A candidate reference measurement procedure for quantifying serum concentrations of 25-hydroxyvitamin D3 and 25-hydroxyvitamin D2 using isotope-dilution liquid chromatography-tandem mass spectrometry
Open accessThe DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
AbstractThe inaccuracy of routine serum 25-hydroxyvitamin D measurements hampers the interpretation of data in patient care and public health research. We developed and validated a candidate reference measurement procedure (RMP) for highly accurate quantitation of two clinically important 25-hydroxyvitamin D metabolites in serum, 25-hydroxyvitamin D2 [25(OH)D2] and 25-hydroxyvitamin D3 [25(OH)D3]. The two compounds of interest together with spiked deuterium-labeled internal standards [d3-25(OH)D2 and d6-25(OH)D3] were extracted from serum via liquid-liquid extraction. The featured isotope-dilution LC-MS/MS method used reversed-phase chromatography and atmospheric pressure chemical ionization in positive ion mode. A pentafluorophenylpropyl-packed UHPLC column together with isocratic elution allowed for complete baseline resolution of 25(OH)D2 and 25(OH)D3 from their structural C-3 isomers within 12 min. We evaluated method trueness, precision, potential interferences, matrix effects, limits of quantitation, and measurement uncertainty. Calibration materials were, or were traceable to, NIST Standard Reference Materials 2972. Within-day and total imprecision (CV) averaged 1.9 and 2.0 % for 25(OH)D3, respectively, and 2.4 and 3.5 % for 25(OH)D2, respectively. Mean trueness was 100.3 % for 25(OH)D3 and 25(OH)D2. The limits of quantitation/limits of detection were 4.61/1.38 nmol/L for 25(OH)D3 and 1.46/0.13 nmol/L for 25(OH)D2. When we compared our RMP results to an established RMP using 40 serum samples, we found a nonsignificant mean bias of 0.2 % for total 25(OH)D. This candidate RMP for 25(OH)D metabolites meets predefined method performance specifications (≤5 % total CV and ≤1.7 % bias) and provides sufficient sample throughput to meet the needs of the Centers for Disease Control and Prevention Vitamin D Standardization Certification Program. Graphical abstractBias assessment using NIST standard reference materials. Legend CDC mean mass fractions (ng/g) ± U95 (6 replicates per mean). NIST-certified mass fractions (ng/g) ± U95 from the Certificates of Analysis
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Standardization of testosterone measurements in humans
Vesper, H. W., & Botelho, J. C. (2010) · The Journal of Steroid Biochemistry and Molecular Biology · doi:10.1016/j.jsbmb.2010.03.032 · Cited by 113 · cited 1×
“This program uses approaches and procedures successfully employed in CDC’s Hormone Standardization Program [16, 17].”
Self-citation
Cited by
Vitamin D status in the United States, 2011–2014
Herrick, K. A., Storandt, R. J., Afful, J., Pfeiffer, C. M., Schleicher, R. L., Gahche, J. J., & Potischman, N. (2019) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/nqz037 · Cited by 597
Self-citationSempos, C. T., Heijboer, A. C., Bikle, D. D., Bollerslev, J., Bouillon, R., Brannon, P. M., DeLuca, H. F., Jones, G., Munns, C. F., Bilezikian, J. P., et al. (2018) · British Journal of Clinical Pharmacology · doi:10.1111/bcp.13652 · Cited by 331
An Update on Vitamin D Metabolism
Saponaro, F., Saba, A., & Zucchi, R. (2020) · International Journal of Molecular Sciences · doi:10.3390/ijms21186573 · Cited by 308
Dirks, N. F., Vesper, H. W., van Herwaarden, A. E., van den Ouweland, J. M. W., Kema, I. P., Krabbe, J. G., & Heijboer, A. C. (2016) · Clinica Chimica Acta · doi:10.1016/j.cca.2016.08.016 · Cited by 40
Self-citationAnalytical Performance Specifications for 25-Hydroxyvitamin D Examinations
Cavalier, E., Fraser, C., Bhattoa, H., Heijboer, A., Makris, K., Ulmer, C., Vesper, H., Vasikaran, S., Lukas, P., Delanaye, P., et al. (2021) · Nutrients · doi:10.3390/nu13020431 · Cited by 23
“The CDC Vitamin D Standardization-Certification Program (VDSCP) determines the calibration bias and imprecision by evaluating the data obtained from 40 blinded singleunit, fresh-frozen serum samples measured in batches of 10 samples over four consecutive quarters.”
MethodInfluentialSelf-citation
Metadata
- Authors
- Mineva, E. M., Schleicher, R. L., Chaudhary-Webb, M., Maw, K. L., Botelho, J. C., Vesper, H. W., & Pfeiffer, C. M.
- Venue
- Analytical and Bioanalytical Chemistry
- Published
- 2015-05-13
- Volume
- 407
- Issue
- 19
- Pages
- 5615-5624
- ISSN
- 1618-2642 (ISSN-L)
- 1618-2650
- OpenAlex
- W2118839330
- Cited by
- 103
- References
- 1
- FWCI
- 6.42
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A candidate reference measurement procedure for quantifying serum concentrations of 25-hydroxyvitamin D3 and 25-hydroxyvitamin D2 using isotope-dilution liquid chromatography-tandem mass spectrometry. https://app.cochat.ai/papers/10.1007/s00216-015-8733-z
- The paper
- Mineva, E. M., Schleicher, R. L., Chaudhary-Webb, M., Maw, K. L., Botelho, J. C., Vesper, H. W., & Pfeiffer, C. M. (2015). A candidate reference measurement procedure for quantifying serum concentrations of 25-hydroxyvitamin D3 and 25-hydroxyvitamin D2 using isotope-dilution liquid chromatography-tandem mass spectrometry. Analytical and Bioanalytical Chemistry. https://doi.org/10.1007/s00216-015-8733-z
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.