Cited by 38Thyroid Disorders and TreatmentsAnalytical Chemistry and ChromatographyAnalytical Methods in Pharmaceuticals

A direct comparison of liquid chromatography-mass spectrometry with clinical routine testing immunoassay methods for the detection and quantification of thyroid hormones in blood serum

Open access
Bowerbank, Samantha L., Carlin, Michelle G., Dean, John R.·2019·Analytical and Bioanalytical Chemistry

Abstract

A new and improved method was developed for the determination and quantification of four "free" thyroid hormones (i.e. 3,5-diiodothyronine (T2), 3,3',5-triiodothyronine (T3), 3,3',5'-triiodothyrone (rT3) and 3,5,3',5'-tetraiodothyronine (T4)) in human serum by low- and high-resolution liquid chromatography-mass spectrometry (LC-MS). Several sample preparation strategies were investigated to obtain matrix-independent results. These strategies included solid phase extraction and matrix dilution. The developed analytical methods were then directly compared, in a blind study using patient-derived human blood serum samples, to the current clinical routine testing methods, i.e. electrochemiluminescence immunoassay and enzyme-linked immunosorbent assay. Chromatographic separation was achieved on a pentafluorophenyl (F5) column with an isocratic method of 30% aqueous phase, 70% organic phase where mobile phase A is 0.1% formic acid in water (pH 4) and mobile phase B is 0.1% formic acid in methanol (pH 4) (v/v). The high-resolution LC-MS was able to give a significant improvement in sensitivity with limits of quantification of 0.002 to 0.008 pmol/L for all four "free" thyroid hormones, as well as reduced sample preparation, making this the preferred method. However, the increase in capital cost may be beyond the capabilities of some laboratories. The LC-MS methods allow for the analysis of "free" thyroid hormones to be carried out in a significantly reduced analysis time. Clinical sample analysis showed that there was no statistical difference between the results obtained by ECLIA/ELISA and both LC-MS methods. Graphical abstract.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1007/s00216-019-01724-2

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Several laboratories have developed procedures to accurately determine total T4 and total T3 in human plasma or serum and largely agree on concentration ranges, but this consensus has not been achieved for most other TH metabolites.

    “While meanwhile several laboratories developed procedures to accurately determine total T4 and total T3 in human plasma or serum and largely agree on concentrations ranges (25, 67–73), this consensus has not been achieved yet for most of the other TH metabolites, which occur at much lower concentrations.”

    From 3,5-T2—A Janus-Faced Thyroid Hormone Metabolite Exerts Both Canonical T3-Mimetic Endocrine and Intracrine Hepatic Action (2020) · cites this paper as [69]

    The Abstract describes a procedure for measuring T3 and T4 in human serum, and Results and discussion reports strong agreement between that method and routine immunoassays. But the paper measures *free*, not total, hormones and does not establish that multiple laboratories agree on concentration ranges for total T3 and T4. The citation is attached to this first clause, not the subsequent statement about other metabolites. (LLM, conf=0.82)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 27

