Comparison of the point-of-care blood gas analyzer versus the laboratory auto-analyzer for the measurement of electrolytes

Open access
Jain, Anunaya, Subhan, Imron, Joshi, Mahesh·2009·International Journal of Emergency Medicine

Abstract

BACKGROUND: Electrolyte values are measured both by arterial blood gas (ABG) analyzers and central laboratory auto-analyzers (AA), but a significant time gap exists between the availability of both these results, with the ABG giving faster results than the AA. The authors hypothesized that there is no difference between the results obtained after measurement of electrolytes by the blood gas and auto-analyzers. METHODS: After approval by the ethics committee, an observational cohort study was conducted in which 200 paired venous and arterial samples from patients admitted to the Medical Intensive Care Unit (ICU) of Apollo Hospital, Hyderabad, India, were analyzed for electrolytes on the ABG machine and the AA. Analyses were done on the ABL555 blood gas analyzer and the Dade Dimension RxL Max, both located in the central laboratory. Statistical analyses were performed using paired Student's t test. RESULTS: A total of 200 paired samples were analyzed. The mean ABG sodium value was 131.28 (SD 7.33), and the mean AA sodium value was 136.45 (SD 6.50) (p < 0.001). The mean ABG potassium value was 3.74 (SD 1.92), and the mean AA potassium value was 3.896 (SD 1.848) (p = 0.2679). CONCLUSION: Based on the above analysis, the authors found no significant difference between the potassium values measured by the blood gas machine and the auto-analyzer. However, the difference between the measured sodium was found to be significant. We therefore conclude that critical decisions can be made by trusting the potassium values obtained from the arterial blood gas analysis.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1007/s12245-009-0091-1

Cites

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Whole-blood POCT analyzers tend to report slightly elevated potassium levels compared to serum-based laboratory systems due to differences in sample handling and electrode sensitivity.

    “This minor overestimation bias is consistent with earlier work by Chhapola et al. [3] and Jain et al. [4], who observed that whole-blood POCT analyzers tend to report slightly elevated K⁺ levels compared to serum-based lab systems due to differences in sample handling and electrode sensitivity.”

    From Reliability of Point-of-Care Venous Blood Gas Testing for Potassium and Sodium Compared to Central Laboratory Results in Emergency Patients: A Clinical Audit (2025) · cites this paper as Jain et al. [4]

    The Results report the opposite direction: mean potassium was 3.7 mmol/l on the blood gas analyzer versus 3.9 mmol/l on the laboratory auto-analyzer, and 71.5% of blood gas readings were lower. The Introduction describes differences in sample type and electrode technology, but does not attribute an upward potassium bias to them. The paper’s conclusion says the potassium measurements did not differ significantly. (LLM, conf=0.90)

    Checked by a reasoning model — may be wrong.

    Experimental — accuracy not yet measured
  • Multiple factors contribute to discrepancies between POCT and laboratory values: variation in sample matrix, pre-analytical handling, device calibration intervals, and analytical methodology.

    “Multiple factors contribute to discrepancies between POCT and laboratory values: variation in sample matrix (whole blood vs. serum/plasma), pre-analytical handling, device calibration intervals, and analytical methodology [4,5].”

    From Reliability of Point-of-Care Venous Blood Gas Testing for Potassium and Sodium Compared to Central Laboratory Results in Emergency Patients: A Clinical Audit (2025) · cites this paper as [4]

    The Introduction (Table 1; page not provided) contrasts whole blood with serum, direct with indirect electrode methods, and different sample dilutions; it also explains how heparin dilution can lower measured electrolyte values. Materials and methods (page not provided) describes sample handling and says both analyzers were calibrated before each measurement cycle. It does not identify variation in device calibration intervals as a cause of discrepancies. (LLM, conf=0.88)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

No references could be resolved yet.

