slice orientation, spatial resolution enhancement, acquisition time reduction, signal-to-noise ratio, inversion recovery
- MeSH terms
- Adult, Algorithms, Brain, Brain Mapping, Computer Simulation, Humans, Image Interpretation, Computer-Assisted, Image Processing, Computer-Assisted, Magnetic Resonance Imaging, Male, Motion, Reproducibility of Results, Models, Statistical, Anisotropy, Phantoms, Imaging, Imaging, Three-Dimensional, Signal-To-Noise Ratio
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review