  1. [1]Sinha RA Singh BK Yen PM Thyroid hormone regulation of hepatic lipid and carbohydrate metabolism Trends Endocrinol Metab 2014 25 10 538 545 10.1016/j.tem.2014.07.001 25127738 · DOI
  2. [2]Goodman HM Basic medical endocrinology 2009 4 Amsterdam Academic Press
  3. [3]Gu J Soldin OP Soldin SJ Simultaneous quantification of free triiodothyronine and free thyroxine by isotope dilution tandem mass spectrometry Clin Biochem 2007 40 18 1386 1391 10.1016/j.clinbiochem.2007.08.007 17936741 · DOI
  4. [4]Kapelari K Kirchlechner C Hogler W Schweitzer K Virgolini I Moncayo R Pediatric reference intervals for thyroid hormone levels from birth to adulthood: a retrospective study BMC Endocr Disord 2008 8 15 10.1186/1472-6823-8-15 19036169 PMC2645400 · DOI · Paper page
  5. [5]Lem AJ de Rijke YB van Toor H de Ridder MA Visser TJ Hokken-Koelega AC Serum thyroid hormone levels in healthy children from birth to adulthood and in short children born small for gestational age J Clin Endocrinol Metab 2012 97 9 3170 3178 10.1210/jc.2012-1759 22736771 · DOI
  6. [10]Chaker L, Bianco AC, Jonklaas J, Peeters RP. Hypothyroidism. Lancet. 2017. PMC6619426 28336049
  7. [11]Dunn D Turner C Hypothyroidism in women Nurs Womens Health 2016 20 1 93 98 10.1016/j.nwh.2015.12.002 26902444 · DOI
  8. [12]Gilbert J Hypothyroidism Medicine. 2017 45 8 506 509 10.1016/j.mpmed.2017.05.009 · DOI
  9. [14]Sanchez-Carbayo M Mauri M Alfayate R Miralles C Soria F Analytical and clinical evaluation of TSH and thyroid hormones by electrochemiluminescent immunoassays Clin Biochem 1999 32 6 395 403 10.1016/S0009-9120(99)00032-6 10667473 · DOI
  10. [15]Soldin OP Hilakivi-Clarke L Weiderpass E Soldin SJ Trimester-specific reference intervals for thyroxine and triiodothyronine in pregnancy in iodine-sufficient women using isotope dilution tandem mass spectrometry and immunoassays Clin Chim Acta 2004 349 1–2 181 189 10.1016/j.cccn.2004.06.021 15469872 PMC3625638 · DOI
  11. [16]Kazerouni F Amirrasouli H Performance characteristics of three automated immunoassays for thyroid hormones Capsian J Intern Med 2012 3 2 400 404 PMC3861902 24358433
  12. [17]Welsh KJ Soldin SJ How reliable are free thyroid and total T3 hormone assays? Eur J Endocrinol 2016 175 6 255 263 10.1530/EJE-16-0193 27737898 PMC5113291 · DOI · Paper page
  13. [22]Jonklaas J Sathasivam A Wang H Gu J Burman KD Soldin SJ Total and free thyroxine and triiodothyronine: measurement discrepancies, particularly in inpatients Clin Biochem 2014 47 13–14 1272 1278 10.1016/j.clinbiochem.2014.06.007 24936679 PMC4187214 · DOI · Paper page
  14. [23]Wu AH French D Implementation of liquid chromatography/mass spectrometry into the clinical laboratory Clin Chim Acta 2013 420 4 10 10.1016/j.cca.2012.10.026 23085380 · DOI
  15. [24]Yong S Chen Y Lee TK Lee HK Determination of total thyroxine in human serum by hollow fiber liquid-phase microextraction and liquid chromatography-tandem mass spectrometry Talanta. 2014 126 163 169 10.1016/j.talanta.2014.03.058 24881548 · DOI
  16. [25]Zhang Y Conrad AH Conrad GW Detection and quantification of 3,5,3′-triiodothyronine and 3,3′,5′-triiodothyronine by electrospray ionization tandem mass spectrometry J Am Soc Mass Spectrom 2005 16 11 1781 1786 10.1016/j.jasms.2005.06.017 16182556 · DOI
  17. [26]Couldwell AM Thomas MC Mitchell TW Hulbert AJ Blanksby SJ Tandem mass spectrometry of deprotonated iodothyronines Rapid Commun Mass Sp 2005 19 16 2295 2304 10.1002/rcm.2061 16021613 · DOI
  18. [27]Grebenstein N Frank J Rapid baseline-separation of all eight tocopherols and tocotrienols by reversed-phase liquid-chromatography with a solid-core pentafluorophenyl column and their sensitive quantification in plasma and liver J Chromatogr A 2012 1243 39 46 10.1016/j.chroma.2012.04.042 22560347 · DOI
  19. [28]Lemasson E Bertin S Hennig P Lesellier E West C Impurity profiling of drug candidates: analytical strategies using reversed-phase and mixed-mode high-performance liquid chromatography methods J Chromatogr A 2018 1535 101 113 10.1016/j.chroma.2018.01.014 29329884 · DOI
  20. [29]Smidova B Satinsky D Dostalova K Solich P The pentafluorophenyl stationary phase shows a unique separation efficiency for performing fast chromatography determination of highbush blueberry anthocyanins Talanta. 2017 166 249 254 10.1016/j.talanta.2017.01.061 28213230 · DOI
  21. [30]Fibigr J Satinsky D Solich P A new approach to the rapid separation of isomeric compounds in a Silybum marianum extract using UHPLC core-shell column with F5 stationary phase J Pharmaceut Biomed 2017 134 203 213 10.1016/j.jpba.2016.11.042 27915198 · DOI
  22. [31]Yang M Fazio S Munch D Drumm P Impact of methanol and acetonitrile on separations based on pi-pi interactions with a reversed-phase phenyl column J Chromatogr A 2005 1097 1–2 124 129 10.1016/j.chroma.2005.08.028 16298191 · DOI
  23. [32]Welsh KJ Stolze BR Yu X Podsiadlo TR Kim LS Soldin SJ Assessment of thyroid function in intensive care unit patients by liquid chromatography tandem mass spectrometry methods Clin Biochem 2017 50 6 318 322 10.1016/j.clinbiochem.2016.11.022 27890823 PMC5429196 · DOI · Paper page
  24. [33]Gounden V Jonklaas J Soldin SJ A pilot study: subclinical hypothyroidism and free thyroid hormone measurement by immunoassay and mass spectrometry Clin Chim Acta 2014 430 121 124 10.1016/j.cca.2013.12.034 24389098 PMC3971640 · DOI
  25. [34]Masika LS Zhao Z Soldin SJ Is measurement of TT3 by immunoassay reliable at low concentrations? A comparison of the Roche Cobas 6000 vs. LC-MSMS Clin Biochem 2016 49 12 846 849 10.1016/j.clinbiochem.2016.02.004 26879528 PMC4972640 · DOI · Paper page
  26. [35]Tanoue R Kume I Yamamoto Y Takaguchi K Nomiyama K Tanabe S Determination of free thyroid hormones in animal serum/plasma using ultrafiltration in combination with ultra-fast liquid chromatography-tandem mass spectrometry J Chromatogr A 2018 1539 30 40 10.1016/j.chroma.2018.01.044 29395158 · DOI
  27. [36]European Medicines Agency guideline bioanalytical method validation. 2011. 22533559