Research works printed but not linked: 15

  1. [1]Price CP Medical and economic outcomes of point of care testing Clin Chem Lab Med 2002 40 246 251 10.1515/CCLM.2002.040 12005214 · DOI
  2. [2]Cox CJ Acute care testing. Bood gases and electrolytes at the point of care Clin Lab Med. 2001 21 2 321 335 11396086
  3. [3]Flegar-Mestric Z Perkov S Comparability of point of care whole blood electrolyte and substrate testing using a Stat Profile Critical Care Xpress analyzer and standard laboratory methods Clin Chem Lab Med 2006 44 7 898 903 10.1515/CCLM.2006.148 16776641 · DOI
  4. [4]Pungor E Working mechanism of ion-selective electrodes Pure & Appl. Chem 1992 64 4 503 507 10.1351/pac199264040503 · DOI
  5. [6]Morimatsu H Rocktäschel J Bellomo R Uchino S Goldsmith D Gutteridge G Comparison of point-of-care versus central laboratory measurement of electrolyte concentrations on calculations of the anion gap and the strong ion difference Anesthesiology 2003 98 1077 1084 10.1097/00000542-200305000-00009 12717128 · DOI
  6. [7]Prichard JS French JS Alvar N Clinical evaluation of the ABL-77 for point of care analysis in the cardiovascular operating room J Extra Corpor Technol 2006 38 2 128 133 16921685 PMC4680748
  7. [10]Standardization of sodium and potassium ion-selective electrode systems to the flame photometric reference method; Approved Standard—Second Edition NCCLS.
  8. [11]Lam HS Chan MH Ng PC Wong W Cheung RC So AK Fok TF Lam CW Are your hands clean enough for point of care electrolyte analysis? Pathology 2005 37 4 299 304 10.1080/00313020500169156 16194829 PMC7131148 · DOI
  9. [12]Steinfelder-Visscher J Weerwind PW Teerenstra S Pop GA Brouwer RM Conductivity based hematocrit meaurement during cardiopulmonary bypass J Clin Monit Comput 2007 21 1 7 12 10.1007/s10877-006-9052-x 17086448 · DOI
  10. [13]Vos G Engel M Ramsay G Waardenburg D Point of care blood analyzer during interhospital transport of critically ill children Eur J Emerg Med 2006 13 5 304 307 10.1097/00063110-200610000-00013 16969239 · DOI
  11. [14]Myers GJ Browne J Point of care hematocrit and hemoglobin in cardiac surgery: a review Perfusion 2007 22 179 183 10.1177/0267659107080826 18018397 · DOI
  12. [15]Magny E Beaudeux JL Launay JM Point care testing in blood gas and electrolyte analysis: examples of implementation and cost analysis Ann Biol Clin (Paris) 2003 61 3 344 351 12805014
  13. [16]Dimeski G Barnett RJ Effects of total plasma protein concentration on plasma sodium, potassium and chloride measurements by an indirect ion selective electrode measuring system Crit Care Resusc 2005 7 1 12 15 16548813
  14. [17]Chow E Fox N Gama R Effect of low serum total protein on sodium and potassium measurement by ion-selective electrodes in critically ill patients Br J Biomed Sci 2008 65 3 128 131 18986099
  15. [18]Daures MF Combescure C Cristol JP Comparitive study of siz blood gas analyzers Ann Biol Clin (Paris) 2007 65 5 505 518 17913669

Cited by

Sorted by citations

Authors
Jain, A., Subhan, I., & Joshi, M.
Affiliations
  • Mayo Clinic (US)
  • Apollo Hospitals (IN)
Corresponding author
Jain, Anunaya
Venue
International Journal of Emergency Medicine
Published
2009-02-24
Volume
2
Issue
2
Pages
117-120
ISSN
  • 1865-1372 (ISSN-L)
  • 1865-1380
OpenAlex
W2051694322
Cited by
108
References
0
Citations by year
2026: 3 · 2025: 9 · 2024: 10 · 2023: 6 · 2022: 11 · 2021: 7 · 2020: 4 · 2019: 8 · 2018: 7 · 2017: 4 · 2016: 8 · 2015: 8 · 2014: 10 · 2013: 4 · 2012: 5
FWCI
1.99
Type
journal article
Language
en
Open access
Locations
  • International Journal of Emergency Medicine (publishedVersion, cc-by-nc)
  • PubMed Central (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
  • Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Electrolyte and hormonal disorders
  • Life Sciences › Biochemistry, Genetics and Molecular Biology › Clinical Biochemistry › Bacterial Identification and Susceptibility Testing
Keywords
electrolyte measurement, sodium, potassium, arterial blood gases, venous blood samples, intensive care, intensive care unit, method comparison

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review