Measurement of multiply substituted isotopologues ('clumped isotopes') of CO 2 using a 5 kV compact isotope ratio mass spectrometer: Performance, reference frame, and carbonate paleothermometry

Rosenheim, Brad E., Tang, Jianwu, Fernandez, Alvaro · 2013 · Rapid Communications in Mass Spectrometry
Cited by 21

The DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.

Abstract

RATIONALE: The burgeoning field of 'clumped isotope' paleothermometry, which has broad applications in geosciences, depends almost entirely on measurements made on two types of mass spectrometer with the same ion source design. Demonstration that these measurements can be carried out on a retrofitted mass spectrometer with a different type of ion source is important to the growing community of geoscientists considering employing this technique. METHODS: We summarize the sample preparation techniques, mass spectrometry, and data processing involved in establishing an absolute reference frame for the measurement of Δ 47 in CO 2 from carbonate minerals using a prototype 5 kV mass spectrometer. The prototype differs from previously marketed mass spectrometers only in the presence of six extra Faraday cups designed specifically to measure ion beams of masses 44–49. We employ pressure baseline corrections on the small mass 47 signal. RESULTS: The system was fully capable of establishing absolute reference frames for clumped isotope carbonate paleothermometry, despite alternative use of the system for other types of measurements (carbonate δ 18 O and δ 13 C values, water δ 18 O and δ 2 H values via equilibration, and δ 13 C and δ 15 N values of sedimentary organic material) in different reference frames. Repeated measurements of CO 2 from carbonate standards yielded external precisions comparable with those from other laboratories and adequate for paleothermometry to a precision of near or below ±5 °C. The accuracy was demonstrated by reasonable temperatures being obtained for different types of repeatedly measured carbonates. Measurements of the pressure baseline with every sample increased analysis times by 20 min, but decreased the stabilization time of the reference frame by days to weeks, while improving precision. CONCLUSIONS: Simple modifications to existing instrumentation will allow more laboratories to perform clumped isotope measurements. The prototype mass spectrometer used herein provides precision comparable with all published data. We show that, where negatively charged species are associated with large ion beams, making pressure baseline corrections with every measurement is increasingly important no matter what instrument is used. Copyright © 2013 John Wiley & Sons, Ltd.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1002/rcm.6634

Cites

10 of 10 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 10 references

Cited by

See all 21

Metadata

Authors
Rosenheim, B. E., Tang, J., & Fernandez, A.
Venue
Rapid Communications in Mass Spectrometry
Publisher
Wiley
Published
2013-07-15
Volume
27
Issue
16
Pages
1847-1857
ISSN
  • 0951-4198 (ISSN-L)
  • 1097-0231
OpenAlex
W2100981658
Cited by
21
References
10
FWCI
2.60
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Measurement of multiply substituted isotopologues ('clumped isotopes') of CO 2 using a 5 kV compact isotope ratio mass spectrometer: Performance, reference frame, and carbonate paleothermometry. https://app.cochat.ai/papers/10.1002/rcm.6634
The paper
Rosenheim, B. E., Tang, J., & Fernandez, A. (2013). Measurement of multiply substituted isotopologues ('clumped isotopes') of CO 2 using a 5 kV compact isotope ratio mass spectrometer: Performance, reference frame, and carbonate paleothermometry. Rapid Communications in Mass Spectrometry. https://doi.org/10.1002/rcm.6634

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.