Cited by

Sorted by citations

Authors
Bowerbank, S. L., Carlin, M. G., & Dean, J. R.
Affiliations
Northumbria University (GB)
Corresponding author
Dean, John R.
Venue
Analytical and Bioanalytical Chemistry
Published
2019-05
Volume
411
Issue
13
Pages
2839-2853
ISSN
  • 1618-2642 (ISSN-L)
  • 1618-2650
OpenAlex
W2944288092
Cited by
38
References
1
Citations by year
2026: 5 · 2024: 7 · 2023: 6 · 2022: 3 · 2021: 9 · 2020: 8
FWCI
4.06
Type
journal article
Language
en
Open access
Locations
  • Analytical and Bioanalytical Chemistry (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • Northumbria Research Link (Northumbria University) (acceptedVersion)
Funders
Northumbria University
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Thyroid Disorders and Treatments
  • Physical Sciences › Chemistry › Spectroscopy › Analytical Chemistry and Chromatography
  • Physical Sciences › Chemistry › Analytical Chemistry › Analytical Methods in Pharmaceuticals
Keywords
free thyroid hormones, liquid chromatography-mass spectrometry, immunoassay comparison, electrochemiluminescence immunoassay, enzyme-linked immunosorbent assay, 3,5-diiodothyronine, triiodothyronine, reverse T3, thyroxine, solid-phase extraction, method validation
MeSH terms
Animals, Chromatography, High Pressure Liquid, Horses, Humans, Immunoassay, Mass Spectrometry, Thyroid Hormones, Limit of Detection

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review