Simultaneous determination of parabens, triclosan and triclocarban in water by liquid chromatography/electrospray ionisation tandem mass spectrometry
Abstract
A method for the determination of several household biocides in water by liquid chromatography/electrospray ionisation tandem mass spectrometry (LC/ESI-MS/MS) is presented. It permits the simultaneous determination of triclosan (TCS), triclocarban (TCC) and seven parabens, including the distinction between branched and linear isomers of propyl (i-PrP and n-PrP) and butyl parabens (i-BuP and n-BuP). Prior to LC/MS/MS, analytes are preconcentrated by solid-phase extraction (SPE) on Oasis HLB (60 mg) cartridges at natural sample pH and subsequently eluted with 4 mL of methanol. This simple SPE procedure provides extraction recoveries above 85% except for raw wastewater, where it falls to 65% for TCC. The performance of the method was tested with two triple-quadrupole LC/MS instruments from a low/mid and mid/high market range: a Varian 1200L and an API-4000. The latter system provided between 3 and 80 times lower limits of quantification (LOQs) than the first one, in the 0.08-0.44 ng/L range for surface water. Moreover, a comparison of matrix effects on both instruments showed a very different behaviour, particularly in the case of parabens. For these compounds signal suppression was observed in the 1200L instrument and signal enhancement with the 4000 instrument. As a result, different calibration approaches were chosen for them and this pointed to the need of matrix effect re-evaluation in method transfer between different LC/MS systems. The application of the method to real sa…
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Recoveries are comparable with those obtained using conventional SPE/LC-MS.
“Recoveries are comparable with those obtained using conventional SPE/LC-MS [55].”
From Automated dispersive liquid-liquid microextraction based on the solidification of the organic phase (2018) · cites this paper as [55]
Cites
Concentrations of parabens in human breast tumours
Darbre, P. D., Aljarrah, A., Miller, W. R., Coldham, N. G., Sauer, M. J., & Pope, G. S. (2004) · Journal of Applied Toxicology · doi:10.1002/jat.958 · Cited by 805 · see citing works
Some Alkyl Hydroxy Benzoate Preservatives (Parabens) Are Estrogenic
Routledge, E. J., Parker, J., Odum, J., Ashby, J., & Sumpter, J. P. (1998) · Toxicology and Applied Pharmacology · doi:10.1006/taap.1998.8544 · Cited by 721 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Wastewater-based epidemiology biomarkers: Past, present and future
Choi, P. M., Tscharke, B. J., Donner, E., O'Brien, J. W., Grant, S. C., Kaserzon, S. L., Mackie, R., O'Malley, E., Crosbie, N. D., Thomas, K. V., et al. (2018) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2018.06.004 · Cited by 486 · see citing works
Parabens as chemicals of emerging concern in the environment and humans: a review
Wei, F., Mortimer, M., Cheng, H., Sang, N., & Guo, L.-H. (2021) · Science of The Total Environment · doi:10.1016/j.scitotenv.2021.146150 · Cited by 321 · see citing works
Quantity and quality characteristics of greywater: A review
Shaikh, I. N., & Ahammed, M. M. (2020) · Journal of Environmental Management · doi:10.1016/j.jenvman.2020.110266 · Cited by 175 · see citing works
Ocaña-González, J. A., Villar-Navarro, M., Ramos-Payán, M., Fernández-Torres, R., & Bello-López, M. A. (2015) · Analytica Chimica Acta · doi:10.1016/j.aca.2014.07.002 · Cited by 129 · see citing works
Automated dispersive liquid-liquid microextraction based on the solidification of the organic phase
Medina, D. A. V., Santos-Neto, Á. J., Cerdà, V., & Maya, F. (2018) · Talanta · doi:10.1016/j.talanta.2018.06.081 · Cited by 55 · see citing works
- Authors
- González‐Mariño, I., Quintana, J. B., Rodríguez, I., & Cela, R.
- Affiliations
- Universidade de Santiago de Compostela (ES)
- Corresponding author
- Quintana, José Benito
- Venue
- Rapid Communications in Mass Spectrometry
- Publisher
- Wiley
- Published
- 2009-05-12
- Volume
- 23
- Issue
- 12
- Pages
- 1756-1766
- ISSN
- 0951-4198 (ISSN-L)
- 1097-0231
- OpenAlex
- W2149987399
- Cited by
- 160
- References
- 2
- Citations by year
- 2026: 1 · 2025: 3 · 2024: 9 · 2023: 4 · 2022: 7 · 2021: 11 · 2020: 6 · 2019: 15 · 2018: 21 · 2017: 11 · 2016: 14 · 2015: 10 · 2014: 20 · 2013: 5 · 2012: 6
- FWCI
- 12.18
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Rapid Communications in Mass Spectrometry (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Spanish Ministry of Science and Innovation (CTQ2006-03334)
- FEDER
- Topics
- Life Sciences › Agricultural and Biological Sciences › Food Science › Pesticide Residue Analysis and Safety
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical chemistry methods development
- Physical Sciences › Environmental Science › Health, Toxicology and Mutagenesis › Effects and risks of endocrine disrupting chemicals
- Keywords
- parabens, triclosan, triclocarban, water pollution, solid-phase extraction, matrix effects, isomer identification, propylparaben, butylparaben, methylparaben, raw sewage, surface water, method transfer
- MeSH terms
- Carbanilides, Chromatography, Liquid, Parabens, Triclosan, Water Pollutants, Chemical, Spectrometry, Mass, Electrospray Ionization, Tandem Mass Spectrometry
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review